Novel scintillation detector design and performance for proton radiography and computed tomography

被引:63
|
作者
Bashkirov, V. A. [1 ]
Schulte, R. W. [1 ]
Hurley, R. F. [1 ]
Johnson, R. P. [2 ]
Sadrozinski, H. F. -W. [2 ]
Zatserklyaniy, A. [2 ]
Plautz, T. [2 ]
Giacometti, V. [3 ]
机构
[1] Loma Linda Univ, Dept Basic Sci, 11175 Campus St, Loma Linda, CA 92354 USA
[2] Univ Calif Santa Cruz, Dept Phys, 1156 High St, Santa Cruz, CA 95064 USA
[3] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
proton computed tomography; proton radiography; proton therapy; WEPL; proton energy detector; proton range measurement;
D O I
10.1118/1.4939255
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Proton computed tomography (pCT) will enable accurate prediction of proton and ion range in a patient while providing the benefit of lower radiation exposure than in x-ray CT. The accuracy of the range prediction is essential for treatment planning in proton or ion therapy and depends upon the detector used to evaluate the water-equivalent path length (WEPL) of a proton passing through the object. A novel approach is presented for an inexpensive WEPL detector for pCT and proton radiography. Methods: A novel multistage detector with an aperture of 10 x 37.5 cm was designed to optimize the accuracy of the WEPL measurements while simplifying detector construction and the performance requirements of its components. The design of the five-stage detector was optimized through simulations based on the GEANT4 detector simulation toolkit, and the fabricated prototype was calibrated in water-equivalent millimeters with 200 MeV protons in the research beam line of the clinical proton synchrotron at Loma Linda University Medical Center. A special polystyrene step phantom was designed and built to speed up and simplify the calibration procedure. The calibrated five-stage detector was tested in the 200 MeV proton beam as part of the pCT head scanner, using a water phantom and polystyrene slabs to verify the WEPL reconstruction accuracy. Results: The beam-test results demonstrated excellent performance of the new detector, in good agreement with the simulation results. The WEPL measurement accuracy is about 3 0 mm per proton in the 0-260 mm WEPL range required for a pCT head scan with a 200 MeV proton beam. Conclusions: The new multistage design approach to WEPL measurements for proton CT and radiography has been prototyped and tested. The test results show that the design is competitive with much more expensive calorimeter and range-counter designs. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:664 / 674
页数:11
相关论文
共 50 条
  • [31] Cosine fitting radiography and computed tomography
    李盼云
    张凯
    黄万霞
    袁清习
    王研
    鞠在强
    吴自玉
    朱佩平
    Chinese Physics B, 2015, (06) : 683 - 690
  • [32] COMPUTED TOMOGRAPHY VERSUS SKULL RADIOGRAPHY
    WEINSTEIN, MA
    ALFIDI, RJ
    DUCHESNEAU, PM
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1977, 128 (05) : 873 - 873
  • [33] Cosine fitting radiography and computed tomography
    Li Pan-Yun
    Zhang Kai
    Huang Wan-Xia
    Yuan Qing-Xi
    Wang Yan
    Ju Zai-Qiang
    Wu Zi-Yu
    Zhu Pei-Ping
    CHINESE PHYSICS B, 2015, 24 (06)
  • [34] Performance of a prototype of liquid xenon scintillation detector system for positron emission tomography
    Nishikido, F
    Doke, T
    Kikuchi, J
    Mori, T
    Takizawa, K
    Yamamoto, M
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 2004, 43 (02): : 779 - 784
  • [35] CONCEPTUAL DESIGN OF PROTON COMPUTED-TOMOGRAPHY WITH MAGNETIC SPECTROMETER
    AKISADA, M
    OHASHI, J
    KONDO, K
    KURIHARA, D
    MIYASHITA, S
    TACHIKAWA, A
    TAKADA, Y
    TAKIKAWA, K
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1983, 22 (04): : 752 - 758
  • [36] Pencil beam approach to proton computed tomography: a performance study
    Rescigno, Regina
    Bopp, Cecile
    Rousseau, Marc
    Brasse, David
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [37] Imaging performance of a LaBr3:Ce scintillation detector for photon counting x-ray computed tomography: Simulation study
    Taguchi, Katsuyuki
    Schaart, Dennis R.
    Goorden, Marlies C.
    Hsieh, Scott S.
    MEDICAL PHYSICS, 2024,
  • [38] Optimization of a Scintillation Detector for Positron Emission Tomography
    Belov, M. V.
    Zavertyaev, M. V.
    Kozlov, V. A.
    Tskhay, V. S.
    BULLETIN OF THE LEBEDEV PHYSICS INSTITUTE, 2024, 51 (07) : 268 - 271
  • [39] Scintillation Detector Design Study for Prompt Gamma Photon Detection in Proton Therapy Monitoring
    Zhang, Hongyang
    Fan, Peng
    Xu, Tianpeng
    Wei, Qingyang
    Ma, Tianyu
    Liu, Yaqiang
    Wu, Zhaoxia
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [40] Assembling and Test of a Proton Computed Radiography Apparatus
    Menichelli, David
    Bruzzi, Mara
    Bucciolini, Marta
    Candiano, Giuliana
    Cirrone, Giulio A. P.
    Civinini, Carlo
    Cuttone, Giacomo
    Lo Presti, Domenico
    Marrazzo, Livia
    Pallotta, Stefania
    Randazzo, Nunzio
    Scaringella, Monica
    Sipala, Valeria
    Talamonti, Cinzia
    Brianzi, Mirko
    Tesi, Mauro
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 4181 - +