Technical Note: CT calibration for proton treatment planning by cross-calibration with proton CT data

被引:7
|
作者
Farace, Paolo [1 ,2 ]
Tommasino, Francesco [2 ,3 ]
Righetto, Roberto [1 ,2 ]
Fracchiolla, Francesco [1 ,2 ]
Scaringella, Monica [4 ,5 ]
Bruzzi, Mara [4 ]
Civinini, Carlo [4 ]
机构
[1] Hosp Trento, Protontherapy Unit, Azienda Prov & Serv Sanit APSS, Trento, Italy
[2] Ist Nazl Fis Nucl TIFPA, Via Sommar 14, Trento, Italy
[3] Univ Trento, Dept Phys, Via Sommar 14, Trento, Italy
[4] Ist Nazl Fis Nucl, Sez Firenze, Via G Sansone 1, Sesto Fiorentino, Italy
[5] Univ Firenze, Dipartimento Fis & Astron, Via G Sansone 1, Sesto Fiorentino, Italy
关键词
CT; proton CT; proton therapy;
D O I
10.1002/mp.14698
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose This study explores the possibility of a new method for x-ray computed tomography (CT) calibration by means of cross-calibration with proton CT (pCT) data. The proposed method aims at a more accurate conversion of CT Hounsfield Units (HU) into proton stopping power ratio (SPR) relative to water to be used in proton-therapy treatment planning. Methods X-ray CT scan was acquired on a synthetic anthropomorphic phantom, composed of different tissue equivalent materials (TEMs). A pCT apparatus was instead adopted to obtain a reference three-dimensional distribution of the phantom's SPR values. After rigid registration, the x-ray CT was artificially blurred to the same resolution of pCT. Then a scatter plot showing voxel-by-voxel SPR values as a function of HU was employed to link the two measurements and thus obtaining a cross-calibrated x-ray CT calibration curve. The cross-calibration was tested at treatment planning system and then compared with a conventional calibration based on exactly the same TEMs constituting the anthropomorphic phantom. Results Cross-calibration provided an accurate SPR mapping, better than by conventional TEMs calibration. The dose distribution of single beams optimized on the reference SPR map was recomputed on cross-calibrated CT, showing, with respect to conventional calibration, minor deviation at the dose fall-off (lower than 1%). Conclusions The presented data demonstrated that, by means of reference pCT data, a heterogeneous phantom can be used for CT calibration, paving the way to the use of biological samples, with their accurate description of patients' tissues. This overcomes the limitations of conventional CT calibration requiring homogenous samples, only available by synthetic TEMs, which fail in accurately mimicking the properties of biological tissues. Once a heterogeneous biological sample is provided with its corresponding reference SPR maps, a cross-calibration procedure could be adopted by other PT centers, even when not equipped with a pCT system.
引用
收藏
页码:1349 / 1355
页数:7
相关论文
共 50 条
  • [31] Robustness of three stoichiometric CT-calibration parameterizations used in proton therapy
    Oden, J.
    Zimmerman, J.
    Poludniowski, G.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S223 - S224
  • [32] Impact of Image Reconstruction Kernel On CT Number to Proton Stopping Power Calibration
    Chacko, M.
    Grewal, H.
    Rana, S.
    Wu, D.
    Sonnad, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E625 - E625
  • [33] A critical look at the stoichiometric single-energy CT calibration for proton therapy
    Goma, C.
    Almeida, I. P.
    Verhaegen, F.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1139 - S1140
  • [34] CROSS-CALIBRATION OF LIQUID AND SOLID QCT CALIBRATION STANDARDS - CORRECTIONS TO THE UCSF NORMATIVE DATA
    FAULKNER, KG
    GLUER, CC
    GRAMPP, S
    GENANT, HK
    OSTEOPOROSIS INTERNATIONAL, 1993, 3 (01) : 36 - 42
  • [35] A constancy test to monitor cross-calibration factors for a small animal SPECT-CT scanner
    Vines, Douglass
    Green, David
    Keller, Harald
    Breen, Stephen
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [36] Validation of the Implementation of Different CT Scanners in Proton Treatment Planning
    Fuentes, C. Llina
    Geismer, D.
    Christiaens, M.
    Vermeren, X.
    Ding, X.
    MEDICAL PHYSICS, 2015, 42 (06) : 3226 - 3226
  • [37] Impact of CT Metal Artifacts on Proton Radiotherapy Treatment Planning
    Verburg, J. M.
    Joshi, M. C.
    Madden, T. M.
    Kooy, H. M.
    Seco, J. C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S809 - S810
  • [38] Review of 3D image data calibration for heterogeneity correction in proton therapy treatment planning
    Jiahua Zhu
    Scott N. Penfold
    Australasian Physical & Engineering Sciences in Medicine, 2016, 39 : 379 - 390
  • [39] Review of 3D image data calibration for heterogeneity correction in proton therapy treatment planning
    Zhu, Jiahua
    Penfold, Scott N.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (02) : 379 - 390
  • [40] The LSO/KSO Hα prominence catalogue: cross-calibration of data
    Rybak, J.
    Goemoery, P.
    Macura, R.
    Kucera, A.
    Rusin, V.
    Poetzi, W.
    Baumgartner, D.
    Hanslmeier, A.
    Veronig, A.
    Temmer, M.
    CONTRIBUTIONS OF THE ASTRONOMICAL OBSERVATORY SKALNATE PLESO, 2011, 41 (02): : 133 - 136