Analytical evaluation of direct bicarbonate measurement with the new gem premier chemstat in hemodialysis patients

被引:0
|
作者
Balboni, Fiamma [1 ]
Terreni, Alessandro [2 ]
Gallo, Marco [3 ]
Guzzi, Francesco [3 ]
Caparrini, Chiara [4 ]
Burbui, Stefania [5 ]
Vezzosi, Marco [1 ]
Galora, Silvia [1 ]
Lori, Giulia [1 ]
Lippi, Giuseppe [6 ]
机构
[1] Ist Fiorentino Cura & Assistenza IFCA, Lab Serv, Florence, Italy
[2] SOD Sicurezza & Qualita Careggi Hosp, Florence, Italy
[3] IFCA, UO Nefrol & Dialisi Ist Fiorentino Cura & Assiste, Florence, Italy
[4] IFCA, UO Anestesia & Rianimaz Ist Fiorentino Cura & Ass, Florence, Italy
[5] IFCA, Outpatients Serv, Florence, Italy
[6] Univ Verona, Sect Clin Biochem, Verona, Italy
关键词
Performance; measure error; bed-side analysis; potentiometry; SERUM BICARBONATE; METABOLIC-ACIDOSIS; RISK; ASSOCIATION; MORTALITY;
D O I
10.1080/00365513.2021.1930142
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
GEM Premier ChemSTAT is a whole-blood analyzer designed for providing a rapid basic metabolic panel, inclusive of creatinine and blood urea nitrogen, with the unique characteristic of providing measured bicarbonate (HCO3-) levels. The aim of this work was to evaluate the clinical performance of HCO3- assessment with this analyser in a real-life hemodialysis setting. Imprecision was calculated at different HCO3- levels, along with assay comparison with Gem Premier 4000 analysers. GEM Premier ChemSTAT displayed an imprecision and a bias (in comparison to GEM Premier 4000) for HCO3- of 0.4% and 37.3% at 20.8 mmol/L, 1.2% and 25.6% at 16.4 mmol/L, and 2.1% and 11.6% at 11.5 mmol/L, respectively, using three levels of HCO3- quality control sample ChemSTAT System Evaluator. At direct comparison with the GEM Premier 4000 in the hemodialysis setting, Bland-Altman analysis of HCO3- levels evidenced a bias (mu) of -4.9 (95% CI, -5.2 to -4.7) mmol/L. Such difference was attenuated by recalculating the GEM ChemSTAT expected HCO3- values from pH and pCO(2) using the Henderson Hasselbach equation, mu=-0.07 (95%CI, -0.19 to 0.05) mmol/L (p = .24). In conclusion, our results show a remarkable difference between the HCO3- values reported by GEM ChemSTAT or GEM 4000. New reference values for GEM ChemSTAT HCO3- shall hence be defined according to our findings. We suggest that further investigation and a re-evaluation of the reference range should be made before extending the clinical use of this device.
引用
收藏
页码:418 / 421
页数:4
相关论文
共 50 条
  • [41] Clinical evaluation of a new noninvasive method of cardiac output measurement - Preliminary results in CABG patients
    Guzzi, L
    Jaffe, MB
    Orr, JA
    ANESTHESIOLOGY, 1998, 89 (3A) : U409 - U409
  • [42] Evaluation of a new assay for measurement of remnant lipoprotein cholesterol in patients with coronary artery disease.
    Devaraj, S
    Vega, GL
    Grundy, SM
    Jialal, I
    CLINICAL CHEMISTRY, 1997, 43 : 743 - 743
  • [43] FGF23 measurement in burosumab-treated patients: an emerging treatment may induce a new analytical interference
    Piketty, Marie-Liesse
    Brabant, Severine
    Souberbielle, Jean-Claude
    Maruani, Gerard
    Audrain, Christelle
    Rothenbuhler, Anya
    Prie, Dominique
    Linglart, Agnes
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2020, 58 (11) : E267 - E269
  • [44] Measurement of the international normalized ratio (INR) in hemodialysis patients with heparin-locked central venous catheters: evaluation of a novel blood sampling method
    Rioux, Jean-Philippe
    De Bortoli, Bruno
    Querin, Serge
    Deziel, Clement
    Troyanov, Stephan
    Madore, Francois
    JOURNAL OF VASCULAR ACCESS, 2009, 10 (03): : 180 - 182
  • [45] Evaluation of analytical performance and comparison of clinical results of the new generation method AccuTnI+3 for the measurement of cardiac troponin I using both patients and quality control plasma samples
    Storti, Simona
    Masotti, Silvia
    Prontera, Concetta
    Franzini, Maria
    Buzzi, Paola
    Casagranda, Ivo
    Ciofini, Enrica
    Zucchelli, Gian Carlo
    Ndreu, Rudina
    Passino, Claudio
    Clerico, Aldo
    CLINICA CHIMICA ACTA, 2015, 451 : 129 - 134
  • [46] EVALUATION OF TREATMENT THAT ESTABLISHES BLOOD FLUID, ENTERAL ADMINISTRATION OF FE AND NEW TRANSFUSIONAL POLICY IN PATIENTS UNDERGOING CHRONIC-HEMODIALYSIS
    DUCLOS, J
    REVISTA MEDICA DE CHILE, 1983, 111 (11) : 17 - 17
  • [47] 6.8 Application of a New Chemiluminescent Method for Direct Renin Measurement to the Calculation of Aldosterone/Renin Ratio in Patients with Essential Hypertension
    C. Lonati
    A. Gritti
    B. Gidaro
    A. Morganti
    High Blood Pressure & Cardiovascular Prevention, 2008, 15 (3) : 243 - 243
  • [48] Analytical and clinical evaluation of a novel ELISA assay for the measurement of serum PIVKA-II in patients with hepatocellular carcinoma of viral-etiology
    Guariglia, Marta
    Gaia, Silvia
    Rolle, Emanuela
    Abate, Maria L.
    Carucci, Patrizia
    Caviglia, Gian Paolo
    MINERVA BIOTECHNOLOGY AND BIOMOLECULAR RESEARCH, 2024, 36 (04): : 181 - 187
  • [49] MEASUREMENT OF PLATELET-RELEASE PROTEINS (PF4, BETA-TG) IN THE EVALUATION OF A NEW THROMBIN INHIBITOR (MD 805) DURING HEMODIALYSIS
    MATSUO, T
    NAKAO, K
    OKINAGA, Y
    OHOKI, Y
    THROMBOSIS AND HAEMOSTASIS, 1985, 54 (01) : 3 - 3
  • [50] DIRECT QUANTITATIVE MEASUREMENT OF TREMOR - 1ST RESULTS WITH A NEW METHOD OF MEASURING IN PATIENTS UNDER LITHIUM-THERAPY
    ENGEL, RR
    FISCHER, J
    GREIL, W
    ARZNEIMITTEL-FORSCHUNG/DRUG RESEARCH, 1976, 26 (06): : 1126 - 1128