Simultaneous Hip Head-Stem Taper Junction Measurements of Electrochemical Corrosion and Micromotion: A Comparison of Taper Geometry and Stem Material

被引:2
|
作者
Swaminathan, V. [1 ]
Scholl, L. [1 ]
Lee, R. [1 ]
Faizan, A. [1 ]
Thakore, M. [1 ]
TenHuisen, K. [1 ]
Nevelos, J. [1 ]
机构
[1] Stryker Orthopaed, 325 Corp Dr, Mahwah, NJ 07430 USA
关键词
corrosion; micromotion; electrochemistry; onset load; tapers; FRETTING CORROSION; CREVICE CORROSION; INTERFACE;
D O I
10.1520/STP159120140150
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Fretting corrosion or mechanically assisted crevice corrosion is a current topic of research in total hip arthroplasty highlighted by the modularity of the head-neck taper junction. The extent of corrosion may be influenced by a multitude of factors, and in this study, we examined two specific variables: material combination and taper geometry. Corrosion performance was quantitatively assessed by measuring fretting corrosion at the taper junction as well as relative micromotion between the head and stem components. To examine the effect of material combinations, stems composed of either Ti-6AI-4V or TMZF (R) were tested with CoCrMo femoral heads. To study taper geometry, a V40 (TM) taper design was compared with a C-taper design. Cyclic fatigue loading was applied to the test specimens, increasing load from 100 to 3200 N and then continuing loading to 1 million cycles. All tests were performed in phosphate-buffered saline solution at room temperature and electrochemical data analysis included determination of onset load, average current, average open-circuit potential, peak-to-peak current, and absolute change in potential. Micromotion fixtures were attached rigidly to the head and proximal stem. The micromotion output was used to calculate pistoning of the head along the stem, rocking of the head about the stem, and subsidence (or seating or slipping) of the head against the stem. The electrochemical data showed that the mean onset load for different groups was in the range of 1075 to 1194 N and average currents up to 2 mu A were recorded for different sample groups during cyclic loading. Fretting corrosion and micromotion measurements in this study showed that the variation produced by material combination and taper design within the range of test groups tested were less than the variation that existed in each test group.
引用
收藏
页码:321 / 335
页数:15
相关论文
共 48 条
  • [41] Is taper corrosion in modular revision hip stem junctions associated with patient or implant specific factors? A retrieval analysis
    Bormann, Therese
    Kretzer, J. Philippe
    Jaeger, Sebastian
    Lohmann, Christoph H.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2024, 150
  • [42] Strength of the taper junction of modular revision hip stems: The influence of contamination of the taper junction and the joining condition on the relative motion and pull-off force between the stem and neck part of the MRP prosthesis
    Boettcher J.M.
    Sellenschloh K.
    Strube A.
    Huber G.
    Morlock M.M.
    Die Orthopädie, 2024, 53 (1) : 47 - 55
  • [43] Influence of geometry and materials on the axial and torsional strength of the head-neck taper junction in modular hip replacements: A finite element study
    Fallahnezhad, Khosro
    Farhoudi, Hamidreza
    Oskouei, Reza H.
    Taylor, Marlz
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 60 : 118 - 126
  • [44] A novel method to improve femoral head and stem taper stability intraoperatively in total hip arthroplasty - a proof of concept study
    Dun, Shouchen
    Lim, Boon Him
    Swope, Stephen W.
    Whitaker, Dustin R.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2022, 108 (08)
  • [45] Clinical Results of Revision Hip Arthroplasty for Neck-Taper Corrosion and Adverse Local Tissue Reactions Around a Modular Neck Stem
    Inoue, Daisuke
    Restrepo, Camilo
    Nourie, Blake
    Hozack, William J.
    JOURNAL OF ARTHROPLASTY, 2020, 35 (06): : S289 - S293
  • [46] Material loss at the femoral head taper A COMPARISON STUDY OF THE EXETER METAL-ON-POLYETHYLENE AND CONTEMPORARY METAL-ON-METAL TOTAL HIP ARTHROPLASTY
    Langton, D. J.
    Wells, S. R.
    Joyce, T. J.
    Bowsher, J. G.
    Deehan, D.
    Green, S.
    Nargol, A. V. F.
    Holland, J. P.
    BONE & JOINT JOURNAL, 2018, 100B (10): : 1310 - 1319
  • [47] Does a micro-grooved trunnion stem surface finish improve fixation and reduce fretting wear at the taper junction of total hip replacements? A finite element evaluation
    Ashkanfar, Ariyan
    Langton, David J.
    Joyce, Thomas J.
    JOURNAL OF BIOMECHANICS, 2017, 63 : 47 - 54
  • [48] Adverse Local Tissue Reaction Arising from Corrosion at the Femoral Neck-Body Junction in a Dual-Taper Stem with a Cobalt-Chromium Modular Neck
    Cooper, H. John
    Urban, Robert M.
    Wixson, Richard L.
    Meneghini, R. Michael
    Jacobs, Joshua J.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (10): : 865 - 872