Cross-calibration of ionization chambers in proton and carbon beams

被引:34
|
作者
Kanai, T
Fukumura, A
Kusano, Y
Shimbo, M
Nishio, T
机构
[1] Natl Inst Radiol Sci, Inage Ku, Chiba 2638555, Japan
[2] Accelerator Engn Corp, Inage Ku, Chiba 2630043, Japan
[3] East Hosp, Natl Canc Ctr, Kashiwa, Chiba 2778577, Japan
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2004年 / 49卷 / 05期
关键词
D O I
10.1088/0031-9155/49/5/008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The calibration coefficients of a parallel plate ionization chamber are examined by comparing the coefficients obtained through three methods: a calculation from a Co-60 calibration coefficient, N-D,N-W,N-60Co, a cross-calibration of a parallel plate ionization chamber using a cylindrical ionization chamber at the plateau region of a mono-energetic beam and a cross-calibration of the chamber using a cylindrical chamber at the middle of the SOBP of the therapeutic beams. This paper also examines reference conditions for determining absorbed dose to water in the cases of therapeutic carbon and proton beams. In the dose calibration procedure recommended by IAEA, irradiation fields should be larger than 10 cm in diameter and the water phantom should extend by at least 5 cm beyond each side of the field. These recommendations are experimentally verified for proton and carbon beams. For proton beams, the calibration coefficients obtained by these three methods approximately agreed. For carbon beams, the calibration coefficients obtained by the second method were about 1.0% larger than those obtained by the third method, and the calibration coefficients obtained by cross-calibration using 290 MeV/u beams were 0.5% lower than those obtained using 400 MeV/u beams. The calibration coefficient obtained by the first method agreed roughly with the results obtained by SOBP beams.
引用
收藏
页码:771 / 781
页数:11
相关论文
共 50 条
  • [1] Direct calibration of ionization chambers in linac photon beams
    McEwen, M
    Ross, C
    MEDICAL PHYSICS, 2005, 32 (06) : 2128 - 2128
  • [2] Determination of Magnetic Field Correction Factors K(Q,B) for Ionization Chambers Using Alanine Cross-Calibration
    Pojtinger, S.
    Nachbar, M.
    Kapsch, R-P
    Thorwarth, D.
    MEDICAL PHYSICS, 2020, 47 (06) : E740 - E740
  • [3] Technical Note: CT calibration for proton treatment planning by cross-calibration with proton CT data
    Farace, Paolo
    Tommasino, Francesco
    Righetto, Roberto
    Fracchiolla, Francesco
    Scaringella, Monica
    Bruzzi, Mara
    Civinini, Carlo
    MEDICAL PHYSICS, 2021, 48 (03) : 1349 - 1355
  • [4] CROSS-CALIBRATION TECHNIQUES
    HOLLAND, JZ
    BULLETIN OF THE AMERICAN METEOROLOGICAL SOCIETY, 1971, 52 (11) : 1159 - &
  • [5] Determination of the Uncertainty in the Cross-calibration of an Ionization Chamber Used in Radiation Therapy
    Cardoso, Pedro H. B.
    Tietz, Gustavo F.
    Neves-Junior, Wellington F. P.
    Haddad, Cecilia M. K.
    Terini, Ricardo A.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 756 - 758
  • [6] The use of plane-parallel chambers in electron dosimetry without any cross-calibration
    Christ, G
    Dohm, OS
    Bruggmoser, G
    Schüle, E
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (09): : N121 - N126
  • [7] Investigation of the response of cylindrical ionization chambers in magnetic field in proton beams
    Blum, Isabel
    Wong, Jing Syuen
    Padre, Krishna Godino
    Baumann, Kilian-Simon
    Fuchs, Hermann
    Poppe, Bjoern
    Looe, Hui Khee
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3381 - S3383
  • [8] The Effective Depth of Cylindrical Ionization Chambers in Water for Clinical Proton Beams
    Bhullar, A.
    Watchman, C.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [9] ION RECOMBINATION LOSS IN IONIZATION CHAMBERS IRRADIATED BY PROTON-BEAMS
    HIRAOKA, T
    KAWASHIMA, K
    HOSHINO, K
    BRITISH JOURNAL OF RADIOLOGY, 1982, 55 (656): : 585 - 587
  • [10] The effective depth of cylindrical ionization chambers in water for clinical proton beams
    Bhullar, Amarjeet S.
    Watchman, Christopher J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (01): : 273 - 286