Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses

被引:123
|
作者
Rehmer, Annelie [1 ]
Bishop, Nicholas E. [1 ]
Morlock, Michael M. [1 ]
机构
[1] Hamburg Univ Technol, TUHH, Biomech Sect, D-21073 Hamburg, Germany
关键词
Head-taper connection; Assembly force; Pull-off forces; Turn-off capacity; Modular hip prosthesis; FEMORAL-HEAD; ARTHROPLASTY; REPLACEMENTS; COMPONENT; BEARINGS; STEM; CORROSION; FRICTION; IMPLANTS; WEAR;
D O I
10.1016/j.clinbiomech.2011.08.002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: A stable fixation between femoral head and endoprosthesis taper is necessary to prevent relative motions and corrosion at the taper junction. Although the importance of the component assembly has been recognised, no definitive instructions are available. The purpose of this study was to assess the influence of assembly force, assembly tool and number of hammer strokes on the taper junction strength of various material combinations. Methods: Co-Cr29-Mo (n=10) and Ti-6Al-4V (n=10) neck tapers were assembled with Co-Cr and Al2O3 ceramic heads either by push-on or by impaction with single or multiple hammer blows. The strength of the taper-head connection was evaluated by measuring the head pull-off forces according to ISO 7206-10 and the turn-off moment capacity. Findings: The taper strength linearly increased with assembly forces (P<0.001). Co-Cr heads combined with Co-Cr tapers showed significantly lower pull-off forces and turn-off moments than the combination with Ti tapers (0.001<P<0.025). Multiple impaction did not increase taper strength (0.063<P<0.995). Ceramic and Co-Cr heads showed similar fixation patterns on Ti tapers. Turn-off moments varied between 6 Nm and 19 Nm, dependent on material combination and assembly force. Interpretation: It is suggested that sufficient head-taper junction strength in all bearing conditions is achieved by impaction forces of at least 4 kN. A single impact is sufficient to achieve fixation. Special attention should be paid to the assembly of Co-Cr heads on Co-Cr tapers. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:77 / 83
页数:7
相关论文
共 22 条
  • [1] Influence of geometry and materials on the axial and torsional strength of the head-neck taper junction in modular hip replacements: A finite element study
    Fallahnezhad, Khosro
    Farhoudi, Hamidreza
    Oskouei, Reza H.
    Taylor, Marlz
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 60 : 118 - 126
  • [2] Fixation strength of taper connection at head-neck junction in retrieved carbon fiber-reinforced PEEK hip stems
    Nakahara, Ichiro
    Takao, Masaki
    Bandoh, Shunichi
    Sugano, Nobuhiko
    JOURNAL OF ARTIFICIAL ORGANS, 2014, 17 (04) : 358 - 363
  • [3] Influence of surface topography on junction strength of modular hip endoprostheses
    Denkena, Berend
    Hurschler, Christof
    Bergmann, Benjamin
    Legutko, Beate
    Gustav, Marco
    Welke, Bastian
    V CIRP CONFERENCE ON BIOMANUFACTURING, 2022, 110 : 64 - 69
  • [4] Finite element analysis of the head-neck taper interface of modular hip prostheses
    Dyrkacz, R. M. R.
    Brandt, J. M.
    Morrison, J. B.
    O' Brien, S. T.
    Ojo, O. A.
    Turgeon, T. R.
    Wyss, U. P.
    TRIBOLOGY INTERNATIONAL, 2015, 91 : 206 - 213
  • [5] Influence of the use of an adhesive connection on the joint strength of modular hip endoprostheses
    Einfeldt, Ann-Kathrin
    Legutko, Beate
    Pott, Philipp-Cornelius
    Bergmann, Benjamin
    Denkena, Berend
    Hurschler, Christof
    Welke, Bastian
    PLOS ONE, 2024, 19 (11):
  • [6] The Influence of Assembly Force on the Material Loss at the Metallic Head-Neck Junction of Hip Implants Subjected to Cyclic Fretting Wear
    Fallahnezhad, Khosro
    Oskouei, Reza H.
    Badnava, Hojjat
    Taylor, Mark
    METALS, 2019, 9 (04):
  • [7] Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers
    Bechstedt, Maren
    Gustafson, Jonathan A.
    Mell, Steven P.
    Guehrs, Julian
    Morlock, Michael M.
    Levine, Brett R.
    Lundberg, Hannah J.
    JOURNAL OF BIOMECHANICS, 2020, 103
  • [8] Fixation strength of taper connection at head–neck junction in retrieved carbon fiber-reinforced PEEK hip stems
    Ichiro Nakahara
    Masaki Takao
    Shunichi Bandoh
    Nobuhiko Sugano
    Journal of Artificial Organs, 2014, 17 : 358 - 363
  • [9] Evidence based recommendations for reducing head-neck taper connection fretting corrosion in hip replacement prostheses
    Wight, Christian M.
    Lanting, Brent
    Schemitsch, Emil H.
    HIP INTERNATIONAL, 2017, 27 (06) : 523 - 531
  • [10] Model validation for estimating taper microgroove deformation during total hip arthroplasty head-neck assembly
    Godoy, Michael
    Gustafson, Jonathan A.
    Hertzler, Justin S.
    Bischoff, Jeffrey E.
    Pourzal, Robin
    Lundberg, Hannah J.
    JOURNAL OF BIOMECHANICS, 2022, 140