Technical Note: CT calibration for proton treatment planning by cross-calibration with proton CT data

被引:7
|
作者
Farace, Paolo [1 ,2 ]
Tommasino, Francesco [2 ,3 ]
Righetto, Roberto [1 ,2 ]
Fracchiolla, Francesco [1 ,2 ]
Scaringella, Monica [4 ,5 ]
Bruzzi, Mara [4 ]
Civinini, Carlo [4 ]
机构
[1] Hosp Trento, Protontherapy Unit, Azienda Prov & Serv Sanit APSS, Trento, Italy
[2] Ist Nazl Fis Nucl TIFPA, Via Sommar 14, Trento, Italy
[3] Univ Trento, Dept Phys, Via Sommar 14, Trento, Italy
[4] Ist Nazl Fis Nucl, Sez Firenze, Via G Sansone 1, Sesto Fiorentino, Italy
[5] Univ Firenze, Dipartimento Fis & Astron, Via G Sansone 1, Sesto Fiorentino, Italy
关键词
CT; proton CT; proton therapy;
D O I
10.1002/mp.14698
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose This study explores the possibility of a new method for x-ray computed tomography (CT) calibration by means of cross-calibration with proton CT (pCT) data. The proposed method aims at a more accurate conversion of CT Hounsfield Units (HU) into proton stopping power ratio (SPR) relative to water to be used in proton-therapy treatment planning. Methods X-ray CT scan was acquired on a synthetic anthropomorphic phantom, composed of different tissue equivalent materials (TEMs). A pCT apparatus was instead adopted to obtain a reference three-dimensional distribution of the phantom's SPR values. After rigid registration, the x-ray CT was artificially blurred to the same resolution of pCT. Then a scatter plot showing voxel-by-voxel SPR values as a function of HU was employed to link the two measurements and thus obtaining a cross-calibrated x-ray CT calibration curve. The cross-calibration was tested at treatment planning system and then compared with a conventional calibration based on exactly the same TEMs constituting the anthropomorphic phantom. Results Cross-calibration provided an accurate SPR mapping, better than by conventional TEMs calibration. The dose distribution of single beams optimized on the reference SPR map was recomputed on cross-calibrated CT, showing, with respect to conventional calibration, minor deviation at the dose fall-off (lower than 1%). Conclusions The presented data demonstrated that, by means of reference pCT data, a heterogeneous phantom can be used for CT calibration, paving the way to the use of biological samples, with their accurate description of patients' tissues. This overcomes the limitations of conventional CT calibration requiring homogenous samples, only available by synthetic TEMs, which fail in accurately mimicking the properties of biological tissues. Once a heterogeneous biological sample is provided with its corresponding reference SPR maps, a cross-calibration procedure could be adopted by other PT centers, even when not equipped with a pCT system.
引用
收藏
页码:1349 / 1355
页数:7
相关论文
共 50 条
  • [41] Relative electron density calibration of CT scanners for radiotherapy treatment planning
    Thomas, SJ
    BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (860): : 781 - 786
  • [42] Validation of the proton range accuracy and optimization of CT calibration curves utilizing range probing
    Meijers, A.
    Free, J.
    Wagenaar, D.
    Deffet, S.
    Knopf, A. C.
    Langendijk, J. A.
    Both, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (03):
  • [43] Dual energy CT proton stopping power ratio calibration: Validation with animal tissues
    Xie, Y.
    Yin, L.
    Ainsley, C.
    McDonough, J.
    Solberg, T.
    Lin, A.
    Teo, B. K.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S35 - S36
  • [44] Dual Energy CT Proton Stopping Power Ratio Calibration and Validation with Animal Tissues
    Xie, Y.
    Yin, L.
    Ainsley, C.
    McDonough, J.
    Solberg, T.
    Lin, A.
    Teo, B.
    MEDICAL PHYSICS, 2016, 43 (06) : 3756 - 3756
  • [45] A novel range probing-based optimization of CT calibration curve for Proton Therapy
    Meijers, A.
    Free, J.
    Wagenaar, D.
    Deffet, S.
    Knopf, A. C.
    Langendijk, J. A.
    Both, S.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S294 - S295
  • [46] Deformable image registration of the treatment planning CT with proton radiographies in perspective of adaptive proton therapy
    Palaniappan, Prasannakumar
    Meyer, Sebastian
    Kamp, Florian
    Belka, Claus
    Riboldi, Marco
    Parodi, Katia
    Gianoli, Chiara
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (04):
  • [47] Improving CT calibration for proton range prediction by dual-energy CT based patient-cohort analysis
    Wohlfahrt, P.
    Moehler, C.
    Enghardt, W.
    Krause, M.
    Troost, E. G. C.
    Greilich, S.
    Richter, C.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S43 - S44
  • [48] Merging proton radiographies with treatment planning CT for adaptive radiation therapy
    Gianoli, C.
    Dedes, G.
    Meyer, S.
    Magallanes, L.
    Landry, G.
    Nijhuis, R.
    Ganswindt, U.
    Thieke, C.
    Belka, C.
    Parodi, K.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S438 - S438
  • [49] The improvement of cross-calibration of IIM data and band selection for FeO inversion
    Wang Chao
    Wu YunZhao
    Shi RunHe
    Gao Wei
    SCIENCE CHINA-PHYSICS MECHANICS & ASTRONOMY, 2013, 56 (11) : 2216 - 2225
  • [50] Breast specific cone-beam CT calibration for daily dose verification in proton therapy
    Omand Kirkegaard Sorensen, T.
    Muren, L.
    Elstrom, U. V.
    Kronborg, C.
    Offersen, B. V.
    Petersen, S. E.
    Stick, L. B.
    Taasti, V. T.
    Jensen, M. F.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S130 - S132