Physical and biological pretreatment quality assurance of the head and neck cancer plan with the volumetric modulated arc therapy

被引:0
|
作者
So-Hyun Park
Dong-Soo Lee
Yun-Hee Lee
Seu-Ran Lee
Min-Ju Kim
Tae-Suk Suh
机构
[1] Uijeongbu St. Mary’s hospital,Department of Radiation Oncology
[2] Seoul St. Mary’s hospital,Department of Radiation Oncology
[3] The Catholic University of Korea College of Medicine,Department of Biomedical Engineering
来源
关键词
Quality assurance; Gamma passing rates; False negative; EUD; TCP/NTCP;
D O I
暂无
中图分类号
学科分类号
摘要
The aim of this work is to demonstrate both the physical and the biological quality assurance (QA) aspects as pretreatment QA of the head and neck (H&N) cancer plan for the volumetric modulated arc therapy (VMAT). Ten H&N plans were studied. The COMPASS® dosimetry analysis system and the tumor control probability (TCP) and the normal tissue complication probability (NTCP) calculation free program were used as the respective measurement and calculation tools. The reliability of these tools was verified by a benchmark study in accordance with the TG-166 report. For the physical component of QA, the gamma passing rates and the false negative cases between the calculated and the measured data were evaluated. The biological component of QA was performed based on the equivalent uniform dose (EUD), TCP and NTCP values. The evaluation was performed for the planning target volumes (PTVs) and the organs at risks (OARs), including the eyes, the lens, the parotid glands, the esophagus, the spinal cord, and the brainstem. All cases had gamma passing rates above 95% at an acceptance tolerance level with the 3%/3 mm criteria. In addition, the false negative instances were presented for the PTVs and OARs. The gamma passing rates exhibited a weak correlation with false negative cases. For the biological QA, the physical dose errors affect the EUD and the TCP for the PTVs, but no linear correlation existed between them. The EUD and NTCP for the OARs were shown the random differences that could not be attributed to the dose errors from the physical QA. The differences in the EUD and NTCP between the calculated and the measured results were mainly demonstrated for the parotid glands. This study describes the importance and the necessity of improved QA to accompany both the physical and the biological aspects for accurate radiation treatment.
引用
收藏
页码:946 / 955
页数:9
相关论文
共 50 条
  • [11] Plan Robustness Study of Volumetric-Modulated Arc Therapy Vs. Intensity-Modulated Radiation Therapy for Head and Neck Cancer
    Liu, W.
    Patel, S.
    Shen, J.
    Harrington, D.
    Stoker, J.
    Ding, X.
    Hu, Y.
    Wong, W.
    Halyard, M.
    Schild, S.
    Ezzell, G.
    Bues, M.
    MEDICAL PHYSICS, 2015, 42 (06) : 3478 - 3478
  • [12] Quality assurance of dynamic parameters in volumetric modulated arc therapy
    Manikandan, A.
    Sarkar, B.
    Holla, R.
    Vivek, T. R.
    Sujatha, N.
    BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1015): : 1002 - 1010
  • [13] Evidence based quality assurance of volumetric modulated arc therapy
    Tuemmler, H.
    Bellmann, C.
    Schilling, P.
    Schreiber, A.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S117 - S117
  • [14] Exploratory study of the association of volumetric modulated arc therapy (VMAT) plan robustness with local failure in head and neck cancer
    Liu, Wei
    Patel, Samir H.
    Harrington, Daniel P.
    Hu, Yanle
    Ding, Xiaoning
    Shen, Jiajian
    Halyard, Michele Y.
    Schild, Steven E.
    Wong, William W.
    Ezzell, Gary E.
    Bues, Martin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (04): : 76 - 83
  • [15] Quality assurance of volumetric modulated arc therapy for a prostate cancer patient with a hip prosthesis
    Ikeda, T.
    Moriwaki, H.
    Sakumi, A.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S595 - S595
  • [16] Early clinical experience with volumetric modulated arc therapy in head and neck cancer patients
    Scorsetti, Marta
    Fogliata, Antonella
    Castiglioni, Simona
    Bressi, Caterina
    Bignardi, Mario
    Navarria, Piera
    Mancosu, Pietro
    Cozzi, Luca
    Pentimalli, Sara
    Alongi, Filippo
    Santoro, Armando
    RADIATION ONCOLOGY, 2010, 5
  • [17] Impact of shoulder deformation on volumetric modulated arc therapy doses for head and neck cancer
    Tachiban, Hidenobu
    Motegi, Kana
    Moriya, Shunsuke
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 53 : 118 - 128
  • [18] Early clinical experience with volumetric modulated arc therapy in head and neck cancer patients
    Marta Scorsetti
    Antonella Fogliata
    Simona Castiglioni
    Caterina Bressi
    Mario Bignardi
    Piera Navarria
    Pietro Mancosu
    Luca Cozzi
    Sara Pentimalli
    Filippo Alongi
    Armando Santoro
    Radiation Oncology, 5
  • [19] Automatic interactive optimization for volumetric modulated arc therapy planning of head and neck cancer
    Tol, J.
    Dahele, M.
    Doornaert, P.
    Slotman, B. J.
    Verbakel, W. F. A. R.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S35 - S35
  • [20] Predicting the complexity of head-and-neck volumetric-modulated arc therapy planning using a radiation therapy planning quality assurance software
    Sasaki, Motoharu
    Nakaguchi, Yuji
    Kamomae, Takeshi
    Ueda, Shoji
    Endo, Yuto
    Sato, Daisuke
    Ikushima, Hitoshi
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2022, 27 (06) : 963 - 972