Thermionic sodium ion source versus corona discharge in detection of alkaloids using ion mobility spectrometry

被引:4
|
作者
Asadi, Masood [1 ]
Valadbeigi, Younes [2 ]
Tabrizchi, Mahmoud [1 ]
机构
[1] Isfahan Univ Technol, Dept Chem, Esfahan 8415683111, Iran
[2] Imam Khomeini Int Univ, Fac Sci, Dept Chem, POB 288, Qazvin, Iran
关键词
Thermal ionization; Ion mobility spectrometry; Sodium ion; Alkaloids; Ionization source; IONIZATION SOURCE;
D O I
10.1007/s12127-019-00249-9
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
IMS spectra obtained by using a thermal ionization sodium ion (SI) source and a corona discharge (CD) source were compared in terms of resolution, selectivity and sensitivity. The reactant ion peak (RIP) for the SI source is a single narrow peak which appears in a shorter drift time compared to the main RIP of the CD ion source, (H2O)(n)H3O+. Comparison of the IMS spectra of selected alkaloids obtained with both the CD and the SI sources shows that the latter exhibits simpler pattern, including a narrow single peak for each compound, well separated from other samples peak. Furthermore, the product ion peaks (PIP) in the SI source appeared at higher drift times leading to higher peak-to-peak resolution. However, the sodium ion source is less sensitive to the selected alkaloids compared to the CD ion source.
引用
收藏
页码:51 / 58
页数:8
相关论文
共 50 条
  • [11] Corona discharge ion mobility spectrometry at reduced pressures
    Tabrizchi, M
    Rouholahnejad, F
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (11): : 4656 - 4661
  • [12] Corona Discharge Ion Mobility Spectrometry of Ten Alcohols
    Han, Hai-yan
    Wang, Hong-mei
    Jiang, Hai-he
    Stano, Michal
    Sabo, Martin
    Matejcik, Stefan
    Chu, Yan-nan
    CHINESE JOURNAL OF CHEMICAL PHYSICS, 2009, 22 (06) : 605 - 610
  • [13] Detection of Volatile Organic Compound with an Ion Mobility Spectrometry Cell Type Device with a Corona Discharge Ionization Source
    Boggio, N. G.
    Pierpauli, K.
    Rinaldi, C. A.
    Lamagna, A.
    SENSOR LETTERS, 2011, 9 (02) : 866 - 868
  • [14] Towards metals analysis using corona discharge ionization ion mobility spectrometry
    Jafari, Mohammad T.
    Saraji, Mohammad
    Sherafatmand, Hossein
    ANALYTICA CHIMICA ACTA, 2016, 909 : 84 - 90
  • [15] Using corona discharge-ion mobility spectrometry for detection of 2,4,6-Trichloroanisole
    Lichvanova, Zuzana
    Ilbeigi, Vahideh
    Sabo, Martin
    Tabrizchi, Mahmoud
    Matejcik, Stefan
    TALANTA, 2014, 127 : 239 - 243
  • [16] Corona discharge ion mobility spectrometry of aliphatic and aromatic hydrocarbons
    Borsdorf, H
    Schelhorn, H
    Flachowsky, J
    Döring, HR
    Stach, J
    ANALYTICA CHIMICA ACTA, 2000, 403 (1-2) : 235 - 242
  • [17] Direct Liquid Sampling for Corona Discharge Ion Mobility Spectrometry
    Sabo, Martin
    Malaskova, Michaela
    Harmathova, Olga
    Hradski, Jasna
    Masar, Marian
    Radjenovic, Branislav
    Matejcik, Stefan
    ANALYTICAL CHEMISTRY, 2015, 87 (14) : 7389 - 7394
  • [18] Laser desorption with corona discharge ion mobility spectrometry for direct surface detection of explosives
    Sabo, M.
    Malaskova, M.
    Matejcik, S.
    ANALYST, 2014, 139 (20) : 5112 - 5117
  • [19] Comparison of the positive and negative modes of corona discharge ion source for direct determination of aspirin in urine by ion mobility spectrometry
    Valadbeigi, Younes
    Ilbeigi, Vahideh
    Afgar, Azizulhaq
    Soleimani, Majid
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2021, 470
  • [20] Direct analysis of human breath ammonia using corona discharge ion mobility spectrometry
    Jazan, Elham
    Mirzaei, Hadi
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2014, 88 : 315 - 320