Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses

被引:123
|
作者
Rehmer, Annelie [1 ]
Bishop, Nicholas E. [1 ]
Morlock, Michael M. [1 ]
机构
[1] Hamburg Univ Technol, TUHH, Biomech Sect, D-21073 Hamburg, Germany
关键词
Head-taper connection; Assembly force; Pull-off forces; Turn-off capacity; Modular hip prosthesis; FEMORAL-HEAD; ARTHROPLASTY; REPLACEMENTS; COMPONENT; BEARINGS; STEM; CORROSION; FRICTION; IMPLANTS; WEAR;
D O I
10.1016/j.clinbiomech.2011.08.002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: A stable fixation between femoral head and endoprosthesis taper is necessary to prevent relative motions and corrosion at the taper junction. Although the importance of the component assembly has been recognised, no definitive instructions are available. The purpose of this study was to assess the influence of assembly force, assembly tool and number of hammer strokes on the taper junction strength of various material combinations. Methods: Co-Cr29-Mo (n=10) and Ti-6Al-4V (n=10) neck tapers were assembled with Co-Cr and Al2O3 ceramic heads either by push-on or by impaction with single or multiple hammer blows. The strength of the taper-head connection was evaluated by measuring the head pull-off forces according to ISO 7206-10 and the turn-off moment capacity. Findings: The taper strength linearly increased with assembly forces (P<0.001). Co-Cr heads combined with Co-Cr tapers showed significantly lower pull-off forces and turn-off moments than the combination with Ti tapers (0.001<P<0.025). Multiple impaction did not increase taper strength (0.063<P<0.995). Ceramic and Co-Cr heads showed similar fixation patterns on Ti tapers. Turn-off moments varied between 6 Nm and 19 Nm, dependent on material combination and assembly force. Interpretation: It is suggested that sufficient head-taper junction strength in all bearing conditions is achieved by impaction forces of at least 4 kN. A single impact is sufficient to achieve fixation. Special attention should be paid to the assembly of Co-Cr heads on Co-Cr tapers. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:77 / 83
页数:7
相关论文
共 22 条
  • [21] Influence of material coupling and assembly condition on the magnitude of micromotion at the stem-neck interface of a modular hip endoprosthesis
    Jauch, S. Y.
    Huber, G.
    Hoenig, E.
    Baxmann, M.
    Grupp, T. M.
    Morlock, M. M.
    JOURNAL OF BIOMECHANICS, 2011, 44 (09) : 1747 - 1751
  • [22] Contact Mechanics and Plastic Deformation at the Local Surface Topography Level After Assembly of Modular Head-Neck Junctions in Modern Total Hip Replacement Devices
    Lundberg, Hannah J.
    Ha, Nguyen Q.
    Hall, Deborah J.
    Urban, Robert M.
    Levine, Brett R.
    Pourzal, Robin
    MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 59 - 82