Effect of Automatic Tube Current Modulation on Radiation Dose and Image Quality for Low Tube Voltage Multidetector Row CT Angiography: Phantom Study

被引:46
|
作者
Schindera, Sebastian T. [1 ]
Nelson, Rendon C. [2 ]
Yoshizumi, Terry [3 ]
Toncheva, Greta [3 ]
Nguyen, Giao [3 ]
DeLong, David M. [2 ]
Szucs-Farkas, Zsolt [1 ]
机构
[1] Univ Bern, Univ Hosp Bern, Inselspital, Dept Diagnost Intervent & Pediat Radiol, CH-3010 Bern, Switzerland
[2] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[3] Duke Univ, Med Ctr, Div Radiat Safety, Durham, NC USA
关键词
CT angiography; automatic tube current modulation; low tube voltage; radiation dose; image quality; COMPUTED-TOMOGRAPHY; PULMONARY-EMBOLISM; CONTRAST MATERIAL; PATIENT SIZE; REDUCTION; SETTINGS; 16-MDCT; ENHANCEMENT; PROTOCOL; WEIGHT;
D O I
10.1016/j.acra.2009.02.021
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. To evaluate the effect of automatic tube current modulation on radiation dose and image quality for low tube voltage computed tomography (CT) angiography. Materials and Methods. An anthropomorphic phantom was scanned with a 64-section CT scanner using following tube voltages: 140 kVp (Protocol A), 120 kVp (Protocol B), 100 kVp (Protocol C), and 80 kVp (Protocol D). To achieve similar noise, combined z-axis and xy-axes automatic tube current modulation was applied. Effective dose (ED) for the four tube voltages was assessed. Three plastic vials filled with different concentrations of iodinated solution were placed on the phantom's abdomen to obtain attenuation measurements. The signal-to-noise ratio (SNR) was calculated and a figure of merit (FOM) for each iodinated solution was computed as SNR2/ED. Results. The ED was kept similar for the four different tube voltages: (A) 5.4 mSv +/- 0.3, (B) 4.1 mSv +/- 0.6, (C) 3.9 mSv +/- 0.5, and (D) 4.2 mSv +/- 0.3 (P >.05). As the tube voltage decreased from 140 to 80 kVp, image noise was maintained (range, 13.8-14.9 HU) (P >.05). SNR increased as the tube voltage decreased, with an overall gain of 119% for the 80-kVp compared to the 140-kVp protocol (P < .05). The FOM results indicated that with a reduction of the tube voltage from 140 to 120, 100, and 80 kVp, at constant SNR, ED was reduced by a factor of 2.1, 3.3, and 5.1, respectively, (P < .001). Conclusions. As tube voltage decreases, automatic tube current modulation for CT angiography yields either a significant quality. increase in image quality at constant radiation dose or a significant decrease in radiation dose at a constant image quality.
引用
收藏
页码:997 / 1002
页数:6
相关论文
共 50 条
  • [31] Radiation dose reduction and image quality in pediatric paranasal sinus CT: with automatic tube current modulation and iterative reconstruction technique
    Chen, Xiao-Fang
    Zhang, Yi-Chi
    Ding, Ning
    Liu, Li
    Ji, Yi-Ding
    Zhang, Chao
    Chi, Jing
    RADIATION PROTECTION DOSIMETRY, 2024, 200 (15) : 1470 - 1476
  • [32] Effect of low tube kV on radiation dose and image quality in retrospective ECG-gated coronary CT angiography
    Abou-Issa, Ahmed H.
    Elganayni, Fady
    Al-Azzazy, Mohammed Zakaria
    EGYPTIAN JOURNAL OF RADIOLOGY AND NUCLEAR MEDICINE, 2011, 42 (3-4): : 327 - 333
  • [33] Iterative Reconstruction Algorithm for Abdominal Multidetector CT at Different Tube Voltages: Assessment of Diagnostic Accuracy, Image Quality, and Radiation Dose in a Phantom Study
    Schindera, Sebastian T.
    Diedrichsen, Lars
    Mueller, Hubert C.
    Rusch, Oliver
    Marin, Daniele
    Schmidt, Bernhard
    Raupach, Rainer
    Vock, Peter
    Szucs-Farkas, Zsolt
    RADIOLOGY, 2011, 260 (02) : 454 - 462
  • [34] Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine Intake
    Pan, Yue-ying
    Zhou, Shu-chang
    Wang, Yu-jin
    Li, Qian
    Zhu, Ting-ting
    Liu, Chun-xia
    Guan, Han-xiong
    CURRENT MEDICAL SCIENCE, 2020, 40 (01) : 178 - 183
  • [35] Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine Intake
    Yue-ying Pan
    Shu-chang Zhou
    Yu-jin Wang
    Qian Li
    Ting-ting Zhu
    Chun-xia Liu
    Han-xiong Guan
    Current Medical Science, 2020, 40 : 178 - 183
  • [36] New technique to avoid an increase in CT radiation dose from metallic prostheses with automatic tube current modulation: Phantom and patient study
    Dalal, T
    Rizzo, SMR
    Kalra, MK
    Schmidt, B
    Suess, C
    Flohr, T
    Blake, MA
    Saini, S
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2005, 184 (04) : 51 - 51
  • [37] Effect of X-ray Tube Parameters and Iodine Concentration on Image Quality and Radiation Dose in Cerebral Pediatric and Adult CT Angiography A Phantom Study
    Papadakis, Antonios E.
    Perisinakis, Kostas
    Raissaki, Maria
    Damilakis, John
    INVESTIGATIVE RADIOLOGY, 2013, 48 (04) : 192 - 199
  • [38] Coronary CT angiography by modifying tube voltage and contrast medium concentration: Evaluation of image quality and radiation dose
    Gazafroudi, Seyede Shokoofeh Mousavi
    Tavakkoli, Mohammad Bagher
    Moradi, Maryam
    Gazafroudi, Seyede Shabnam Mousavi
    Yadegarfar, Ghasem
    Behjati, Mohaddeseh
    Karimian, Mohammad
    Khajouei, Amirreza Sajjadieh
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2019, 36 (07): : 1391 - 1396
  • [39] Hypervascular liver tumors: Low tube voltage, high tube current multi-detector row CT for enhanced detection - Phantom study
    Schindera, Sebastian T.
    Nelson, Rendon C.
    Mukundan, Srinivasan, Jr.
    Paulson, Erik K.
    Jaffe, Tracy A.
    Miller, Chad M.
    DeLong, David M.
    Kawaji, Keigo
    Yoshizumi, Terry T.
    Samei, Ehsan
    RADIOLOGY, 2008, 246 (01) : 125 - 132
  • [40] Achieving consistent image quality and overall radiation dose reduction for coronary CT angiography with body mass index-dependent tube voltage and tube current selection
    Wang, G.
    Gao, J.
    Zhao, S.
    Sun, X.
    Chen, X.
    Cui, X.
    CLINICAL RADIOLOGY, 2014, 69 (09) : 945 - 951