Fixation strength of taper connection at head-neck junction in retrieved carbon fiber-reinforced PEEK hip stems

被引:9
|
作者
Nakahara, Ichiro [1 ]
Takao, Masaki [2 ]
Bandoh, Shunichi [3 ]
Sugano, Nobuhiko [2 ,4 ]
机构
[1] Osaka Natl Hosp, Dept Orthopaed Surg, Chuo Ku, Osaka 5400006, Japan
[2] Osaka Univ, Grad Sch Med, Dept Orthopaed Surg, Suita, Osaka 5650871, Japan
[3] BITEC Co Ltd, Kakamigahara, Japan
[4] Osaka Univ, Grad Sch Med, Dept Orthopaed Med Engn, Suita, Osaka 5650871, Japan
关键词
Carbon-fiber-reinforced PEEK; Hip stem; Ceramic head; Head-neck taper junction; Pull-off strength; ORTHOPEDIC IMPLANTS; PROSTHESIS DESIGN; FEMORAL COMPONENT; METAL; ARTHROPLASTY; COMPOSITE; INTERFACE; CORROSION; POLYETHERETHERKETONE; BIOCOMPATIBILITY;
D O I
10.1007/s10047-014-0789-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Carbon fiber-reinforced polyetheretherketone (CFR-PEEK) hip prostheses possess numerous advantages over metal prostheses; however, the security of the taper connection between the CFR-PEEK stem and the modular femoral head in vivo has not been verified. Therefore, we mechanically examined the taper connection of retrieved in vivo loaded CFR-PEEK stems in comparison with in vivo loaded titanium alloy stems. CFR-PEEK and titanium alloy femoral stems with a 12/14 taper trunnion were implanted in ovine hips. A 22-mm ceramic head was intraoperatively impacted to the stem. Retrieved specimens were obtained following weight-bearing conditions for up to 39 postoperative weeks and taper junction pull-off tests were conducted. Postoperative retrieved CFR-PEEK stem pull-off strength was significantly greater than that at time zero. Postoperative retrieved CFR-PEEK stem pull-off strength was also significantly higher than that of postoperative retrieved titanium alloy stem. Microscopic findings of the taper surface revealed no obvious damage in the retrieved CFR-PEEK stems, whereas fretting and corrosion were observed in the retrieved titanium alloy stems. The present findings suggest that the taper connection between the ceramic head and the 12/14 CFR-PEEK stem trunnion is more secure than that between the ceramic head and the titanium alloy trunnion.
引用
收藏
页码:358 / 363
页数:6
相关论文
共 15 条
  • [1] Fixation strength of taper connection at head–neck junction in retrieved carbon fiber-reinforced PEEK hip stems
    Ichiro Nakahara
    Masaki Takao
    Shunichi Bandoh
    Nobuhiko Sugano
    Journal of Artificial Organs, 2014, 17 : 358 - 363
  • [2] AN ANALYSIS OF THE HEAD-NECK TAPER INTERFACE IN RETRIEVED HIP PROSTHESES
    LIEBERMAN, JR
    RIMNAC, CM
    GARVIN, KL
    KLEIN, RW
    SALVATI, EA
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1994, (300) : 162 - 167
  • [3] Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses
    Rehmer, Annelie
    Bishop, Nicholas E.
    Morlock, Michael M.
    CLINICAL BIOMECHANICS, 2012, 27 (01) : 77 - 83
  • [4] In vivo implant fixation of carbon fiber-reinforced PEEK hip prostheses in an ovine model
    Nakahara, Ichiro
    Takao, Masaki
    Bandoh, Shunichi
    Bertollo, Nicky
    Walsh, William R.
    Sugano, Nobuhiko
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (03) : 485 - 492
  • [5] Influence of geometry and materials on the axial and torsional strength of the head-neck taper junction in modular hip replacements: A finite element study
    Fallahnezhad, Khosro
    Farhoudi, Hamidreza
    Oskouei, Reza H.
    Taylor, Marlz
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 60 : 118 - 126
  • [6] Evidence based recommendations for reducing head-neck taper connection fretting corrosion in hip replacement prostheses
    Wight, Christian M.
    Lanting, Brent
    Schemitsch, Emil H.
    HIP INTERNATIONAL, 2017, 27 (06) : 523 - 531
  • [7] Laser Joining of Continuous Carbon Fiber-Reinforced PEEK and Titanium Alloy with High Strength
    Wang, Haipeng
    Ren, Zhongjing
    Guan, Yingchun
    POLYMERS, 2022, 14 (21)
  • [8] Characterization of carbon fiber-reinforced PEEK composite for use as a bearing material in total hip replacements
    Polineni, VK
    Wang, AG
    Essner, A
    Lin, R
    Chopra, A
    Stark, C
    Dumbleton, JH
    ALTERNATIVE BEARING SURFACES IN TOTAL JOINT REPLACEMENT, 1998, 1346 : 266 - 273
  • [9] Characterization of carbon fiber-reinforced peek composite for use as a bearing material in total hip replacements
    Polineni, V. Krishna
    Wang, Aiguo
    Essner, Aaron
    Lin, Ruey
    Chopra, Ash
    Stark, Cas
    Dumbleton, John H.
    ASTM Special Technical Publication, 1999, (1346): : 266 - 273
  • [10] Fracture fixation in extremity trauma with carbon fiber-reinforced polyetheretherketone (CFR-PEEK) plates: evidence today
    Chloros, George D.
    Prodromidis, Apostolos D.
    Wilson, Jo
    Giannoudis, Peter V.
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2022, 48 (03) : 2387 - 2406