Physical and biological pretreatment quality assurance of the head and neck cancer plan with the volumetric modulated arc therapy

被引:0
|
作者
So-Hyun Park
Dong-Soo Lee
Yun-Hee Lee
Seu-Ran Lee
Min-Ju Kim
Tae-Suk Suh
机构
[1] Uijeongbu St. Mary’s hospital,Department of Radiation Oncology
[2] Seoul St. Mary’s hospital,Department of Radiation Oncology
[3] The Catholic University of Korea College of Medicine,Department of Biomedical Engineering
来源
关键词
Quality assurance; Gamma passing rates; False negative; EUD; TCP/NTCP;
D O I
暂无
中图分类号
学科分类号
摘要
The aim of this work is to demonstrate both the physical and the biological quality assurance (QA) aspects as pretreatment QA of the head and neck (H&N) cancer plan for the volumetric modulated arc therapy (VMAT). Ten H&N plans were studied. The COMPASS® dosimetry analysis system and the tumor control probability (TCP) and the normal tissue complication probability (NTCP) calculation free program were used as the respective measurement and calculation tools. The reliability of these tools was verified by a benchmark study in accordance with the TG-166 report. For the physical component of QA, the gamma passing rates and the false negative cases between the calculated and the measured data were evaluated. The biological component of QA was performed based on the equivalent uniform dose (EUD), TCP and NTCP values. The evaluation was performed for the planning target volumes (PTVs) and the organs at risks (OARs), including the eyes, the lens, the parotid glands, the esophagus, the spinal cord, and the brainstem. All cases had gamma passing rates above 95% at an acceptance tolerance level with the 3%/3 mm criteria. In addition, the false negative instances were presented for the PTVs and OARs. The gamma passing rates exhibited a weak correlation with false negative cases. For the biological QA, the physical dose errors affect the EUD and the TCP for the PTVs, but no linear correlation existed between them. The EUD and NTCP for the OARs were shown the random differences that could not be attributed to the dose errors from the physical QA. The differences in the EUD and NTCP between the calculated and the measured results were mainly demonstrated for the parotid glands. This study describes the importance and the necessity of improved QA to accompany both the physical and the biological aspects for accurate radiation treatment.
引用
收藏
页码:946 / 955
页数:9
相关论文
共 50 条
  • [31] Improving plan quality for prostate volumetric-modulated arc therapy
    Wright, Katrina
    Ferrari-Anderson, Janet
    Barry, Tamara
    Bernard, Anne
    Brown, Elizabeth
    Lehman, Margot
    Pryor, David
    MEDICAL DOSIMETRY, 2017, 42 (04) : 348 - 356
  • [32] Limitations of Optimization Algorithm With Volumetric Modulated Arc Technique for Head and Neck Cancer Radiation Therapy
    Xue, J.
    Wu, P.
    Kryger, M.
    Wang, W.
    Cheng, P.
    Das, I. J.
    Hu, K. S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E741 - E741
  • [33] Evaluation of volumetric modulated arc therapy (VMAT) with Oncentra MasterPlan® for the treatment of head and neck cancer
    Alvarez-Moret, Judith
    Pohl, Fabian
    Koelbl, Oliver
    Dobler, Barbara
    RADIATION ONCOLOGY, 2010, 5
  • [34] Robustness quantification methods comparison in volumetric modulated arc therapy to treat head and neck cancer
    Liu, Wei
    Patel, Samir H.
    Shen, Jiajian
    Hu, Yanle
    Harrington, Daniel P.
    Ding, Xiaoning
    Halyard, Michele Y.
    Schild, Steven E.
    Wong, William W.
    Ezzell, Gary A.
    Bues, Martin
    PRACTICAL RADIATION ONCOLOGY, 2016, 6 (06) : E269 - E275
  • [35] Evaluation of volumetric modulated arc therapy (VMAT) with Oncentra MasterPlan® for the treatment of head and neck cancer
    Judith Alvarez-Moret
    Fabian Pohl
    Oliver Koelbl
    Barbara Dobler
    Radiation Oncology, 5
  • [36] Impact of machines on plan quality: volumetric modulated arc therapy and intensity modulated radiation therapy
    Clemente, S.
    Cozzolino, M.
    Oliviero, C.
    Fiorentino, A.
    Chiumento, C.
    Fusco, V.
    CLINICAL & TRANSLATIONAL ONCOLOGY, 2014, 16 (02): : 141 - 146
  • [37] Impact of machines on plan quality: volumetric modulated arc therapy and intensity modulated radiation therapy
    S. Clemente
    M. Cozzolino
    C. Oliviero
    A. Fiorentino
    C. Chiumento
    V. Fusco
    Clinical and Translational Oncology, 2014, 16 : 141 - 146
  • [38] Assessment of combined use of ArcCheck® detector and portal dosimetry for delivery quality assurance of head and neck and prostate volumetric-modulated arc therapy
    Moliner, Gilles
    Sorro, Lise
    Verstraet, Rodolfe
    Daviau, Paul Alexandre
    Casas, Melanie
    Piron, Berengere
    Dubois, Karine
    Debrigode, Charles
    Barrau, Corinne
    Bons, Francoise
    Greffier, Joel
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (06): : 133 - 139
  • [39] A longitudinal evaluation of improvements in treatment plan quality for lung cancer with volumetric modulated arc therapy
    Xia, Wenlong
    Liu, Zhiqiang
    Yan, Lingling
    Han, Fei
    Hu, Zhihui
    Tian, Yuan
    Cui, Weijie
    Ren, Wenting
    Guo, Chenlei
    Miao, Junjie
    Dai, Jianrong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (06): : 33 - 43
  • [40] A novel method for routine quality assurance of volumetric-modulated arc therapy
    Wang, Qingxin
    Dai, Jianrong
    Zhang, Ke
    MEDICAL PHYSICS, 2013, 40 (10)