Bipolar Radiofrequency Ablation: Development of a New Expandable Device

被引:4
|
作者
Ito, Nobutake [1 ,2 ]
Pfeffer, Jochen [1 ]
Isfort, Peter [1 ,2 ]
Penzkofer, Tobias [1 ,2 ]
Kuhl, Christiane K. [1 ]
Mahnken, Andreas H. [3 ]
Schmitz-Rode, Thomas [2 ]
Bruners, Philipp [1 ]
机构
[1] Rhein Westfal TH Aachen, Dept Diagnost Radiol, Aachen, Germany
[2] Rhein Westfal TH Aachen, Helmholtz Inst Biomed Engn, Aachen, Germany
[3] Univ Marburg, Dept Radiol, Marburg, Germany
关键词
Liver; Minimally invasive therapy; Radiofrequency ablation; INTERNALLY COOLED ELECTRODES; ENHANCED HEPATIC BIPOLAR; VIVO BOVINE LIVER; EFFICIENCY; MONOPOLAR; TUMORS; COMPLICATIONS; APPLICATORS; COAGULATION; PRINCIPLES;
D O I
10.1007/s00270-013-0703-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To test the performance of an expandable bipolar probe as a simple technical solution for extending the coagulation volume. On the basis of a commercially available monopolar radiofrequency (RF) probe (LeVeen), an expandable bipolar RF probe was developed by integrating a second electrode into the probe shaft. The influence of length on the second electrode, and the distance between both electrodes and generator output was investigated by performing ten ablations for each condition on a freshly excised bovine liver. Macroscopically quantified coagulation volumes, lesion shape characteristics, and procedure durations were recorded. Results of the prototype featuring the optimal configuration were compared to the original LeVeen probe and commonly used bipolar RF probe (CelonLabPower). Extension of the shaft electrode length, increasing distance between the shaft electrode and the tip electrode, and reduction of generator output resulted in increasing coagulation volumes. The coagulation volumes the prototype generated were significantly smaller and more elliptically shaped than the monopolar probe (9.4 +/- A 1.5 cm(3) vs. 12.1 +/- A 1.6 cm(3)), but were larger than the commercially available bipolar RF probe (vs. 7.3 +/- A 0.5). The procedure duration of the prototype was comparable to the monopolar probe (467 +/- A 31 s vs. 464 +/- A 17 s) and shorter than the bipolar probe (vs. 2009 +/- A 444 s). In comparison to the commercially available bipolar system, the developed prototype exhibited favorable results. The first benchmark testing of the developed bipolar prototype had promising results. However, further optimization of the applicator design and ablation protocol is needed to enlarge the achievable coagulation volume.
引用
收藏
页码:770 / 776
页数:7
相关论文
共 50 条
  • [31] Thin-layer ablation of human esophageal epithelium using a bipolar radiofrequency balloon device
    Dunkin, BJ
    Martinez, J
    Bejarano, PA
    Smith, CD
    Chang, K
    Livingstone, AS
    Melvin, WS
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2006, 20 (01): : 125 - 130
  • [32] Performance of a Novel Dual-Electrode Bipolar Radiofrequency Ablation Device A Chronic Porcine Study
    Voeller, Rochus K.
    Zierer, Andreas
    Schuessler, Richard B.
    Damiano, Ralph J., Jr.
    INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY, 2011, 6 (01) : 17 - 22
  • [33] Expandable radiofrequency electrodes for ablation of centrally placed hepatocellular carcinoma
    Ahmad, Medhat Ibraheem Mohammad
    EGYPTIAN JOURNAL OF RADIOLOGY AND NUCLEAR MEDICINE, 2018, 49 (01): : 92 - 98
  • [34] Evaluation of bipolar radiofrequency ablation for occluded self-expandable metal stents in the bile duct: in vivo and in vitro study
    Yoon, Won Jae
    Kim, Yong-Tae
    Daglilar, Ebubekir S.
    Mino-Kenudson, Mari
    Brugge, William R.
    ENDOSCOPY, 2015, 47 (12) : 1167 - 1170
  • [35] Predictors of a Successful Bipolar Radiofrequency Endometrial Ablation
    Eisele, Lilly
    Kochli, Lea
    Stadele, Patricia
    Welter, JoEllen
    Fehr-Kuhn, Maja
    Fehr, Mathias K.
    GEBURTSHILFE UND FRAUENHEILKUNDE, 2019, 79 (03) : 286 - 292
  • [36] An update to an update on radiofrequency bipolar catheter ablation
    Futyma, Piotr
    Kulakowski, Piotr
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (05) : 1238 - 1239
  • [37] Histopathology of Bipolar Radiofrequency Ablation in the Human Atrium
    Kasirajan, Vigneshwar
    Sayeed, Sadia
    Filler, Elizabeth
    Knarik, Arkun
    Koneru, Jayanthi N.
    Ellenbogen, Kenneth A.
    ANNALS OF THORACIC SURGERY, 2016, 101 (02): : 638 - 643
  • [38] Bipolar Radiofrequency Ablation of Typical Atrial Flutter
    Futyma, Piotr
    Futyma, Marian
    Maciolek, Marcin
    Kulakowski, Piotr
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2016, 27 (07) : 874 - 875
  • [39] Complete Right Atrial Ablation With Bipolar Radiofrequency
    Benussi, Stefano
    Galanti, Andrea
    Nascimbene, Simona
    Fumero, Andrea
    Dorigo, Enrica
    Zerbi, Valerio
    Alfieri, Ottavio
    ANNALS OF THORACIC SURGERY, 2009, 87 (05): : 1573 - 1576
  • [40] Complete left atrial ablation with bipolar radiofrequency
    Benussi, Stefano
    Nascimbene, Simona
    Galanti, Andrea
    Fumero, Andrea
    Dorigo, Enrica
    Zerbi, Valerio
    Cioni, Micaela
    Alfieri, Ottavio
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2008, 33 (04) : 590 - 595