Effect of X-ray Tube Parameters and Iodine Concentration on Image Quality and Radiation Dose in Cerebral Pediatric and Adult CT Angiography A Phantom Study

被引:22
|
作者
Papadakis, Antonios E. [1 ]
Perisinakis, Kostas [2 ]
Raissaki, Maria [3 ]
Damilakis, John [2 ]
机构
[1] Univ Hosp Herakl, Dept Med Phys, Iraklion 71110, Greece
[2] Univ Crete, Dept Med Phys, Iraklion, Greece
[3] Univ Crete, Dept Radiol, Iraklion, Greece
关键词
CT angiography; automatic exposure control; imaging optimization; radiation protection; COMPUTED-TOMOGRAPHY-ANGIOGRAPHY; AUTOMATIC EXPOSURE CONTROL; CURRENT MODULATION; SCANNING PARAMETERS; RATIONAL APPROACH; 80; KVP; REDUCTION; CONTRAST; PROTOCOLS; VOLTAGE;
D O I
10.1097/RLI.0b013e31827efc17
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The aim of the present phantom study was to investigate the effect of x-ray tube parameters and iodine concentration on image quality and radiation dose in cerebral computed tomographic (CT) angiographic examinations of pediatric and adult individuals. Materials and Methods: Four physical anthropomorphic phantoms that represent the average individual as neonate, 1-year-old, 5-year-old, and 10-year-old children and the RANDO phantom that simulates the average adult individual were used. Cylindrical vessels were bored along the brain-equivalent plugs of each physical phantom. To simulate the brain vasculature, vessels of 0.6, 1, 2, and 3 mm in diameter were created. These vessels were filled with contrast medium (CM) solutions at different iodine concentrations, that is, 5.6, 4.2, 2.7, and 1.4 mg I/mL. The phantom heads were scanned at 120, 100, and 80 kV. The applied quality reference tube current-time product values ranged from a minimum of 45 to a maximum of 680. The CT acquisitions were performed on a 16-slice CT scanner using the automatic exposure control system. Image quality was evaluated on the basis of image noise and contrast-to-noise ratio (CNR) between the contrast-enhanced iodinated vessels and the unenhanced regions of interest. Dose reduction was calculated as the percentage difference of the CT dose index value at the quality reference tube current-time product and the CT dose index at the mean modulated tube current-time product. Results: Image noise that was measured using the preset tube current-time product settings varied significantly among the different phantoms (P < 0.0001). Hounsfield unit number of iodinated vessels was linearly related to CM concentration (r(2) = 0.907) and vessel diameter (r(2) = 0.918). The Hounsfield unit number of iodinated vessels followed a decreasing trend from the neonate phantom to the adult phantom at all kilovoltage settings. For the same image noise level, a CNR improvement of up to 69% and a dose reduction of up to 61% may be achieved when CT acquisition is performed at 80 kV compared with 120 kV. For the same CNR, a reduction by 25% of the administered CM concentration may be achieved when CTacquisition is performed at 80 kV compared with 120 kV. Conclusions: In cerebral CT angiographic studies, appropriate adjustment of the preset tube current-time product settings is required to achieve the same image noise level among participants of different age. Cerebral CT angiography at 80 kV significantly improves CNR and significantly reduces radiation dose. Moreover, at 80 kV, a considerable reduction of the administered amount of the CM may be reached, thus reducing potential risks for contrast-induced nephropathy.
引用
收藏
页码:192 / 199
页数:8
相关论文
共 50 条
  • [1] Effect of x-ray tube parameters, iodine concentration, and patient size on image quality in pulmonary computed tomography angiography: A chest-phantom-study
    Szucs-Farkas, Zsolt
    Verdun, Francis R.
    von Allmen, Gabriel
    Mini, Roberto L.
    Vock, Peter
    INVESTIGATIVE RADIOLOGY, 2008, 43 (06) : 374 - 381
  • [2] X-ray tube voltage and image quality in adult and pediatric CT
    Huda, W.
    Ogden, K. M.
    Scalzetti, E. M.
    Lavallee, R. L.
    Samei, E.
    MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
  • [3] Image Quality and Radiation Dose in Propagation Based Phase Contrast Mammography with a Microfocus X-ray Tube: A Phantom Study
    Castriconi, Roberta
    Mettivier, Giovanni
    Russo, Paolo
    BREAST IMAGING, IWDM 2016, 2016, 9699 : 618 - 624
  • [4] Effect of Automatic Tube Current Modulation on Radiation Dose and Image Quality for Low Tube Voltage Multidetector Row CT Angiography: Phantom Study
    Schindera, Sebastian T.
    Nelson, Rendon C.
    Yoshizumi, Terry
    Toncheva, Greta
    Nguyen, Giao
    DeLong, David M.
    Szucs-Farkas, Zsolt
    ACADEMIC RADIOLOGY, 2009, 16 (08) : 997 - 1002
  • [5] Pediatric thoracic CT angiography at 70 kV: a phantom study to investigate the effects on image quality and radiation dose
    Robert D. MacDougall
    Patricia L. Kleinman
    Lifeng Yu
    Edward Y. Lee
    Pediatric Radiology, 2016, 46 : 1114 - 1119
  • [6] Pediatric thoracic CT angiography at 70 kV: a phantom study to investigate the effects on image quality and radiation dose
    MacDougall, Robert D.
    Kleinman, Patricia L.
    Yu, Lifeng
    Lee, Edward Y.
    PEDIATRIC RADIOLOGY, 2016, 46 (08) : 1114 - 1119
  • [7] Effect of Tube Voltage and Radiation Dose on Image Quality in Pediatric Abdominal CT Using Deep Learning Reconstruction: A Phantom Study
    Kim, Daehong
    Jeon, Pil-Hyun
    Lee, Chang-Lae
    Chung, Myung-Ae
    SYMMETRY-BASEL, 2023, 15 (02):
  • [8] Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: A phantom study
    Schindera, S. T.
    Winklehner, A.
    Alkadhi, H.
    Goetti, R.
    Fischer, M.
    Gnannt, R.
    Szucs-Farkas, Z.
    CLINICAL RADIOLOGY, 2013, 68 (02) : E79 - E86
  • [9] Effect of low tube voltage and low iodine concentration abdominal CT on image quality and radiation dose in children: preliminary study
    You, Sun Kyoung
    Choi, Young Hun
    Cheon, Jung-Eun
    Kim, Woo Sun
    Kim, In-One
    Lee, So Mi
    Cho, Hyun-Hae
    ABDOMINAL RADIOLOGY, 2019, 44 (05) : 1928 - 1935
  • [10] Effect of low tube voltage and low iodine concentration abdominal CT on image quality and radiation dose in children: preliminary study
    Sun Kyoung You
    Young Hun Choi
    Jung-Eun Cheon
    Woo Sun Kim
    In-One Kim
    So Mi Lee
    Hyun-Hae Cho
    Abdominal Radiology, 2019, 44 : 1928 - 1935