Performance evaluation of a retrofit digital detector-based mammography system

被引:9
|
作者
Marshall, Nicholas W. [1 ]
van Ongeval, Chantal [1 ]
Bosmans, Hilde [1 ]
机构
[1] Univ Hosp Gasthuisberg, Dept Radiol, Herestr 49, B-3000 Leuven, Belgium
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2016年 / 32卷 / 02期
关键词
Mammography; Image quality; Detector characterization; IMAGE QUALITY MEASUREMENTS; BREAST SCREENING-PROGRAM; FILM MAMMOGRAPHY; NOISE; PART; OPTIMIZATION; PARAMETERS; VARIANCE; SOFTWARE; CONTRAST;
D O I
10.1016/j.ejmp.2016.01.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A retrofit flat panel detector was integrated with a GE DMR+ analog mammography system and characterized using detective quantum efficiency (DQE). Technical system performance was evaluated using the European Guidelines protocol, followed by a limited evaluation of clinical image quality for 20 cases using image quality criteria in the European Guidelines. Optimal anode/filter selections were established using signal difference-to-noise ratio measurements. Only small differences in peak DQE were seen between the three anode/filter settings, with an average value of 0.53. For poly(methyl methacrylate) (PMMA) thicknesses above 60 mm, the Rh/Rh setting was the optimal anode/filter setting. The system required a mean glandular dose of 0.54 mGy at 30 kV Rh/Rh to reach the Acceptable gold thickness limit for 0.1 mm details. Imaging performance of the retrofit unit with the GE DMR+ is notably better than of powder based computed radiography systems and is comparable to current flat panel FFDM systems. (C) 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:312 / 322
页数:11
相关论文
共 50 条
  • [31] Flat panel detector-based cone-beam volume CT breast imaging: detector evaluation
    Yu, Y
    Conover, D
    Ning, RL
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 711 - 719
  • [32] An Absolute Detector-based Spectral Radiance Source
    Brown, Steven W.
    Saunders, Robert D.
    Li, Zhigang
    Fein, Abra
    Barnes, Robert A.
    EARTH OBSERVING SYSTEMS XV, 2010, 7807
  • [33] A NEW DETECTOR-BASED SPECTRAL EMISSION SCALE
    ANDERSON, VE
    FOX, NP
    METROLOGIA, 1991, 28 (03) : 135 - 139
  • [34] Mixed Signal System Design Verification Accelerated With Detector-Based Diagnostic Method
    Dammers, D.
    Domingues, C.
    Schollaen, D.
    Vosskaemper, L. M.
    BMAS 2009: PROCEEDINGS OF THE 2009 IEEE INTERNATIONAL BEHAVIORAL MODELING AND SIMULATION WORKSHOP, 2009, : 66 - +
  • [35] Toward a detector-based universal phone recognizer
    Siniscalchi, Sabato Marco
    Svendsen, Torbiorn
    Lee, Chin-Hui
    2008 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH AND SIGNAL PROCESSING, VOLS 1-12, 2008, : 4261 - +
  • [36] Quantitative analysis and qualitative performance of fiat-panel detector for digital mammography
    Debrie, A
    Caron, M
    Savard, S
    Hansroul, M
    Wong, K
    Loustauneau, V
    Shukri, Z
    Martin, JP
    IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 25 - 31
  • [37] Diagnostic medical physics performance evaluation of a portable digital flat detector system
    Hussain, I.
    MEDICAL PHYSICS, 2006, 33 (06) : 2222 - 2222
  • [38] Contrast Enhance Digital Mammography Based On Photon-Counting Detector
    Lee, M.
    Lee, D.
    Kim, H.
    Kim, D.
    Kim, H.
    MEDICAL PHYSICS, 2018, 45 (06) : E517 - E517
  • [39] CCD-based detector for full-field digital mammography
    Stanton, M
    Phillips, W
    Stewart, A
    Smilowitz, L
    Williams, MB
    Simoni, P
    Ingersoll, C
    McCauley, T
    Qian, H
    MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 : 740 - 748
  • [40] A prototype CCD-based detector for whole breast digital mammography
    Rosen, D
    Phillips, W
    Stanton, M
    Xie, JH
    OMara, D
    Stewart, A
    PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 : 52 - 58