Verification of high dose rate 192Ir source position during brachytherapy treatment

被引:16
|
作者
Batic, M. [1 ]
Burger, J. [2 ]
Cindro, V. [1 ]
Kramberger, G. [1 ]
Mandic, I. [1 ]
Mikuz, M. [1 ,3 ]
Studen, A. [1 ]
Zavrtanik, M. [1 ]
机构
[1] Jozef Stefan Inst, Expt Particle Phys Dept, Ljubljana, Slovenia
[2] Inst Oncol, Dept Radiophys, Ljubljana, Slovenia
[3] Univ Ljubljana, Dept Phys, Ljubljana 61000, Slovenia
关键词
Source localization; Brachytherapy; Silicon pixel detectors; Verification; CAMERA;
D O I
10.1016/j.nima.2009.09.122
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A system for in vivo tracking of 1 Ci Ir-192 source during brachytherapy treatment has been built using high resistivity silicon pad detectors as image sensors and knife-edge lead pinholes as collimators. The sensors consist of 256 pads arranged in 32 x 8 grid with pad size 1.4 x 1.4 mm(2) and 1 mm thickness. The sensors have two metal layers, enabling connection of readout electronics (VATAGP3_1 chips) at the edge of the detector. With source self-images obtained from a dual-pinhole system, location of the source can be reconstructed in three dimensions in real time, allowing on-line detection of deviations from planned treatment. The system was tested with 1 Ci Ir-192 clinical source in air and plexi-glass phantom. The movements of the source could be tracked in a field of view of approximately 20 x 20 x 20 cm(3) with absolute precision of about 5 mm. Positions of the source, relative to the first measured source position, could be mapped with precision of around 3 mm. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:206 / 208
页数:3
相关论文
共 50 条
  • [41] An easy and novel method for safer brachytherapy: real-time fluoroscopic verification of high-dose-rate 192Ir source position using a flat-panel detector
    Nose, Takayuki
    Masui, Koji
    Takenaka, Tadashi
    Yamazaki, Hideya
    Nakata, Katsuya
    Otani, Yuki
    Kumita, Shinichiro
    JOURNAL OF RADIATION RESEARCH, 2019, 60 (03) : 412 - 415
  • [42] Dosimetric perturbations at high-Z interfaces with high dose rate 192Ir source
    Zhang, Hualin
    Das, Indra J.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (07): : 782 - 790
  • [43] Calibration of new high dose rate 192Ir sources
    Stump, KE
    DeWerd, LA
    Micka, JA
    Anderson, DR
    MEDICAL PHYSICS, 2002, 29 (07) : 1483 - 1488
  • [44] AIR KERMA STANDARD AND MEASUREMENT COMPARISON ON SOURCE STRENGTH DETERMINATION FOR HIGH-DOSE RATE 192Ir BRACHYTHERAPY IN TAIWAN
    Lee, Jeng-Hung
    Su, Shi-Hwa
    Hsieh, Ming-Tsung
    Chen, Ing-Jane
    Liang, Ji-An
    Hsu, Shih-Ming
    RADIATION PROTECTION DOSIMETRY, 2011, 146 (1-3) : 100 - 102
  • [45] Tumor dose enhancement by nanoparticles during high dose rate 192 Ir brachytherapy
    Zabihzadeh, Mansour
    Arefian, Sahar
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2015, 11 (04) : 752 - 759
  • [46] Feasibility of aponeurectomy in combination with perioperative 192Ir high dose rate brachytherapy for Dupuytren's disease
    Ciernik, I. Frank
    Goldschmidt, Philipp
    Woesle, Markus
    Winter, Jochen
    STRAHLENTHERAPIE UND ONKOLOGIE, 2021, 197 (10) : 903 - 908
  • [47] In vivo thermoluminescence dosimetry dose verification of transperineal 192Ir high-dose-rate brachytherapy using CT-based planning for the treatment of prostate cancer
    Anagnostopoulos, G
    Baltas, D
    Geretschlaeger, A
    Martin, T
    Papagiannis, P
    Tselis, N
    Zamboglou, N
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (04): : 1183 - 1191
  • [48] Dose error from deviation of dwell time and source position for high dose-rate 192Ir in remote afterloading system
    Okamoto, Hiroyuki
    Aikawa, Ako
    Wakita, Akihisa
    Yoshio, Kotaro
    Murakami, Naoya
    Nakamura, Satoshi
    Hamada, Minoru
    Abe, Yoshihisa
    Itami, Jun
    JOURNAL OF RADIATION RESEARCH, 2014, 55 (04) : 780 - 787
  • [49] Clinical application of 192Ir High-Dose-Rate brachytherapy in metastatic lymph nodes of the neck
    Lu, Hongling
    Sun, Yunchuan
    Gao, Yan
    Xiao, Li
    Zhou, Jianxi
    Yin, Xiaoming
    Guo, Wei
    Fan, Kui
    DISCOVER ONCOLOGY, 2023, 14 (01)
  • [50] Feasibility of aponeurectomy in combination with perioperative 192Ir high dose rate brachytherapy for Dupuytren’s disease
    I. Frank Ciernik
    Philipp Goldschmidt
    Markus Wösle
    Jochen Winter
    Strahlentherapie und Onkologie, 2021, 197 : 903 - 908