Analytical evaluation of direct bicarbonate measurement with the new gem premier chemstat in hemodialysis patients

被引:0
|
作者
Balboni, Fiamma [1 ]
Terreni, Alessandro [2 ]
Gallo, Marco [3 ]
Guzzi, Francesco [3 ]
Caparrini, Chiara [4 ]
Burbui, Stefania [5 ]
Vezzosi, Marco [1 ]
Galora, Silvia [1 ]
Lori, Giulia [1 ]
Lippi, Giuseppe [6 ]
机构
[1] Ist Fiorentino Cura & Assistenza IFCA, Lab Serv, Florence, Italy
[2] SOD Sicurezza & Qualita Careggi Hosp, Florence, Italy
[3] IFCA, UO Nefrol & Dialisi Ist Fiorentino Cura & Assiste, Florence, Italy
[4] IFCA, UO Anestesia & Rianimaz Ist Fiorentino Cura & Ass, Florence, Italy
[5] IFCA, Outpatients Serv, Florence, Italy
[6] Univ Verona, Sect Clin Biochem, Verona, Italy
关键词
Performance; measure error; bed-side analysis; potentiometry; SERUM BICARBONATE; METABOLIC-ACIDOSIS; RISK; ASSOCIATION; MORTALITY;
D O I
10.1080/00365513.2021.1930142
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
GEM Premier ChemSTAT is a whole-blood analyzer designed for providing a rapid basic metabolic panel, inclusive of creatinine and blood urea nitrogen, with the unique characteristic of providing measured bicarbonate (HCO3-) levels. The aim of this work was to evaluate the clinical performance of HCO3- assessment with this analyser in a real-life hemodialysis setting. Imprecision was calculated at different HCO3- levels, along with assay comparison with Gem Premier 4000 analysers. GEM Premier ChemSTAT displayed an imprecision and a bias (in comparison to GEM Premier 4000) for HCO3- of 0.4% and 37.3% at 20.8 mmol/L, 1.2% and 25.6% at 16.4 mmol/L, and 2.1% and 11.6% at 11.5 mmol/L, respectively, using three levels of HCO3- quality control sample ChemSTAT System Evaluator. At direct comparison with the GEM Premier 4000 in the hemodialysis setting, Bland-Altman analysis of HCO3- levels evidenced a bias (mu) of -4.9 (95% CI, -5.2 to -4.7) mmol/L. Such difference was attenuated by recalculating the GEM ChemSTAT expected HCO3- values from pH and pCO(2) using the Henderson Hasselbach equation, mu=-0.07 (95%CI, -0.19 to 0.05) mmol/L (p = .24). In conclusion, our results show a remarkable difference between the HCO3- values reported by GEM ChemSTAT or GEM 4000. New reference values for GEM ChemSTAT HCO3- shall hence be defined according to our findings. We suggest that further investigation and a re-evaluation of the reference range should be made before extending the clinical use of this device.
引用
收藏
页码:418 / 421
页数:4
相关论文
共 50 条
  • [1] Performance evaluation of a point of care cartridge of the new GEM Premier ChemSTAT analyzer
    Garcia-Fernandez, Alba E.
    Barquin, Raquel
    Martinez, Mariano
    Ferrer, Roser
    Casis, Ernesto
    Xu, Clarke
    PRACTICAL LABORATORY MEDICINE, 2022, 31
  • [2] Analytical performance evaluation of the new GEM® Premier™ 5000 analyzer in comparison to the GEM® Premier™ 4000 and the RapidPoint® 405 systems
    Mion, Monica Maria
    Bragato, Giada
    Zaninotto, Martina
    Alessandroni, Jhessica
    Bernardini, Sergio
    Plebani, Mario
    CLINICA CHIMICA ACTA, 2018, 486 : 313 - 319
  • [3] Analytical Performance Evaluation of the New GEM® Premier™ 5000 in Comparison to the Epoc® Blood Gas Analyzer in Horses
    Sandersen, Charlotte
    Dmitrovic, Petra
    Dupont, Julien
    Cesarini, Carla
    Guyot, Hugues
    Serteyn, Didier
    Kirsch, Katharina
    VETERINARY SCIENCES, 2023, 10 (02)
  • [4] An Evaluation of a Point-of-Care GEM Premier ChemSTAT Analyzer in an Emergency Department: Prevention of Contrast-Induced Nephropathy and Optimization of Patient Flow
    Brailova, Marina
    Audin, Marie
    Raconnat, Julien
    Bouillon-Minois, Jean-Baptiste
    Schmidt, Jeannot
    Pereira, Bruno
    Bouvier, Damien
    Sapin, Vincent
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (23)
  • [5] Evaluation of a new measurement method of indoxyl sulfate in hemodialysis patients
    Abe, Takaya
    Onoda, Mitsutaka
    Matsuura, Tomohiko
    Sugimura, Jun
    Obara, Wataru
    Sasaki, Nariyuki
    Kato, Tetsuo
    Tatsumi, Kenta
    Maruyama, Toru
    THERAPEUTIC APHERESIS AND DIALYSIS, 2021, 25 (01) : 44 - 49
  • [6] New Chemistry Cartridge for the GEM Premier 4000 Critical Care Analyzer, Including Electrochemical Sensors for Creatinine, Urea, and Measured Bicarbonate
    Cervera, Jose
    Conant, Joann
    POINT OF CARE, 2010, 9 (01): : 1 - 7
  • [7] Performance evaluation of a new high volume cartridge for the GEM Premier, 3000 critical care analyzer.
    Ancy, JJ
    D'Orazio, P
    Alexander, A
    Hanford, H
    CLINICAL CHEMISTRY, 2002, 48 (06) : A152 - A152
  • [8] Evaluation of the direct renin inhibitor, aliskiren, in patients on hemodialysis
    Iturralde Iriso, Jesus
    Ocharan Corcuera, Julen
    Mediavilla Tris, Gregorio
    Rodriguez Fernandez, Alfonso
    Ruiz De Gauna, Ramon
    Jimeno, Isabel
    Aurrekoetxea, Begona
    Chena, Alejandro
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2009, 137 : S32 - S32
  • [9] EVALUATION OF PHOSPHOCALCIC PROFILE IN PATIENTS ON CHRONIC HEMODIALYSIS: COMPRATIVE STUDY BETWEEN HEMODIALYSIS BICARBONATE AND ACETATE FREE BIOFILTRATION AT 84%
    Abdelkafi, Wissem
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2015, 30
  • [10] Analytical evaluation of a new immunonephelometric method for homocysteine measurement
    Zappacosta, Bruno
    Persichilli, Silvia
    Minucci, Angelo
    Scribano, Donata
    Antenucci, Mirca
    Fasanella, Silvia
    Neri, Paola
    Giardina, Bruno
    De Sole, Pasquale
    CLINICA CHIMICA ACTA, 2007, 375 (1-2) : 165 - 168