A bent Laue-Laue monochromator for a synchrotron-based computed tomography system

被引:13
|
作者
Ren, B [1 ]
Dilmanian, FA
Chapman, LD
Ivanov, I
Wu, XY
Zhong, Z
Huang, X
机构
[1] Brookhaven Natl Lab, Dept Med, Upton, NY 11973 USA
[2] IIT, Ctr Synchrotron Radiat Res & Instrumentat, Chicago, IL 60616 USA
[3] Brookhaven Natl Lab, Natl Synchrotron Light Source Dept, Upton, NY 11973 USA
关键词
bent crystal monochromator; beam smiling; harmonics rejection; Laue spot; CT;
D O I
10.1016/S0168-9002(99)00056-X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We designed and tested a two-crystal bent Laue-Laue monochromator for wide, fan-shaped synchrotron X-ray beams for the program multiple energy computed tomography (MECT) at the National Synchrotron Light Source (NSLS). MECT employs monochromatic X-ray beams from the NSLS's X17B superconducting wiggler beamline for computed tomography (CT) with an improved image quality. MECT uses a fixed horizontal fan-shaped beam with the subject's apparatus rotating around a vertical axis. The new monochromator uses two Czochralski-grown Si[1 1 1] crystals, 0.7 and 1.4 mm thick, respectively, and with thick ribs on their upper and lower ends. The crystals are bent cylindrically, with the axis of the cylinder parallel to the fan beam, using 4-rod benders with two fixed rods and two movable ones. The bent-crystal feature of the monochromator resolved the difficulties we had had with the flat Laue-Laue design previously used in MECT, which included (a) inadequate beam intensity, (b) excessive fluctuations in beam intensity, and (c) instability of the shape of the beam's horizontal profile. Compared with that earlier monochromator, the bent Laue-Laue device tested at 42 and 108 keV showed about a 10-fold larger beam flux, about 5 times better beam stability, 10-fold less harmonic contamination, and a smaller energy bandwidth at certain bending radii. The present work gave us better understanding of the basis for the beam-smiling effect in bent-crystal monochromators, and allowed us to refine the theoretical method of estimating the beam-harmonic contamination in bent-crystal monochromators. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:528 / 550
页数:23
相关论文
共 50 条
  • [1] A bent Laue-Laue monochromator for a synchrotron-based monochromatic x-ray computed tomography system
    Ren, BR
    MEDICAL PHYSICS, 1998, 25 (12) : 2483 - 2483
  • [2] Ultrasound adjustable Laue-Laue double-crystal x-ray monochromator
    Polikarpov, I
    Panov, VV
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1999, 32 (16) : 2083 - 2086
  • [3] PERFORMANCE EVALUATION OF A BENT LAUE MONOCHROMATOR
    SUORTTI, P
    THOMLINSON, W
    CHAPMAN, D
    GMUR, N
    GREENE, R
    LAZARZ, N
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 297 (1-2): : 268 - 274
  • [4] A BENT LAUE CRYSTAL MONOCHROMATOR FOR ANGIOGRAPHY AT THE NSLS
    SUORTTI, P
    THOMLINSON, W
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 269 (03): : 639 - 648
  • [5] BENT LAUE CRYSTAL MONOCHROMATOR FOR ANGIOGRAPHY AT THE NSLS.
    Suortti, P.
    Thomlinson, W.
    Nuclear instruments and methods in physics research, 1988, A269 (03): : 639 - 648
  • [6] A High Quality Bent Crystal Monochromator based on Asymmetric Laue Geometry
    Hu, Hua-Chen
    ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 2005, 61 : C142 - C142
  • [7] Bent Laue analyser crystal for Rayleigh-to-Compton computed tomography
    Schulze, C
    Kleuker, U
    JOURNAL OF SYNCHROTRON RADIATION, 1998, 5 : 1085 - 1087
  • [8] A bent Laue analyser crystal for Rayleigh-to-Compton computed tomography
    Swiss Light Source, PSI, CH-5232 Villigen PSI, Switzerland
    不详
    J. Synchrotron Radiat., 3 (1085-1087):
  • [9] DOUBLE-BEAM BENT LAUE MONOCHROMATOR FOR CORONARY ANGIOGRAPHY
    ILLING, G
    HEUER, J
    REIME, B
    LOHMANN, M
    MENK, RH
    SCHILDWACHTER, L
    DIX, WR
    GRAEFF, W
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02): : 1379 - 1381
  • [10] A SINGLE-CRYSTAL BENT LAUE MONOCHROMATOR FOR CORONARY ANGIOGRAPHY
    SUORTTI, P
    THOMLINSON, W
    CHAPMAN, D
    GMUR, N
    SIDDONS, DP
    SCHULZE, C
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 336 (1-2): : 304 - 309