A bent Laue-Laue monochromator for a synchrotron-based computed tomography system

被引:13
|
作者
Ren, B [1 ]
Dilmanian, FA
Chapman, LD
Ivanov, I
Wu, XY
Zhong, Z
Huang, X
机构
[1] Brookhaven Natl Lab, Dept Med, Upton, NY 11973 USA
[2] IIT, Ctr Synchrotron Radiat Res & Instrumentat, Chicago, IL 60616 USA
[3] Brookhaven Natl Lab, Natl Synchrotron Light Source Dept, Upton, NY 11973 USA
关键词
bent crystal monochromator; beam smiling; harmonics rejection; Laue spot; CT;
D O I
10.1016/S0168-9002(99)00056-X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We designed and tested a two-crystal bent Laue-Laue monochromator for wide, fan-shaped synchrotron X-ray beams for the program multiple energy computed tomography (MECT) at the National Synchrotron Light Source (NSLS). MECT employs monochromatic X-ray beams from the NSLS's X17B superconducting wiggler beamline for computed tomography (CT) with an improved image quality. MECT uses a fixed horizontal fan-shaped beam with the subject's apparatus rotating around a vertical axis. The new monochromator uses two Czochralski-grown Si[1 1 1] crystals, 0.7 and 1.4 mm thick, respectively, and with thick ribs on their upper and lower ends. The crystals are bent cylindrically, with the axis of the cylinder parallel to the fan beam, using 4-rod benders with two fixed rods and two movable ones. The bent-crystal feature of the monochromator resolved the difficulties we had had with the flat Laue-Laue design previously used in MECT, which included (a) inadequate beam intensity, (b) excessive fluctuations in beam intensity, and (c) instability of the shape of the beam's horizontal profile. Compared with that earlier monochromator, the bent Laue-Laue device tested at 42 and 108 keV showed about a 10-fold larger beam flux, about 5 times better beam stability, 10-fold less harmonic contamination, and a smaller energy bandwidth at certain bending radii. The present work gave us better understanding of the basis for the beam-smiling effect in bent-crystal monochromators, and allowed us to refine the theoretical method of estimating the beam-harmonic contamination in bent-crystal monochromators. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:528 / 550
页数:23
相关论文
共 50 条
  • [41] Structural properties in ruptured mitral chordae tendineae measured by synchrotron-based X-ray phase computed tomography
    Koda, Yojiro
    Tsukube, Takuro
    Hoshino, Masato
    Yagi, Naoto
    Ishibashi-Ueda, Hatsue
    Okada, Kenji
    JOURNAL OF SYNCHROTRON RADIATION, 2023, 30 : 995 - 1002
  • [42] Structural properties in ruptured mitral chordae tendineae measured by synchrotron-based X-ray phase computed tomography
    Koda, Yojiro
    Tsukube, Takuro
    Hoshino, Masato
    Yagi, Naoto
    Ishibashi-Ueda, Hatsue
    Okada, Kenji
    JOURNAL OF SYNCHROTRON RADIATION, 2023, 30 (05) : 995 - 1002
  • [43] Fixed-exit monochromator for computed tomography with synchrotron radiation at energies 18-90 keV
    Suortti, P
    Fiedler, S
    Bravin, A
    Brochard, T
    Mattenet, M
    Renier, M
    Spanne, P
    Thomlinson, W
    Charvet, AM
    Elleaume, H
    Schulze-Briese, C
    Thompson, AC
    JOURNAL OF SYNCHROTRON RADIATION, 2000, 7 (05) : 340 - 347
  • [44] Nondestructive material characterization of meteorites with synchrotron-based high energy x-ray phase micro-computed tomography
    Liu, Huiqiang
    Xiao, Tiqiao
    Xie, Honglan
    Fu, Yanan
    Zhang, Xueliang
    Fan, Xiaoxi
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2017, 50 (05)
  • [45] A synchrotron-based local computed tomography combined with data-constrained modelling approach for quantitative analysis of anthracite coal microstructure
    Chen, Wen Hao
    Yang, Sam Y. S.
    Xiao, Ti Qiao
    Mayo, Sherry C.
    Wang, Yu Dan
    Wang, Hai Peng
    JOURNAL OF SYNCHROTRON RADIATION, 2014, 21 : 586 - 593
  • [46] Utilizing Synchrotron-Based X-ray Micro-Computed Tomography to Visualize the Microscopic Structure of Starch Hydrogels In Situ
    Ren, Yikai
    Stobbs, Jarvis A.
    Lee, Dong-Jin
    Li, Dongxing
    Karunakaran, Chithra
    Ai, Yongfeng
    BIOMACROMOLECULES, 2024, 25 (06) : 3302 - 3311
  • [47] INTEGRATION OF A SYNCHROTRON-BASED X-RAY-LITHOGRAPHY SYSTEM
    LI, DC
    SUN, CY
    CULLMANN, E
    JOURNAL OF SCIENTIFIC & INDUSTRIAL RESEARCH, 1994, 53 (10): : 745 - 757
  • [48] A 1200 ELEMENT DETECTOR SYSTEM FOR SYNCHROTRON-BASED CORONARY ANGIOGRAPHY
    THOMPSON, AC
    LAVENDER, WM
    CHAPMAN, D
    GMUR, N
    THOMLINSON, W
    ROSSO, V
    SCHULZE, C
    RUBENSTEIN, E
    GIACOMINI, JC
    GORDON, HJ
    DERVAN, JP
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 347 (1-3): : 545 - 552
  • [49] Investigation of the imaging quality of synchrotron-based phase-contrast mammographic tomography
    Gureyev, T. E.
    Mayo, S. C.
    Nesterets, Ya I.
    Mohammadi, S.
    Lockie, D.
    Menk, R. H.
    Arfelli, F.
    Pavlov, K. M.
    Kitchen, M. J.
    Zanconati, F.
    Dullin, C.
    Tromba, G.
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2014, 47 (36)
  • [50] Convolution reconstruction algorithm for refraction-contrast computed tomography using a Laue-case analyzer for dark-field imaging
    Sunaguchi, Naoki
    Yuasa, Tetsuya
    Huo, Qingkai
    Ando, Masami
    OPTICS LETTERS, 2011, 36 (03) : 391 - 393