Technical Note: CT calibration for proton treatment planning by cross-calibration with proton CT data

被引:7
|
作者
Farace, Paolo [1 ,2 ]
Tommasino, Francesco [2 ,3 ]
Righetto, Roberto [1 ,2 ]
Fracchiolla, Francesco [1 ,2 ]
Scaringella, Monica [4 ,5 ]
Bruzzi, Mara [4 ]
Civinini, Carlo [4 ]
机构
[1] Hosp Trento, Protontherapy Unit, Azienda Prov & Serv Sanit APSS, Trento, Italy
[2] Ist Nazl Fis Nucl TIFPA, Via Sommar 14, Trento, Italy
[3] Univ Trento, Dept Phys, Via Sommar 14, Trento, Italy
[4] Ist Nazl Fis Nucl, Sez Firenze, Via G Sansone 1, Sesto Fiorentino, Italy
[5] Univ Firenze, Dipartimento Fis & Astron, Via G Sansone 1, Sesto Fiorentino, Italy
关键词
CT; proton CT; proton therapy;
D O I
10.1002/mp.14698
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose This study explores the possibility of a new method for x-ray computed tomography (CT) calibration by means of cross-calibration with proton CT (pCT) data. The proposed method aims at a more accurate conversion of CT Hounsfield Units (HU) into proton stopping power ratio (SPR) relative to water to be used in proton-therapy treatment planning. Methods X-ray CT scan was acquired on a synthetic anthropomorphic phantom, composed of different tissue equivalent materials (TEMs). A pCT apparatus was instead adopted to obtain a reference three-dimensional distribution of the phantom's SPR values. After rigid registration, the x-ray CT was artificially blurred to the same resolution of pCT. Then a scatter plot showing voxel-by-voxel SPR values as a function of HU was employed to link the two measurements and thus obtaining a cross-calibrated x-ray CT calibration curve. The cross-calibration was tested at treatment planning system and then compared with a conventional calibration based on exactly the same TEMs constituting the anthropomorphic phantom. Results Cross-calibration provided an accurate SPR mapping, better than by conventional TEMs calibration. The dose distribution of single beams optimized on the reference SPR map was recomputed on cross-calibrated CT, showing, with respect to conventional calibration, minor deviation at the dose fall-off (lower than 1%). Conclusions The presented data demonstrated that, by means of reference pCT data, a heterogeneous phantom can be used for CT calibration, paving the way to the use of biological samples, with their accurate description of patients' tissues. This overcomes the limitations of conventional CT calibration requiring homogenous samples, only available by synthetic TEMs, which fail in accurately mimicking the properties of biological tissues. Once a heterogeneous biological sample is provided with its corresponding reference SPR maps, a cross-calibration procedure could be adopted by other PT centers, even when not equipped with a pCT system.
引用
收藏
页码:1349 / 1355
页数:7
相关论文
共 50 条
  • [21] Treatment Planning and Verification with Proton CT and Proton Radiography to Reduce Range Uncertainties in Proton Therapy
    Schulte, R.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S242 - S242
  • [22] Technical Note: On the calculation of stopping-power ratio for stoichiometric calibration in proton therapy
    Oden, Jakob
    Zimmerman, Jens
    Bujila, Robert
    Nowik, Patrik
    Poludniowski, Gavin
    MEDICAL PHYSICS, 2015, 42 (09) : 5252 - 5257
  • [23] The calibration of CT Hounsfield units for radiotherapy treatment planning - Reply
    Schneider, U
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (08): : 1526 - 1527
  • [24] The calibration of CT Hounsfield units for radiotherapy treatment planning - Comment
    deKock, EA
    Schreuder, AN
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (08): : 1524 - 1526
  • [25] Material decomposition maps based calibration of dual energy CT scanners for proton therapy planning: a phantom study
    Viar-Hernandez, David
    Vera-Sanchez, Juan Antonio
    Schmidt-Santiago, Lucia
    Rodriguez-Vila, Borja
    Lorenzo-Villanueva, Isabel
    Canals-de-las-Casas, Elisabet
    Castro-Novais, Juan
    Perez-Moreno, Juan Maria
    Cerron-Campoo, Fernando
    Malpica, Norberto
    Torrado-Carvajal, Angel
    Mazal, Alejandro
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (04):
  • [26] Cross Calibration of the GPS Constellation CXD Proton Data With GOES EPS
    Carver, Matthew R.
    Sullivan, John P.
    Morley, Steven K.
    Rodriguez, Juan V.
    SPACE WEATHER-THE INTERNATIONAL JOURNAL OF RESEARCH AND APPLICATIONS, 2018, 16 (03): : 273 - 288
  • [27] Water-equivalent path length calibration of a prototype proton CT scanner
    Hurley, R. F.
    Schulte, R. W.
    Bashkirov, V. A.
    Wroe, A. J.
    Ghebremedhin, A.
    Sadrozinski, H. F. -W.
    Rykalin, V.
    Coutrakon, G.
    Koss, P.
    Patyal, B.
    MEDICAL PHYSICS, 2012, 39 (05) : 2438 - 2446
  • [28] THE CROSS-CALIBRATION METHOD FOR "SICH-2" DATA
    Basarab, R. M.
    SPACE SCIENCE AND TECHNOLOGY-KOSMICNA NAUKA I TEHNOLOGIA, 2014, 20 (01): : 44 - 50
  • [29] Adjustment of CT calibration in presence of titanium implants by pencil beam proton radiography
    Righetto, R.
    Meijers, A.
    Vander Stappen, F.
    Farace, P.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S441 - S442
  • [30] Validation of CT Calibration for Proton Therapy Using Water-Based Materials
    Cohilis, M.
    Sterpin, E.
    Salvo, K.
    Depuydt, T.
    Lee, J.
    Souris, K.
    MEDICAL PHYSICS, 2020, 47 (06) : E877 - E877