Bipolar Radiofrequency Ablation: Development of a New Expandable Device

被引:4
|
作者
Ito, Nobutake [1 ,2 ]
Pfeffer, Jochen [1 ]
Isfort, Peter [1 ,2 ]
Penzkofer, Tobias [1 ,2 ]
Kuhl, Christiane K. [1 ]
Mahnken, Andreas H. [3 ]
Schmitz-Rode, Thomas [2 ]
Bruners, Philipp [1 ]
机构
[1] Rhein Westfal TH Aachen, Dept Diagnost Radiol, Aachen, Germany
[2] Rhein Westfal TH Aachen, Helmholtz Inst Biomed Engn, Aachen, Germany
[3] Univ Marburg, Dept Radiol, Marburg, Germany
关键词
Liver; Minimally invasive therapy; Radiofrequency ablation; INTERNALLY COOLED ELECTRODES; ENHANCED HEPATIC BIPOLAR; VIVO BOVINE LIVER; EFFICIENCY; MONOPOLAR; TUMORS; COMPLICATIONS; APPLICATORS; COAGULATION; PRINCIPLES;
D O I
10.1007/s00270-013-0703-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To test the performance of an expandable bipolar probe as a simple technical solution for extending the coagulation volume. On the basis of a commercially available monopolar radiofrequency (RF) probe (LeVeen), an expandable bipolar RF probe was developed by integrating a second electrode into the probe shaft. The influence of length on the second electrode, and the distance between both electrodes and generator output was investigated by performing ten ablations for each condition on a freshly excised bovine liver. Macroscopically quantified coagulation volumes, lesion shape characteristics, and procedure durations were recorded. Results of the prototype featuring the optimal configuration were compared to the original LeVeen probe and commonly used bipolar RF probe (CelonLabPower). Extension of the shaft electrode length, increasing distance between the shaft electrode and the tip electrode, and reduction of generator output resulted in increasing coagulation volumes. The coagulation volumes the prototype generated were significantly smaller and more elliptically shaped than the monopolar probe (9.4 +/- A 1.5 cm(3) vs. 12.1 +/- A 1.6 cm(3)), but were larger than the commercially available bipolar RF probe (vs. 7.3 +/- A 0.5). The procedure duration of the prototype was comparable to the monopolar probe (467 +/- A 31 s vs. 464 +/- A 17 s) and shorter than the bipolar probe (vs. 2009 +/- A 444 s). In comparison to the commercially available bipolar system, the developed prototype exhibited favorable results. The first benchmark testing of the developed bipolar prototype had promising results. However, further optimization of the applicator design and ablation protocol is needed to enlarge the achievable coagulation volume.
引用
收藏
页码:770 / 776
页数:7
相关论文
共 50 条
  • [21] Left Atrial Ablation for Atrial Fibrillation Creating the "Box Lesion" with a Bipolar Radiofrequency Device
    Sternik, Leonid
    Schaff, Hartzel V.
    Luria, David
    Glikson, Michael
    Kogan, Alexander
    Malachy, Ateret
    First, Maya
    Raanani, Ehud
    TEXAS HEART INSTITUTE JOURNAL, 2011, 38 (02) : 127 - 131
  • [22] Localised radiofrequency ablation of small breast tumours: In vivo evaluation of a bipolar application device
    Waaijer, L.
    Kreb, D. L.
    Gallardo, M. A. Fernandez
    Koelemij, R.
    Van Diest, P. J.
    Van Hillegersberg, R.
    Witkamp, A. J.
    EUROPEAN JOURNAL OF CANCER, 2014, 50 : S126 - S126
  • [23] New ablation procedure for a radiofrequency liver tissue coagulation system using an expandable needle
    Hirakawa, Miharu
    Ikeda, Kenji
    Kawamura, Yusuke
    Kobayashi, Masahiro
    Hosaka, Tetsuya
    Yatsuji, Hiromi
    Sezaki, Hitomi
    Akuta, Norio
    Suzuki, Fumitaka
    Suzuki, Yoshiyuki
    Saitoh, Satoshi
    Arase, Yasuji
    Kumada, Hiromitsu
    LIVER INTERNATIONAL, 2008, 28 (02) : 214 - 219
  • [24] Incomplete transmural ablation caused by bipolar radiofrequency ablation devices
    Kato, Wataru
    Usui, Akihiko
    Oshima, Hideki
    Akita, Toshiaki
    Ueda, Yuichi
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 137 (01): : 251 - 252
  • [25] Liver resection with a new multiprobe bipolar radiofrequency device - Invited critique
    Kemeny, M. Margaret
    ARCHIVES OF SURGERY, 2008, 143 (04) : 401 - 401
  • [26] RADIOFREQUENCY ASSISTED HEPATIC RESECTION: A NEW TECHNIQUE USING A BIPOLAR DEVICE
    Elgendi, A.
    Vavra, P.
    ElKayal, E.
    Habib, N.
    Baki, N. Abdel
    JOURNAL OF HEPATOLOGY, 2010, 52 : S189 - S189
  • [27] Clinical evaluation of a new fractionated bipolar radiofrequency device for the treatment of photoaging
    Gold, Michael
    Foster, Amy
    Biron, Julie
    JOURNAL OF THE AMERICAN ACADEMY OF DERMATOLOGY, 2011, 64 (02) : AB171 - AB171
  • [28] Treatment of Metastatic Spinal Lesions with a Navigational Bipolar Radiofrequency Ablation Device: A Multicenter Retrospective Study
    Anchala, Praveen R.
    Irving, Winston D.
    Hillen, Travis J.
    Friedman, Michael V.
    Georgy, Bassem A.
    Coldwell, Douglass M.
    Tran, Nam D.
    Vrionis, Frank D.
    Brook, Allan
    Jennings, Jack W.
    PAIN PHYSICIAN, 2014, 17 (04) : 317 - 327
  • [29] Thin-layer ablation of human esophageal epithelium using a bipolar radiofrequency balloon device
    B. J. Dunkin
    J. Martinez
    P. A. Bejarano
    C. D. Smith
    K. Chang
    A. S. Livingstone
    W. S. Melvin
    Surgical Endoscopy And Other Interventional Techniques, 2006, 20 : 125 - 130
  • [30] Performance of a Novel Bipolar/Monopolar Radiofrequency Ablation Device on the Beating Heart in an Acute Porcine Model
    Saint, Lindsey L.
    Lawrance, Christopher P.
    Okada, Shoichi
    Kazui, Toshinobu
    Robertson, Jason O.
    Schuessler, Richard B.
    Damiano, Ralph J., Jr.
    INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY, 2013, 8 (04) : 276 - 283