The comparison of ion exchange-high performance liquid chromatography (IE-HPLC) and capillary electrophoresis (CE) for HbA1C measurement

被引:0
|
作者
Sutrisnani, Catur Suci [1 ]
Darmawan, Edwin [2 ]
Widijanti, Anik [2 ]
Soehita, Sidarti [3 ]
机构
[1] Airlangga Univ, Clin Pathol Dept, Fac Med, Surabaya, Indonesia
[2] Brawijaya Univ, Dept Clin Pathol, Fac Med, Dr Saiful Anwar Hosp, Malang, Indonesia
[3] Airlangga Univ, Dept Clin Pathol, Fac Med, Dr Soetomo Hosp, Surabaya, Indonesia
关键词
HbA1c; measurement; HPLC; CE; HEMOGLOBIN A(1C) MEASUREMENT; PROGRAM; GOALS;
D O I
10.15562/bmj.v8i2.1379
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: HbA1c is glycosylated hemoglobin that routinely performed in a patient with Diabetes Mellitus. There are several methods for HbA1c measurement. One of them is based on the molecular charge such as High-Performance Liquid Chromatography (HPLC) and Capillary Electrophoresis (CE). Each method has its advantages and disadvantages. Thus, the selection of HbA1c measurement method is an important thing. This study was aimed to compare HbA1c measurement with HPLC and CE method. Methods: HbA1c measurement was done by HPLC and CE methods in 110 patient samples which have data of average blood glucose level during three months. We investigate the correlation of HbA1c results with average blood glucose level during three months, precision, methods comparison and interference including Hb variant. This study was conducted in the Central Laboratory of dr. Saiful Anwar Hospital during March-August 2018. Correlation analysis was done by Pearson analysis while precision test and methods comparison were done by Bland Altman Plot and Passing-Bablok Regression analysis. Results: There were correlation between estimated Average Glucose from HbA1c (eAGHbA1c) calculation with an average blood glucose level during 3 months in both methods. From precision test we found intraassay coefficient of variation in normal level (HbA1c +/- 5%) by HPLC vs CE was 1,7% vs 0.63% and in high level (HbA1c +/- 11%) was 0,59% vs 0,61% and inter-assay coefficient of variation by HPLC vs CE was 2,47% vs 2,17%. Method comparison results showed a strong correlation on both methods. The results of HbA1c in the group without Hb variant by CE method was slightly lower than HPLC. In Hb variant group, the CE method was slightly higher than HPLC, but it was not significant for HbA1c interpretation. Conclusion: HbA1c measurement by HPLC and CE method are comparable, and there is no significant difference in the results obtained.
引用
收藏
页码:316 / 322
页数:7
相关论文
共 50 条
  • [41] Evaluation of hemoglobin J-Baltimore interference on HbA1c measurement using capillary electrophoresis
    Kittanakom, Saranya
    Halchuk, Linda M.
    McFarland, Andrew
    Macri, Joseph
    CLINICAL BIOCHEMISTRY, 2014, 47 (12) : 1152 - 1152
  • [42] Metabolism of substance P in rat striatum studies using a combination of high performance liquid chromatography (HPLC) and capillary electrophoresis (CE)
    MichaelTitus, AT
    Whelpton, R
    Stephens, SM
    Yau, KW
    BRITISH JOURNAL OF PHARMACOLOGY, 1997, 122 : P79 - P79
  • [43] Measurement of HbA1c in Gingival Crevicular Blood Using a High-Pressure Liquid Chromatography Procedure
    Pesce, Michael A.
    Strauss, Shiela M.
    Rosedale, Mary
    Netterwald, Jane
    Wang, Hangli
    LABMEDICINE, 2015, 46 (04): : 290 - 298
  • [44] HIGH-PERFORMANCE ION-EXCHANGE CHROMATOGRAPHY AS A PREPARATIVE ADJUNCT TO CAPILLARY ELECTROPHORESIS
    GUARINO, BC
    PHILLIPS, D
    AMERICAN LABORATORY, 1991, 23 (05) : 68 - 69
  • [45] Harmonisation of HbA1c in Sweden using mono S ion-exchange chromatography
    Larsson, B
    Malm, J
    Nordin, G
    Jeppsson, JO
    CLINICA CHIMICA ACTA, 2005, 355 : S414 - S414
  • [46] Strong correlation and high comparability of capillary electrophoresis and three different methods for HbA1c measurement in a population without hemoglobinopathy
    Sriwimol, Wilaiwan
    Choosongsang, Pensiri
    Choosongsang, Phattanapong
    Treerut, Pittaya
    Muenniam, Baralee
    Makkong, Paspimon
    Pocathikorn, Anothai
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2020, 80 (02): : 139 - 150
  • [47] Glycation rate of haemoglobins S, C, D, E, J and G, and analytical interference on the measurement of HbA1c with affinity chromatography and capillary electrophoresis
    Weykamp, Cas
    Kemna, Erwin
    Leppink, Sanne
    Siebelder, Carla
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (09) : E207 - E210
  • [48] PURIFIED HEMOGLOBIN PREPARATIONS IN THE EVALUATION OF HBA1C DETERMINATION BY ION-EXCHANGE CHROMATOGRAPHY
    COHEN, MP
    WITT, J
    WU, VY
    ANNALS OF CLINICAL BIOCHEMISTRY, 1993, 30 : 265 - 271
  • [49] DETERMINATION OF NITRATE AND NITRITE IONS IN HUMAN-PLASMA BY ION EXCHANGE-HIGH PERFORMANCE LIQUID-CHROMATOGRAPHY
    OSTERLOH, J
    GOLDFIELD, D
    JOURNAL OF LIQUID CHROMATOGRAPHY, 1984, 7 (04): : 753 - 763
  • [50] Capillary blood on filter paper for determination of HbA(1c) by ion exchange chromatography
    Jeppsson, JO
    Jerntorp, P
    Almer, LO
    Persson, R
    Ekberg, G
    Sundkvist, G
    DIABETES CARE, 1996, 19 (02) : 142 - 145