AIR KERMA STANDARD AND MEASUREMENT COMPARISON ON SOURCE STRENGTH DETERMINATION FOR HIGH-DOSE RATE 192Ir BRACHYTHERAPY IN TAIWAN

被引:3
|
作者
Lee, Jeng-Hung [2 ]
Su, Shi-Hwa [2 ]
Hsieh, Ming-Tsung [2 ]
Chen, Ing-Jane [2 ]
Liang, Ji-An [1 ]
Hsu, Shih-Ming [1 ]
机构
[1] China Med Univ, Dept Biomed Imaging & Radiol Sci, Taichung 40402, Taiwan
[2] Inst Nucl Energy Res, Hlth Phys Div, Longtan Township 32546, Taoyuan County, Taiwan
关键词
D O I
10.1093/rpd/ncr121
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
This paper describes the establishment by the Institute of Nuclear Energy Research (INER, Taiwan) of the reference air kerma rate (RAKR) calibration standard for measurement of high-dose rate (HDR) Ir-192 brachytherapy source strength. A bilateral comparison has been made in the RAKR standards for HDR Ir-192 brachytherapy sources at the INER and Physikalisch-Technische Bundesanstalt (PTB, Germany) and the measurement difference was within the overall standard uncertainty and showed good agreement between the two calibration standard systems established at the INER and the PTB. Besides, INER also worked with 20 domestic hospitals to organise an on-site measurement comparison programme to explore the status of HDR Ir-192 brachytherapy source strength determination in Taiwan. The comparison results presented the ratios of RAKR with vendor values, as determined by INER and hospitals from the programme. The ratios fall in all cases within the +/-3 % guaranteed by the vendors for a coverage factor of k=2 or at 95 % confidence level.
引用
收藏
页码:100 / 102
页数:3
相关论文
共 50 条
  • [1] Determination of air-kerma strength for the 192Ir GammaMedplus iX pulsed-dose-rate brachytherapy source
    Riley, A. D.
    Pike, T. L.
    Micka, J. A.
    Fulkerson, R. K.
    DeWerd, L. A.
    MEDICAL PHYSICS, 2013, 40 (07)
  • [2] Determination of the reference air kerma rate for 192Ir brachytherapy sources and the related uncertainty
    van Dijk, E
    Kolkman-Deurloo, IKK
    Damen, PMG
    MEDICAL PHYSICS, 2004, 31 (10) : 2826 - 2833
  • [3] The Measurement of the Air-Kerma Rate in Air and a Solid Phantom with Ionization Chambers for a 192Ir HDR Brachytherapy Source
    Zeng, Jing
    Qu, Pengpeng
    Pang, Qingsong
    Wang, Ping
    CANCER MANAGEMENT AND RESEARCH, 2020, 12 : 10821 - 10828
  • [4] The IPEM code of practice for determination of the reference air kerma rate for HDR 192Ir brachytherapy sources based on the NPL air kerma standard
    Bidmead, A. M.
    Sander, T.
    Locks, S. M.
    Lee, C. D.
    Aird, E. G. A.
    Nutbrown, R. F.
    Flynn, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (11): : 3145 - 3159
  • [5] A technique for calibrating a high dose rate 192Ir brachytherapy source
    Fourie, O. L.
    Crabtree, T. G.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) : 85 - 92
  • [6] A technique for calibrating a high dose rate 192Ir brachytherapy source
    O. L. Fourie
    T. G. Crabtree
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 85 - 92
  • [7] Reference air kerma rate calibration system for high dose rate Ir-192 brachytherapy sources in Taiwan
    Chu, Wei-Han
    Yuan, Ming-Chen
    Lee, Jeng-Hung
    Lin, Yi-Chun
    RADIATION PHYSICS AND CHEMISTRY, 2017, 140 : 361 - 364
  • [8] A standard graphite calorimeter for dosimetry in brachytherapy with high dose rate 192Ir sources
    Guerra, A. S.
    Loreti, S.
    Pimpinella, M.
    Quini, M.
    D'Arienzo, M.
    Astefanoaei, I.
    Caporali, C.
    Bolzan, C.
    Pagliari, M.
    METROLOGIA, 2012, 49 (05) : S179 - S183
  • [9] Comparison of air-kerma strength determinations for HDR 192Ir sources
    Rasmussen, Brian E.
    Davis, Stephen D.
    Schmidt, Cal R.
    Micka, John A.
    DeWerd, Larry A.
    MEDICAL PHYSICS, 2011, 38 (12) : 6721 - 6729
  • [10] Dosimetric characteristics of the 192Ir high-dose-rate afterloading brachytherapy source
    Sadeghi, Mahdi
    Taghdiri, Fatemah
    Saidi, Pooneh
    JAPANESE JOURNAL OF RADIOLOGY, 2011, 29 (05) : 324 - 329