The comparison of ion exchange-high performance liquid chromatography (IE-HPLC) and capillary electrophoresis (CE) for HbA1C measurement

被引:0
|
作者
Sutrisnani, Catur Suci [1 ]
Darmawan, Edwin [2 ]
Widijanti, Anik [2 ]
Soehita, Sidarti [3 ]
机构
[1] Airlangga Univ, Clin Pathol Dept, Fac Med, Surabaya, Indonesia
[2] Brawijaya Univ, Dept Clin Pathol, Fac Med, Dr Saiful Anwar Hosp, Malang, Indonesia
[3] Airlangga Univ, Dept Clin Pathol, Fac Med, Dr Soetomo Hosp, Surabaya, Indonesia
关键词
HbA1c; measurement; HPLC; CE; HEMOGLOBIN A(1C) MEASUREMENT; PROGRAM; GOALS;
D O I
10.15562/bmj.v8i2.1379
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: HbA1c is glycosylated hemoglobin that routinely performed in a patient with Diabetes Mellitus. There are several methods for HbA1c measurement. One of them is based on the molecular charge such as High-Performance Liquid Chromatography (HPLC) and Capillary Electrophoresis (CE). Each method has its advantages and disadvantages. Thus, the selection of HbA1c measurement method is an important thing. This study was aimed to compare HbA1c measurement with HPLC and CE method. Methods: HbA1c measurement was done by HPLC and CE methods in 110 patient samples which have data of average blood glucose level during three months. We investigate the correlation of HbA1c results with average blood glucose level during three months, precision, methods comparison and interference including Hb variant. This study was conducted in the Central Laboratory of dr. Saiful Anwar Hospital during March-August 2018. Correlation analysis was done by Pearson analysis while precision test and methods comparison were done by Bland Altman Plot and Passing-Bablok Regression analysis. Results: There were correlation between estimated Average Glucose from HbA1c (eAGHbA1c) calculation with an average blood glucose level during 3 months in both methods. From precision test we found intraassay coefficient of variation in normal level (HbA1c +/- 5%) by HPLC vs CE was 1,7% vs 0.63% and in high level (HbA1c +/- 11%) was 0,59% vs 0,61% and inter-assay coefficient of variation by HPLC vs CE was 2,47% vs 2,17%. Method comparison results showed a strong correlation on both methods. The results of HbA1c in the group without Hb variant by CE method was slightly lower than HPLC. In Hb variant group, the CE method was slightly higher than HPLC, but it was not significant for HbA1c interpretation. Conclusion: HbA1c measurement by HPLC and CE method are comparable, and there is no significant difference in the results obtained.
引用
收藏
页码:316 / 322
页数:7
相关论文
共 50 条
  • [31] High-Resolution HbA1c Separation and Hemoglobinopathy Detection With Capillary Electrophoresis
    Urrechaga, Eloisa
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2012, 138 (03) : 448 - 456
  • [32] PERFORMANCE OF THE BECKMAN DIATRAC GEL-ELECTROPHORESIS METHOD FOR THE MEASUREMENT OF HBA1C
    CHOWNE, CW
    STAMP, RJ
    ANNALS OF CLINICAL BIOCHEMISTRY, 1992, 29 : 565 - 566
  • [33] TINIA检测HbA1c的性能评价及和IE-HPLC在HbE患者中的应用比较
    温冬梅
    张秀明
    王伟佳
    陈亚琼
    索明环
    徐全中
    吴剑杨
    李曼
    萧金丽
    阚丽娟
    周甲思
    检验医学, 2014, 29 (05) : 505 - 512
  • [34] Comparison of Glycated Hemoglobin (HbA1c) Values Estimated by High-Performance Liquid Chromatography and Spectrophotometry: A Pilot Study
    Ray, Avik
    Atal, Shubham
    Sharma, Swati
    Sampath, Ananyan
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (03)
  • [35] Incidence of sample storage temperature on HbA1c determination by high performance liquid chromatography method
    Szymezak, J.
    Lavalard, E.
    Leroy, N.
    Gillery, P.
    ANNALES DE BIOLOGIE CLINIQUE, 2009, 67 (06) : 705 - 710
  • [36] Comparison of Enzymatic Assay for HBA1C Measurement (Abbott Architect) With Capillary Electrophoresis (Sebia Minicap Flex Piercing Analyser)
    Kuna, Andrea Tesija
    Dukic, Kristina
    Gabaj, Nora Nikolac
    Miler, Marijana
    Vukasovic, Ines
    Langer, Sanja
    Simundic, Ana-Maria
    Vrkic, Nada
    LABORATORY MEDICINE, 2018, 49 (03) : 231 - 238
  • [37] STABILITY OF HBA1C BY ION-EXCHANGE, AFFINITY, HPLC AND ISOELECTRIC-FOCUSING
    MULLINS, RE
    PAPADEA, C
    AUSTIN, GE
    CLINICAL CHEMISTRY, 1983, 29 (06) : 1230 - 1230
  • [38] Unnoticed shift in the normal range of glycated HbA1c with TOSOH high performance liquid chromatography
    Kloos, C.
    Heller, T.
    Boeer, K.
    Hunger-Battefeld, W.
    Wolf, G.
    Mueller, U. A.
    DIABETOLOGIA, 2011, 54 : S401 - S401
  • [39] Differential Effects on HbA1c Detection by HPLC and Capillary Electrophoresis in Five Types of Hb Variants in China
    Luo Wei
    Zhang Weifeng
    Dong Zuoliang
    LABORATORY MEDICINE, 2023, 54 (03) : 308 - 316
  • [40] New rapid methods for determination of total LAS in sewage sludge by high performance liquid chromatography (HPLC) and capillary electrophoresis (CE)
    Villar, M.
    Callejon, M.
    Jimenez, J. C.
    Alonso, E.
    Guiraum, A.
    ANALYTICA CHIMICA ACTA, 2009, 634 (02) : 267 - 271