Verification of high dose rate 192Ir source position during brachytherapy treatment

被引:16
|
作者
Batic, M. [1 ]
Burger, J. [2 ]
Cindro, V. [1 ]
Kramberger, G. [1 ]
Mandic, I. [1 ]
Mikuz, M. [1 ,3 ]
Studen, A. [1 ]
Zavrtanik, M. [1 ]
机构
[1] Jozef Stefan Inst, Expt Particle Phys Dept, Ljubljana, Slovenia
[2] Inst Oncol, Dept Radiophys, Ljubljana, Slovenia
[3] Univ Ljubljana, Dept Phys, Ljubljana 61000, Slovenia
关键词
Source localization; Brachytherapy; Silicon pixel detectors; Verification; CAMERA;
D O I
10.1016/j.nima.2009.09.122
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A system for in vivo tracking of 1 Ci Ir-192 source during brachytherapy treatment has been built using high resistivity silicon pad detectors as image sensors and knife-edge lead pinholes as collimators. The sensors consist of 256 pads arranged in 32 x 8 grid with pad size 1.4 x 1.4 mm(2) and 1 mm thickness. The sensors have two metal layers, enabling connection of readout electronics (VATAGP3_1 chips) at the edge of the detector. With source self-images obtained from a dual-pinhole system, location of the source can be reconstructed in three dimensions in real time, allowing on-line detection of deviations from planned treatment. The system was tested with 1 Ci Ir-192 clinical source in air and plexi-glass phantom. The movements of the source could be tracked in a field of view of approximately 20 x 20 x 20 cm(3) with absolute precision of about 5 mm. Positions of the source, relative to the first measured source position, could be mapped with precision of around 3 mm. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:206 / 208
页数:3
相关论文
共 50 条
  • [31] Optimization of deterministic transport parameters for the calculation of the dose distribution around a high dose-rate 192Ir brachytherapy source
    Gifford, Kent A.
    Price, Michael J.
    Horton, John L., Jr.
    Wareing, Todd A.
    Mourtada, Firas
    MEDICAL PHYSICS, 2008, 35 (06) : 2279 - 2285
  • [32] Monte Carlo dosimetry of the VariSource high dose rate 192Ir source
    Wang, RQ
    Sloboda, RS
    MEDICAL PHYSICS, 1998, 25 (04) : 415 - 423
  • [33] Dosimetric comparison of brachytherapy sources for high-dose-rate treatment of endometrial cancer: 192Ir, 60Co and an electronic brachytherapy source
    Mobit, Paul N.
    Nguyen, Alex
    Packianathan, Satyaseelan
    He, Rui
    Yang, Claus C.
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1059):
  • [34] A microDiamondfor determination of absorbed dose around high-dose-rate 192Ir brachytherapy sources
    Kaveckyte, V.
    Malusek, A.
    Benmaklouf, H.
    Carlsson, G. Alm
    Tedgren, A. Carlsson
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S407 - S407
  • [35] Dose perturbation due to catheter materials in high-dose-rate interstitial 192Ir brachytherapy
    Gaudreault, Mathieu
    Reniers, Brigitte
    Landry, Guillaume
    Verhaegen, Frank
    Beaulieu, Luc
    BRACHYTHERAPY, 2014, 13 (06) : 627 - 631
  • [36] A novel absorbed dose standard for the calibration of 192Ir sources used in high dose rate brachytherapy
    D'Arienzo, M.
    Guerra, A. S.
    Loreti, S.
    Caporali, C.
    Pimpinella, M.
    Bolzan, C.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S146 - S147
  • [37] Dosimetric characterization of a novel intracavitary mold applicator for 192Ir high dose rate endorectal brachytherapy treatment
    Poon, Emily
    Reniers, Brigitte
    Devic, Slobodan
    Vuong, Te
    Verhaegen, Frank
    MEDICAL PHYSICS, 2006, 33 (12) : 4515 - 4526
  • [38] Monte Carlo and experimental high dose rate 192Ir brachytherapy dosimetry with microDiamond detectors
    Rossi, Giulio
    Gainey, Mark
    Thomann, Benedikt
    Kollefrath, Michael
    Wuerfel, Jan
    Allgaier, Bernd
    Baltas, Dimos
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2019, 29 (03): : 272 - 281
  • [39] Characterization of microMOSFET detectors for in vivo dosimetry in high-dose-rate brachytherapy with 192Ir
    Ruiz-Arrebola, Samuel
    Fabregat-Borras, Rosa
    Rodriguez, Eduardo
    Fernandez-Montes, Manuel
    Perez-Macho, Mercedes
    Ferri, Maria
    Garcia, Ana
    Cardenal, Juan
    Pacheco, Maria T.
    Anchuelo, Javier
    Tornero-Lopez, Ana M.
    Prada, Pedro J.
    Guirado, Damian
    MEDICAL PHYSICS, 2020, 47 (05) : 2242 - 2253
  • [40] Transit dose of an Ir-192 high dose rate brachytherapy stepping source
    Wong, TPY
    Fernando, W
    Johnston, PN
    Bubb, IF
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02): : 323 - 331