Performance evaluation of a retrofit digital detector-based mammography system

被引:9
|
作者
Marshall, Nicholas W. [1 ]
van Ongeval, Chantal [1 ]
Bosmans, Hilde [1 ]
机构
[1] Univ Hosp Gasthuisberg, Dept Radiol, Herestr 49, B-3000 Leuven, Belgium
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2016年 / 32卷 / 02期
关键词
Mammography; Image quality; Detector characterization; IMAGE QUALITY MEASUREMENTS; BREAST SCREENING-PROGRAM; FILM MAMMOGRAPHY; NOISE; PART; OPTIMIZATION; PARAMETERS; VARIANCE; SOFTWARE; CONTRAST;
D O I
10.1016/j.ejmp.2016.01.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A retrofit flat panel detector was integrated with a GE DMR+ analog mammography system and characterized using detective quantum efficiency (DQE). Technical system performance was evaluated using the European Guidelines protocol, followed by a limited evaluation of clinical image quality for 20 cases using image quality criteria in the European Guidelines. Optimal anode/filter selections were established using signal difference-to-noise ratio measurements. Only small differences in peak DQE were seen between the three anode/filter settings, with an average value of 0.53. For poly(methyl methacrylate) (PMMA) thicknesses above 60 mm, the Rh/Rh setting was the optimal anode/filter setting. The system required a mean glandular dose of 0.54 mGy at 30 kV Rh/Rh to reach the Acceptable gold thickness limit for 0.1 mm details. Imaging performance of the retrofit unit with the GE DMR+ is notably better than of powder based computed radiography systems and is comparable to current flat panel FFDM systems. (C) 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:312 / 322
页数:11
相关论文
共 50 条
  • [11] Flat panel detector-based cone beam CT imaging: Preliminary system evaluation
    Ning, RL
    Conover, D
    Yu, Y
    Lu, XH
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 616 - 620
  • [12] Flat panel detector-based cone beam CT for dynamic imaging: System evaluation
    Ning, Ruola
    Conover, David
    Yu, Yong
    Zhang, Yan
    Cai, Weixing
    Yang, Dong
    Lu, Xianghua
    MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
  • [13] Flat panel detector-based volume tomographic angiography imaging: Detector evaluation
    Ning, R
    Zhang, DH
    Chen, B
    Conover, D
    Yu, RF
    DEVELOPMENTS IN X-RAY TOMOGRAPHY II, 1999, 3772 : 167 - 178
  • [14] Digital mammography system design and performance
    Tesic, M
    Piccaro, MF
    CAR '96: COMPUTER ASSISTED RADIOLOGY, 1996, 1124 : 47 - 52
  • [15] Flat panel detector-based cone beam CT lung imaging: preliminary system evaluation
    Ning, R
    Conover, D
    Yu, Y
    Cai, WX
    Lu, XH
    MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2005, 5745 : 147 - 158
  • [16] Flat panel detector-based cone beam volume CT imaging: detector evaluation
    Ning, R
    Tang, XY
    Yu, RF
    Conover, D
    Zhang, DH
    MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 : 192 - 203
  • [17] A CCD based digital detector for whole-breast digital mammography
    Smilowitz, L
    Rosen, D
    Qian, H
    Phillips, W
    Stanton, M
    Stewart, A
    Mangiafico, P
    Simoni, P
    Williams, M
    DIGITAL MAMMOGRAPHY, 1998, 13 : 31 - 34
  • [18] Performance evaluation of six digital mammography systems
    Perez, Alessandra Maia Marques Martinez
    Lopes, Laura Alcantara Silva
    Caron, Renato Franca
    Oliveira, Bruno Beraldo
    Poletti, Martin Eduardo
    RADIATION PHYSICS AND CHEMISTRY, 2024, 218
  • [20] Evaluation of a digital X-ray detector based on a phosphor-coated CCD for mammography
    Taibi, A.
    Del Guerra, A.
    Gambaccini, M.
    Marziani, M.
    Tuffanelli, A.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997, 392 (1-3): : 210 - 213