Novel scintillation detector design and performance for proton radiography and computed tomography

被引:63
|
作者
Bashkirov, V. A. [1 ]
Schulte, R. W. [1 ]
Hurley, R. F. [1 ]
Johnson, R. P. [2 ]
Sadrozinski, H. F. -W. [2 ]
Zatserklyaniy, A. [2 ]
Plautz, T. [2 ]
Giacometti, V. [3 ]
机构
[1] Loma Linda Univ, Dept Basic Sci, 11175 Campus St, Loma Linda, CA 92354 USA
[2] Univ Calif Santa Cruz, Dept Phys, 1156 High St, Santa Cruz, CA 95064 USA
[3] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
proton computed tomography; proton radiography; proton therapy; WEPL; proton energy detector; proton range measurement;
D O I
10.1118/1.4939255
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Proton computed tomography (pCT) will enable accurate prediction of proton and ion range in a patient while providing the benefit of lower radiation exposure than in x-ray CT. The accuracy of the range prediction is essential for treatment planning in proton or ion therapy and depends upon the detector used to evaluate the water-equivalent path length (WEPL) of a proton passing through the object. A novel approach is presented for an inexpensive WEPL detector for pCT and proton radiography. Methods: A novel multistage detector with an aperture of 10 x 37.5 cm was designed to optimize the accuracy of the WEPL measurements while simplifying detector construction and the performance requirements of its components. The design of the five-stage detector was optimized through simulations based on the GEANT4 detector simulation toolkit, and the fabricated prototype was calibrated in water-equivalent millimeters with 200 MeV protons in the research beam line of the clinical proton synchrotron at Loma Linda University Medical Center. A special polystyrene step phantom was designed and built to speed up and simplify the calibration procedure. The calibrated five-stage detector was tested in the 200 MeV proton beam as part of the pCT head scanner, using a water phantom and polystyrene slabs to verify the WEPL reconstruction accuracy. Results: The beam-test results demonstrated excellent performance of the new detector, in good agreement with the simulation results. The WEPL measurement accuracy is about 3 0 mm per proton in the 0-260 mm WEPL range required for a pCT head scan with a 200 MeV proton beam. Conclusions: The new multistage design approach to WEPL measurements for proton CT and radiography has been prototyped and tested. The test results show that the design is competitive with much more expensive calorimeter and range-counter designs. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:664 / 674
页数:11
相关论文
共 50 条
  • [41] Evaluation of a Novel Plastic Scintillation Detector Performance in MR-LINAC
    Yun, K.
    Hwang, M.
    Lee, D.
    Oh, S.
    Feijoo, M.
    Beron, S.
    Pavord, D.
    MEDICAL PHYSICS, 2022, 49 (06) : E912 - E913
  • [42] Construction and performance of a dose-verification scintillation-fiber detector for proton therapy
    Lee, Suhyun
    Hong, Byungsik
    Lee, Kyong Sei
    Mulilo, Benard
    Park, Sung Keun
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 724 : 6 - 11
  • [43] Monte Carlo studies on proton computed tomography using a silicon strip detector telescope
    Johnson, LR
    Keeney, B
    Ross, G
    Sadrozinski, HEW
    Seiden, A
    Williams, DC
    Zhang, L
    Bashkirov, V
    Schulte, RW
    Shahnazi, K
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 916 - 920
  • [44] RADIONUCLIDE EMISSION COMPUTED TOMOGRAPHY WITH A SCINTILLATION CAMERA
    JASZCZAK, R
    HUARD, D
    MURPHY, P
    BURDINE, J
    JOURNAL OF NUCLEAR MEDICINE, 1976, 17 (06) : 551 - 551
  • [45] PROTON COMPUTED-TOMOGRAPHY
    HANSON, KM
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (01) : 1635 - 1640
  • [46] Issues in proton computed tomography
    Sadrozinski, HFW
    Bashkirov, V
    Bruzzi, M
    Johnson, LR
    Keeney, B
    Ross, G
    Schulte, RW
    Seiden, A
    Shahnazi, K
    Williams, DC
    Zhang, L
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 511 (1-2): : 275 - 281
  • [47] Toward proton computed tomography
    Sadrozinski, HFW
    Bashkirov, V
    Keeney, B
    Johnson, LR
    Peggs, SG
    Ross, G
    Satogata, T
    Schulte, RWM
    Seiden, A
    Shanazi, K
    Williams, DC
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (01) : 3 - 9
  • [48] Clinical comparison of imaging performance of a digital radiography system with a new flat-panel detector vs computed radiography
    Mildenberger, P
    Vogel, N
    Herber, SS
    Kaemmerer, M
    Thelen, M
    RADIOLOGY, 2000, 217 : 518 - 518
  • [49] APPLICATION OF COMPUTED AXIAL TOMOGRAPHY TO INDUSTRIAL RADIOGRAPHY
    KRUGER, RP
    CANNON, TM
    MATERIALS EVALUATION, 1977, 35 (06) : 21 - 21
  • [50] Persistent Nephrogram on Abdominal Radiography and Computed Tomography
    Maekawa, Michitaka
    Imaizumi, Takahiro
    Yamakawa, Taishi
    INTERNAL MEDICINE, 2014, 53 (21) : 2551 - 2552