Analytical evaluation of direct bicarbonate measurement with the new gem premier chemstat in hemodialysis patients

被引:0
|
作者
Balboni, Fiamma [1 ]
Terreni, Alessandro [2 ]
Gallo, Marco [3 ]
Guzzi, Francesco [3 ]
Caparrini, Chiara [4 ]
Burbui, Stefania [5 ]
Vezzosi, Marco [1 ]
Galora, Silvia [1 ]
Lori, Giulia [1 ]
Lippi, Giuseppe [6 ]
机构
[1] Ist Fiorentino Cura & Assistenza IFCA, Lab Serv, Florence, Italy
[2] SOD Sicurezza & Qualita Careggi Hosp, Florence, Italy
[3] IFCA, UO Nefrol & Dialisi Ist Fiorentino Cura & Assiste, Florence, Italy
[4] IFCA, UO Anestesia & Rianimaz Ist Fiorentino Cura & Ass, Florence, Italy
[5] IFCA, Outpatients Serv, Florence, Italy
[6] Univ Verona, Sect Clin Biochem, Verona, Italy
关键词
Performance; measure error; bed-side analysis; potentiometry; SERUM BICARBONATE; METABOLIC-ACIDOSIS; RISK; ASSOCIATION; MORTALITY;
D O I
10.1080/00365513.2021.1930142
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
GEM Premier ChemSTAT is a whole-blood analyzer designed for providing a rapid basic metabolic panel, inclusive of creatinine and blood urea nitrogen, with the unique characteristic of providing measured bicarbonate (HCO3-) levels. The aim of this work was to evaluate the clinical performance of HCO3- assessment with this analyser in a real-life hemodialysis setting. Imprecision was calculated at different HCO3- levels, along with assay comparison with Gem Premier 4000 analysers. GEM Premier ChemSTAT displayed an imprecision and a bias (in comparison to GEM Premier 4000) for HCO3- of 0.4% and 37.3% at 20.8 mmol/L, 1.2% and 25.6% at 16.4 mmol/L, and 2.1% and 11.6% at 11.5 mmol/L, respectively, using three levels of HCO3- quality control sample ChemSTAT System Evaluator. At direct comparison with the GEM Premier 4000 in the hemodialysis setting, Bland-Altman analysis of HCO3- levels evidenced a bias (mu) of -4.9 (95% CI, -5.2 to -4.7) mmol/L. Such difference was attenuated by recalculating the GEM ChemSTAT expected HCO3- values from pH and pCO(2) using the Henderson Hasselbach equation, mu=-0.07 (95%CI, -0.19 to 0.05) mmol/L (p = .24). In conclusion, our results show a remarkable difference between the HCO3- values reported by GEM ChemSTAT or GEM 4000. New reference values for GEM ChemSTAT HCO3- shall hence be defined according to our findings. We suggest that further investigation and a re-evaluation of the reference range should be made before extending the clinical use of this device.
引用
收藏
页码:418 / 421
页数:4
相关论文
共 50 条
  • [11] Evaluation of Assays for Measurement of Serum (Anti)oxidants in Hemodialysis Patients
    Ruskovska, Tatjana
    Jansen, Eugene H. J. M.
    Antarorov, Risto
    BIOMED RESEARCH INTERNATIONAL, 2014, 2014
  • [12] Direct measurement of nutrient concentrations in freshwaters with a miniaturized analytical probe: evaluation and validation
    Copetti, D.
    Valsecchi, L.
    Capodaglio, A. G.
    Tartari, G.
    ENVIRONMENTAL MONITORING AND ASSESSMENT, 2017, 189 (04)
  • [13] Direct measurement of nutrient concentrations in freshwaters with a miniaturized analytical probe: evaluation and validation
    D. Copetti
    L. Valsecchi
    A. G. Capodaglio
    G. Tartari
    Environmental Monitoring and Assessment, 2017, 189
  • [14] MEASUREMENT OF DIRECT CURRENTS AND PULSE COMPONENTS FOR ANALYTICAL EVALUATION OF DIFFERENTIAL PULSE POLAROGRAPHY AND VOLTAMMETRY
    ANDERSON, JE
    BOND, AM
    ANALYTICAL CHEMISTRY, 1980, 52 (09) : 1439 - 1445
  • [15] A NEW APPROACH TO THE MEASUREMENT OF THERMAL-ENERGY BALANCE IN HEMODIALYSIS-PATIENTS
    PROVENZANO, R
    SAWAYA, B
    POLASCHEGG, H
    ROY, T
    LEVIN, NW
    KIDNEY INTERNATIONAL, 1988, 33 (01) : 235 - 235
  • [16] A new measurement of energy expenditure and physical activity in patients treated by maintenance hemodialysis
    Deleaval, P.
    Zweiacker, C.
    Bochud, M.
    Burnier, M.
    Teta, D.
    SWISS MEDICAL WEEKLY, 2008, 138 (47-48) : 8S - 9S
  • [17] A New Analytical Method for Direct Measurement of Particulate and Dissolved B-vitamins and Their Congeners in Seawater
    Suffridge, Christopher
    Cutter, Lynda
    Sanudo-Wilhelmy, Sergio A.
    FRONTIERS IN MARINE SCIENCE, 2017, 4
  • [18] Evaluation of Blood Pressure Profile in Chronic Hemodialysis Patients using two Measurement Methods: Home Blood Pressure Measurement and Conventional Measurement
    Faye, Maria
    Faye, Moustapha
    Lemrabott, Ahmed Tall
    Cisse, Mouhamadou Moustapha
    Fall, Kodia
    Keita, Alex Rich Ismael
    Mbengue, Mansour
    Ba, Bacary
    Keita, Niakhaleen
    Diagne, Seynabou
    Niang, Abdou
    Diouf, Boucar
    Ka, El Hadji Fary
    SAUDI JOURNAL OF KIDNEY DISEASES AND TRANSPLANTATION, 2019, 30 (05) : 1038 - 1043
  • [19] Thirty months evaluation of coronary artery calcium progression in patients new to hemodialysis
    Bellasi, Antonio
    Spiegel, David
    Chonchol, Michael
    Ehrlich, James
    Block, Geoffrey
    Raggi, Paolo
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2006, 21 : 448 - 448
  • [20] Analytical evaluation of a new capillary electrophoresis method for carbohydrate-deficient transferrin measurement
    Schellenberg, Francois
    Girre, Catherine
    Nalpas, Bertrand
    Plat, Arnaud
    Tome, Antonio
    Pages, Jean Christophe
    CLINICA CHIMICA ACTA, 2007, 382 (1-2) : 48 - 53