Technical Note: CT calibration for proton treatment planning by cross-calibration with proton CT data

被引:7
|
作者
Farace, Paolo [1 ,2 ]
Tommasino, Francesco [2 ,3 ]
Righetto, Roberto [1 ,2 ]
Fracchiolla, Francesco [1 ,2 ]
Scaringella, Monica [4 ,5 ]
Bruzzi, Mara [4 ]
Civinini, Carlo [4 ]
机构
[1] Hosp Trento, Protontherapy Unit, Azienda Prov & Serv Sanit APSS, Trento, Italy
[2] Ist Nazl Fis Nucl TIFPA, Via Sommar 14, Trento, Italy
[3] Univ Trento, Dept Phys, Via Sommar 14, Trento, Italy
[4] Ist Nazl Fis Nucl, Sez Firenze, Via G Sansone 1, Sesto Fiorentino, Italy
[5] Univ Firenze, Dipartimento Fis & Astron, Via G Sansone 1, Sesto Fiorentino, Italy
关键词
CT; proton CT; proton therapy;
D O I
10.1002/mp.14698
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose This study explores the possibility of a new method for x-ray computed tomography (CT) calibration by means of cross-calibration with proton CT (pCT) data. The proposed method aims at a more accurate conversion of CT Hounsfield Units (HU) into proton stopping power ratio (SPR) relative to water to be used in proton-therapy treatment planning. Methods X-ray CT scan was acquired on a synthetic anthropomorphic phantom, composed of different tissue equivalent materials (TEMs). A pCT apparatus was instead adopted to obtain a reference three-dimensional distribution of the phantom's SPR values. After rigid registration, the x-ray CT was artificially blurred to the same resolution of pCT. Then a scatter plot showing voxel-by-voxel SPR values as a function of HU was employed to link the two measurements and thus obtaining a cross-calibrated x-ray CT calibration curve. The cross-calibration was tested at treatment planning system and then compared with a conventional calibration based on exactly the same TEMs constituting the anthropomorphic phantom. Results Cross-calibration provided an accurate SPR mapping, better than by conventional TEMs calibration. The dose distribution of single beams optimized on the reference SPR map was recomputed on cross-calibrated CT, showing, with respect to conventional calibration, minor deviation at the dose fall-off (lower than 1%). Conclusions The presented data demonstrated that, by means of reference pCT data, a heterogeneous phantom can be used for CT calibration, paving the way to the use of biological samples, with their accurate description of patients' tissues. This overcomes the limitations of conventional CT calibration requiring homogenous samples, only available by synthetic TEMs, which fail in accurately mimicking the properties of biological tissues. Once a heterogeneous biological sample is provided with its corresponding reference SPR maps, a cross-calibration procedure could be adopted by other PT centers, even when not equipped with a pCT system.
引用
收藏
页码:1349 / 1355
页数:7
相关论文
共 50 条
  • [1] Characterization of the INFN proton CT scanner for cross-calibration of x-ray CT
    Fogazzi, Elena
    Trevisan, Diego
    Farace, Paolo
    Righetto, Roberto
    Rit, Simon
    Scaringella, Monica
    Bruzzi, Mara
    Tommasino, Francesco
    Civinini, Carlo
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (12):
  • [2] Proton CT on biological phantoms for x-ray CT calibration in proton treatment planning
    Fogazzi, Elena
    Bruzzi, Mara
    D'Amato, Elvira
    Farace, Paolo
    Righetto, Roberto
    Scaringella, Monica
    Scarpa, Marina
    Tommasino, Francesco
    Civinini, Carlo
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (13):
  • [3] CT Calibration for Proton Therapy Planning of Pediatric Patients
    Baer, E.
    Collins-Fekete, C. A.
    Rompokos, V.
    Warry, A.
    Poynter, A.
    Royle, G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E258 - E259
  • [4] CT calibration for precise proton therapy planning in children
    Baer, E.
    Collins-Fekete, C.
    Rompokos, V.
    Zhang, Y.
    Gaze, M.
    Warry, A.
    Poynter, A.
    Royle, G.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S404 - S406
  • [5] Cross-calibration of ionization chambers in proton and carbon beams
    Kanai, T
    Fukumura, A
    Kusano, Y
    Shimbo, M
    Nishio, T
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (05): : 771 - 781
  • [6] Calibration of a Prototype Proton CT Scanner
    Hurley, F.
    Schulte, R.
    Bashkirov, V.
    Wroe, A.
    Ghebremedhin, A.
    Patyal, B.
    Penfold, S.
    Rykalin, V.
    Sadrozinsk, H.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [7] Peripheral quantitative CT of the forearm: scanner cross-calibration using patient data
    Grampp, S
    Nather, A
    Rintelen, B
    Henk, C
    Resch-Holeczke, A
    Imhof, H
    Resch, H
    BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (867): : 275 - 277
  • [8] Real patient data based cross verification of kilovoltage and megavoltage CT calibration for proton therapy
    Nishioka, Shie
    Park, Joo Han
    Zou, Wei
    Zhang, Miao
    Yue, Ning J.
    Chen, Ting
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (02): : 343 - 352
  • [9] Assessment of the impact of CT calibration procedures for proton therapy planning on pediatric treatments
    Bar, Esther
    Collins-Fekete, Charles-Antoine
    Rompokos, Vasilis
    Zhang, Ying
    Gaze, Mark N.
    Warry, Alison
    Poynter, Andrew
    Royle, Gary
    MEDICAL PHYSICS, 2021, 48 (09) : 5202 - 5218
  • [10] Practical considerations in the calibration of CT scanners for proton therapy
    Ainsley, Christopher G.
    Yeager, Caitlyn M.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (03): : 202 - 220