Simultaneous Hip Head-Stem Taper Junction Measurements of Electrochemical Corrosion and Micromotion: A Comparison of Taper Geometry and Stem Material

被引:2
|
作者
Swaminathan, V. [1 ]
Scholl, L. [1 ]
Lee, R. [1 ]
Faizan, A. [1 ]
Thakore, M. [1 ]
TenHuisen, K. [1 ]
Nevelos, J. [1 ]
机构
[1] Stryker Orthopaed, 325 Corp Dr, Mahwah, NJ 07430 USA
关键词
corrosion; micromotion; electrochemistry; onset load; tapers; FRETTING CORROSION; CREVICE CORROSION; INTERFACE;
D O I
10.1520/STP159120140150
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Fretting corrosion or mechanically assisted crevice corrosion is a current topic of research in total hip arthroplasty highlighted by the modularity of the head-neck taper junction. The extent of corrosion may be influenced by a multitude of factors, and in this study, we examined two specific variables: material combination and taper geometry. Corrosion performance was quantitatively assessed by measuring fretting corrosion at the taper junction as well as relative micromotion between the head and stem components. To examine the effect of material combinations, stems composed of either Ti-6AI-4V or TMZF (R) were tested with CoCrMo femoral heads. To study taper geometry, a V40 (TM) taper design was compared with a C-taper design. Cyclic fatigue loading was applied to the test specimens, increasing load from 100 to 3200 N and then continuing loading to 1 million cycles. All tests were performed in phosphate-buffered saline solution at room temperature and electrochemical data analysis included determination of onset load, average current, average open-circuit potential, peak-to-peak current, and absolute change in potential. Micromotion fixtures were attached rigidly to the head and proximal stem. The micromotion output was used to calculate pistoning of the head along the stem, rocking of the head about the stem, and subsidence (or seating or slipping) of the head against the stem. The electrochemical data showed that the mean onset load for different groups was in the range of 1075 to 1194 N and average currents up to 2 mu A were recorded for different sample groups during cyclic loading. Fretting corrosion and micromotion measurements in this study showed that the variation produced by material combination and taper design within the range of test groups tested were less than the variation that existed in each test group.
引用
收藏
页码:321 / 335
页数:15
相关论文
共 48 条
  • [31] Material Loss at the Taper Junction of Retrieved Large Head Metal-on-Metal Total Hip Replacements
    Matthies, Ashley K.
    Racasan, Radu
    Bills, Paul
    Blunt, Liam
    Cro, Suzie
    Panagiotidou, Anna
    Blunn, Gordon
    Skinner, John
    Hart, Alister J.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (11) : 1677 - 1685
  • [32] Variability in stem taper surface topography affects the degree of corrosion and fretting in total hip arthroplasty
    Kilian Elia Stockhausen
    Christoph Riedel
    Alex Victoria Belinski
    Dorothea Rothe
    Thorsten Gehrke
    Felix Klebig
    Matthias Gebauer
    Michael Amling
    Mustafa Citak
    Björn Busse
    Scientific Reports, 11
  • [33] Variability in stem taper surface topography affects the degree of corrosion and fretting in total hip arthroplasty
    Stockhausen, Kilian Elia
    Riedel, Christoph
    Belinski, Alex Victoria
    Rothe, Dorothea
    Gehrke, Thorsten
    Klebig, Felix
    Gebauer, Matthias
    Amling, Michael
    Citak, Mustafa
    Busse, Bjorn
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [34] The Relationship Between Cobalt/Chromium Ratios and the High Prevalence of Head-Stem Junction Corrosion in Metal-on-Metal Total Hip Arthroplasty
    Hothi, Harry S.
    Berber, Reshid
    Whittaker, Robert K.
    Blunn, Gordon W.
    Skinner, John A.
    Hart, Alister J.
    JOURNAL OF ARTHROPLASTY, 2016, 31 (05): : 1123 - 1127
  • [35] Femoral stem taper geometry and porous coating in cementless direct anterior primary total hip arthroplasty
    McCrosson, Matthew
    Broadfoot, Jackson
    Yeager, Matthew
    Marquess, Ben
    Scheinberg, Mila
    Naranje, Sameer
    JOURNAL OF ORTHOPAEDICS, 2023, 46 : 169 - 173
  • [36] Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers
    Bechstedt, Maren
    Gustafson, Jonathan A.
    Mell, Steven P.
    Guehrs, Julian
    Morlock, Michael M.
    Levine, Brett R.
    Lundberg, Hannah J.
    JOURNAL OF BIOMECHANICS, 2020, 103
  • [37] Stem taper mismatch has a critical effect on ceramic head fracture risk in modular hip arthroplasty
    Guehrs, Julian
    Koerner, Mandy
    Bechstedt, Maren
    Krull, Annika
    Morlock, Michael M.
    CLINICAL BIOMECHANICS, 2017, 41 : 106 - 110
  • [38] Femoral head material loss at the head-neck junction in total hip arthroplasty: the effect of head size, stem material and stem offset
    Valente, Giuseppe
    Lanting, Brent
    MacDonald, Steven
    Teeter, Matthew G.
    Van Citters, Douglas
    Howard, James
    HIP INTERNATIONAL, 2019, 29 (06) : 647 - 651
  • [39] An intelligent system for image-based rating of corrosion severity at stem taper of retrieved hip replacement implants
    Milimonfared, Roohollah
    Oskouei, Reza H.
    Taylor, Mark
    Solomon, Lucian B.
    MEDICAL ENGINEERING & PHYSICS, 2018, 61 : 13 - 24
  • [40] Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses
    Rehmer, Annelie
    Bishop, Nicholas E.
    Morlock, Michael M.
    CLINICAL BIOMECHANICS, 2012, 27 (01) : 77 - 83