Verification of high dose rate 192Ir source position during brachytherapy treatment

被引:16
|
作者
Batic, M. [1 ]
Burger, J. [2 ]
Cindro, V. [1 ]
Kramberger, G. [1 ]
Mandic, I. [1 ]
Mikuz, M. [1 ,3 ]
Studen, A. [1 ]
Zavrtanik, M. [1 ]
机构
[1] Jozef Stefan Inst, Expt Particle Phys Dept, Ljubljana, Slovenia
[2] Inst Oncol, Dept Radiophys, Ljubljana, Slovenia
[3] Univ Ljubljana, Dept Phys, Ljubljana 61000, Slovenia
关键词
Source localization; Brachytherapy; Silicon pixel detectors; Verification; CAMERA;
D O I
10.1016/j.nima.2009.09.122
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A system for in vivo tracking of 1 Ci Ir-192 source during brachytherapy treatment has been built using high resistivity silicon pad detectors as image sensors and knife-edge lead pinholes as collimators. The sensors consist of 256 pads arranged in 32 x 8 grid with pad size 1.4 x 1.4 mm(2) and 1 mm thickness. The sensors have two metal layers, enabling connection of readout electronics (VATAGP3_1 chips) at the edge of the detector. With source self-images obtained from a dual-pinhole system, location of the source can be reconstructed in three dimensions in real time, allowing on-line detection of deviations from planned treatment. The system was tested with 1 Ci Ir-192 clinical source in air and plexi-glass phantom. The movements of the source could be tracked in a field of view of approximately 20 x 20 x 20 cm(3) with absolute precision of about 5 mm. Positions of the source, relative to the first measured source position, could be mapped with precision of around 3 mm. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:206 / 208
页数:3
相关论文
共 50 条
  • [21] A Monte Carlo investigation of the dosimetric characteristics of the VariSource 192Ir high dose rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Baras, P
    Sakelliou, L
    Sandilos, P
    Dardoufas, K
    Vlachos, L
    MEDICAL PHYSICS, 1999, 26 (08) : 1498 - 1502
  • [22] Monte Carlo dosimetry of a new 192Ir pulsed dose rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Pantelis, E
    Papagiannis, P
    Sakelliou, L
    Kouwenhoven, E
    Baltas, D
    MEDICAL PHYSICS, 2003, 30 (01) : 9 - 16
  • [23] Dosimetric comparison of 60Co and 192Ir high dose rate source used in brachytherapy treatment of cervical cancer
    Shukla, Arvind Kumar
    Jangid, Pawan Kumar
    Rajpurohit, Vikram Singh
    Verma, Atul
    Dangayach, Suresh Kumar
    Gagrani, Vaibhav
    Rathore, Narendra Kumar
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (06) : 1212 - 1215
  • [24] Dosimetric investigation of a new high dose rate 192Ir brachytherapy source, IRAsource, by Monte Carlo method
    Rostami, Atefeh
    Hoseini, Mahdi
    Ghorbani, Mahdi
    Knaup, Courtney
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2020, 25 (01) : 139 - 145
  • [25] Comparison of calibration procedures for 192IR high-dose-rate brachytherapy sources
    Baltas, D
    Geramani, K
    Ioannidis, GT
    Hierholz, K
    Rogge, B
    Kolotas, C
    Müller-Sievers, K
    Milickovic, N
    Kober, B
    Zamboglou, N
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (03): : 653 - 661
  • [26] Film dosimetry calibration method for pulsed-dose-rate brachytherapy with an 192Ir source
    Schwob, Nathan
    Orion, Itzhak
    MEDICAL PHYSICS, 2007, 34 (05) : 1678 - 1683
  • [27] A standard graphite calorimeter for dosimetry in brachytherapy with high dose rate 192Ir sources
    Guerra, A. S.
    Loreti, S.
    Pimpinella, M.
    Quini, M.
    D'Arienzo, M.
    Astefanoaei, I.
    Caporali, C.
    Bolzan, C.
    Pagliari, M.
    METROLOGIA, 2012, 49 (05) : S179 - S183
  • [28] Determination of the dose rate around a HDR 192Ir brachytherapy source with the microDiamond and the microSilicon detector
    Rossi, Giulio
    Failing, Thomas
    Gainey, Mark
    Kollefrath, Michael
    Hensley, Frank
    Zink, Klemens
    Baltas, Dimos
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (04): : 463 - 478
  • [29] Monte Carlo dosimetry of the Buchler high dose rate 192Ir source
    Ballester, F
    Pérez-Calatayud, J
    Puchades, V
    Lluch, JL
    Serrano-Andrés, MA
    Limami, Y
    Lliso, F
    Casal, E
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (03): : N79 - N90
  • [30] High dose rate 192Ir source calibration: A single institution experience
    Abdullah, R.
    Abdullah, N. H.
    Mohamed, M.
    Idris, N. R. N.
    Yusoff, A. L.
    Chen, S. C.
    Zakaria, A.
    10TH INTERNATIONAL SEMINAR ON MEDICAL PHYSICS & 1ST AMDI INTERNATIONAL ONCOLOGY SYMPOSIUM, 2017, 851