Determination of air-kerma strength for the 192Ir GammaMedplus iX pulsed-dose-rate brachytherapy source

被引:3
|
作者
Riley, A. D. [1 ]
Pike, T. L. [1 ]
Micka, J. A. [1 ]
Fulkerson, R. K. [1 ]
DeWerd, L. A. [1 ]
机构
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53705 USA
关键词
Pulsed-dose-rate brachytherapy; well chamber; calibration; DOSIMETRY PARAMETERS; AAPM; CALIBRATION;
D O I
10.1118/1.4812420
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Pulsed-dose-rate (PDR) brachytherapy was originally proposed to combine the therapeutic advantages of high-dose-rate (HDR) and low-dose-rate brachytherapy. Though uncommon in the United States, several facilities employ pulsed-dose-rate brachytherapy in Europe and Canada. Currently, there is no air-kerma strength standard for PDR brachytherapy Ir-192 sources traceable to the National Institute of Standards and Technology. Discrepancies in clinical measurements of the air-kerma strength of the PDR brachytherapy sources using HDR source-calibrated well chambers warrant further investigation. Methods: In this research, the air-kerma strength for an Ir-192 PDR brachytherapy source was compared with the University of Wisconsin Accredited Dosimetry Calibration Laboratory transfer standard well chambers, the seven-distance technique [B. E. Rasmussen et al., "The air-kerma strength standard for Ir-192 HDR sources," Med. Phys. 38, 6721-6729 (2011)], and the manufacturer's stated value. Radiochromic film and Monte Carlo techniques were also employed for comparison to the results of the measurements. Results: While the measurements using the seven-distance technique were within + 0.44% from the manufacturer's determination, there was a + 3.10% difference between the transfer standard well chamber measurements and the manufacturer's stated value. Results showed that the PDR brachytherapy source has geometric and thus radiological qualities that exhibit behaviors similar to a point source model in contrast to a conventional line source model. Conclusions: The resulting effect of the pointlike characteristics of the PDR brachytherapy source likely account for the differences observed between well chamber and in-air measurements. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Film dosimetry calibration method for pulsed-dose-rate brachytherapy with an 192Ir source
    Schwob, Nathan
    Orion, Itzhak
    MEDICAL PHYSICS, 2007, 34 (05) : 1678 - 1683
  • [2] The Measurement of the Air-Kerma Rate in Air and a Solid Phantom with Ionization Chambers for a 192Ir HDR Brachytherapy Source
    Zeng, Jing
    Qu, Pengpeng
    Pang, Qingsong
    Wang, Ping
    CANCER MANAGEMENT AND RESEARCH, 2020, 12 : 10821 - 10828
  • [3] AIR KERMA STANDARD AND MEASUREMENT COMPARISON ON SOURCE STRENGTH DETERMINATION FOR HIGH-DOSE RATE 192Ir BRACHYTHERAPY IN TAIWAN
    Lee, Jeng-Hung
    Su, Shi-Hwa
    Hsieh, Ming-Tsung
    Chen, Ing-Jane
    Liang, Ji-An
    Hsu, Shih-Ming
    RADIATION PROTECTION DOSIMETRY, 2011, 146 (1-3) : 100 - 102
  • [4] Spectra and air-kerma strength for encapsulated 192Ir sources
    Borg, J
    Rogers, DWO
    MEDICAL PHYSICS, 1999, 26 (11) : 2441 - 2444
  • [5] Comparison of air-kerma strength determinations for HDR 192Ir sources
    Rasmussen, Brian E.
    Davis, Stephen D.
    Schmidt, Cal R.
    Micka, John A.
    DeWerd, Larry A.
    MEDICAL PHYSICS, 2011, 38 (12) : 6721 - 6729
  • [6] Air-kerma strength determination of a 169Yb high dose rate brachytherapy source
    VanDamme, J. J.
    Culberson, W. S.
    DeWerd, L. A.
    Micka, J. A.
    MEDICAL PHYSICS, 2008, 35 (09) : 3935 - 3942
  • [7] Air-kerma strength determination of a 169Ytterbium high dose rate brachytherapy source
    VanDamme, J.
    DeWerd, L.
    Micka, J.
    Culberson, W.
    Davis, S.
    MEDICAL PHYSICS, 2006, 33 (06) : 1984 - 1984
  • [8] Determination of the reference air kerma rate for 192Ir brachytherapy sources and the related uncertainty
    van Dijk, E
    Kolkman-Deurloo, IKK
    Damen, PMG
    MEDICAL PHYSICS, 2004, 31 (10) : 2826 - 2833
  • [9] TLD DOSIMETRY OF A NEW 192Ir PULSED DOSE RATE BRACHYTHERAPY SOURCE
    Kabiri, F.
    Sheibani, S.
    Ghozzati, B.
    Beigi, H. Pour
    Javanshir, M.
    INTERNAL MEDICINE JOURNAL, 2011, 41 : 45 - 45
  • [10] Air-kerma strength determination of an HDR 192Ir source including a geometric sensitivity study of the seven-distance method
    Smith, Blake R.
    Micka, John A.
    Aima, Manik
    DeWerd, Larry A.
    Culberson, Wesley S.
    MEDICAL PHYSICS, 2017, 44 (01) : 311 - 320