Bipolar Radiofrequency Ablation: Development of a New Expandable Device

被引:4
|
作者
Ito, Nobutake [1 ,2 ]
Pfeffer, Jochen [1 ]
Isfort, Peter [1 ,2 ]
Penzkofer, Tobias [1 ,2 ]
Kuhl, Christiane K. [1 ]
Mahnken, Andreas H. [3 ]
Schmitz-Rode, Thomas [2 ]
Bruners, Philipp [1 ]
机构
[1] Rhein Westfal TH Aachen, Dept Diagnost Radiol, Aachen, Germany
[2] Rhein Westfal TH Aachen, Helmholtz Inst Biomed Engn, Aachen, Germany
[3] Univ Marburg, Dept Radiol, Marburg, Germany
关键词
Liver; Minimally invasive therapy; Radiofrequency ablation; INTERNALLY COOLED ELECTRODES; ENHANCED HEPATIC BIPOLAR; VIVO BOVINE LIVER; EFFICIENCY; MONOPOLAR; TUMORS; COMPLICATIONS; APPLICATORS; COAGULATION; PRINCIPLES;
D O I
10.1007/s00270-013-0703-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To test the performance of an expandable bipolar probe as a simple technical solution for extending the coagulation volume. On the basis of a commercially available monopolar radiofrequency (RF) probe (LeVeen), an expandable bipolar RF probe was developed by integrating a second electrode into the probe shaft. The influence of length on the second electrode, and the distance between both electrodes and generator output was investigated by performing ten ablations for each condition on a freshly excised bovine liver. Macroscopically quantified coagulation volumes, lesion shape characteristics, and procedure durations were recorded. Results of the prototype featuring the optimal configuration were compared to the original LeVeen probe and commonly used bipolar RF probe (CelonLabPower). Extension of the shaft electrode length, increasing distance between the shaft electrode and the tip electrode, and reduction of generator output resulted in increasing coagulation volumes. The coagulation volumes the prototype generated were significantly smaller and more elliptically shaped than the monopolar probe (9.4 +/- A 1.5 cm(3) vs. 12.1 +/- A 1.6 cm(3)), but were larger than the commercially available bipolar RF probe (vs. 7.3 +/- A 0.5). The procedure duration of the prototype was comparable to the monopolar probe (467 +/- A 31 s vs. 464 +/- A 17 s) and shorter than the bipolar probe (vs. 2009 +/- A 444 s). In comparison to the commercially available bipolar system, the developed prototype exhibited favorable results. The first benchmark testing of the developed bipolar prototype had promising results. However, further optimization of the applicator design and ablation protocol is needed to enlarge the achievable coagulation volume.
引用
收藏
页码:770 / 776
页数:7
相关论文
共 50 条
  • [11] Pilot study for a new bipolar radiofrequency ablation/aspirator device in the management of primary and secondary liver cancers
    Zacharoulis, Dimitris
    Khorsandi, Shirin E.
    Vavra, Petr
    Dostalik, Jan
    Navarra, Giuseppe
    Nicholls, Joanna P.
    Jiao, Long R.
    Habib, Nagy A.
    LIVER INTERNATIONAL, 2009, 29 (06) : 824 - 830
  • [12] Liver resection with a new multiprobe bipolar radiofrequency device
    Ayav, Ahmet
    Jiao, Long
    Dickinson, Robert
    Nicholls, Joanna
    Milicevic, Miroslav
    Pellicci, Ricardo
    Bachellier, Philippe
    Habib, Nagy
    ARCHIVES OF SURGERY, 2008, 143 (04) : 396 - 401
  • [13] The Same is Not the Same: Device Effect during Bipolar Radiofrequency Ablation of Atrial Fibrillation
    Niemann, Bernd
    Dominik, Elisabeth
    Rohrbach, Susanne
    Grieshaber, Philippe
    Roth, Peter
    Boening, Andreas
    THORACIC AND CARDIOVASCULAR SURGEON, 2021, 69 (02): : 124 - 132
  • [14] Initial experience using a bipolar radiofrequency ablation device for hemostasis during thyroidectomy
    Baldwin, Keith
    Haniff, Michelle
    Somasundar, Ponnandai
    HEAD AND NECK-JOURNAL FOR THE SCIENCES AND SPECIALTIES OF THE HEAD AND NECK, 2013, 35 (01): : 118 - 122
  • [16] Endovenous radiofrequency ablation using a new bipolar electrode in a canine model: a new endovenous radiofrequency electrode
    Hwang, Jin Ho
    Park, Sang Woo
    Min, Jeeyoung
    Yang, Woo Young
    Kwon, Yong Wonn
    Hwang, Jae Joon
    Kim, Jun Seok
    Lee, Song Am
    Chee, Hyun Keun
    ANNALS OF SURGICAL TREATMENT AND RESEARCH, 2023, 104 (03) : 164 - 169
  • [17] Radiofrequency Ablation of Osteoid Osteoma: Initial Experience with a New Monopolar Ablation Device
    Andreas H. Mahnken
    Philipp Bruners
    Heide Delbrück
    Rolf W. Günther
    CardioVascular and Interventional Radiology, 2011, 34 : 579 - 584
  • [18] Technical Development of a New Semispherical Radiofrequency Bipolar Device (RONJA): Ex Vivo and In Vivo Studies
    Vavra, Petr
    Penhaker, Marek
    Grepl, Jan
    Jurcikova, Jana
    Palecek, Jiri
    Crha, Michal
    Nowakova, Jana
    Hasal, Martin
    Skrobankova, Martina
    Ostruszka, Petr
    Ihnat, Peter
    Delongova, Patricie
    Salounova, Dana
    Habib, Nagy A.
    Zonca, Pavel
    BIOMED RESEARCH INTERNATIONAL, 2014, 2014
  • [19] Radiofrequency Ablation of Osteoid Osteoma: Initial Experience with a New Monopolar Ablation Device
    Mahnken, Andreas H.
    Bruners, Philipp
    Delbrueck, Heide
    Guenther, Rolf W.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 34 (03) : 579 - 584
  • [20] InLine bipolar radiofrequency ablation device-assisted partial nephrectomy in a porcine model
    Yao, Peng
    Gunasegaram, Aravin
    Ladd, Leigh A.
    Morris, David L.
    ANZ JOURNAL OF SURGERY, 2008, 78 (07) : 564 - 567