Effect of Automatic Tube Current Modulation on Radiation Dose and Image Quality for Low Tube Voltage Multidetector Row CT Angiography: Phantom Study

被引:46
|
作者
Schindera, Sebastian T. [1 ]
Nelson, Rendon C. [2 ]
Yoshizumi, Terry [3 ]
Toncheva, Greta [3 ]
Nguyen, Giao [3 ]
DeLong, David M. [2 ]
Szucs-Farkas, Zsolt [1 ]
机构
[1] Univ Bern, Univ Hosp Bern, Inselspital, Dept Diagnost Intervent & Pediat Radiol, CH-3010 Bern, Switzerland
[2] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[3] Duke Univ, Med Ctr, Div Radiat Safety, Durham, NC USA
关键词
CT angiography; automatic tube current modulation; low tube voltage; radiation dose; image quality; COMPUTED-TOMOGRAPHY; PULMONARY-EMBOLISM; CONTRAST MATERIAL; PATIENT SIZE; REDUCTION; SETTINGS; 16-MDCT; ENHANCEMENT; PROTOCOL; WEIGHT;
D O I
10.1016/j.acra.2009.02.021
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. To evaluate the effect of automatic tube current modulation on radiation dose and image quality for low tube voltage computed tomography (CT) angiography. Materials and Methods. An anthropomorphic phantom was scanned with a 64-section CT scanner using following tube voltages: 140 kVp (Protocol A), 120 kVp (Protocol B), 100 kVp (Protocol C), and 80 kVp (Protocol D). To achieve similar noise, combined z-axis and xy-axes automatic tube current modulation was applied. Effective dose (ED) for the four tube voltages was assessed. Three plastic vials filled with different concentrations of iodinated solution were placed on the phantom's abdomen to obtain attenuation measurements. The signal-to-noise ratio (SNR) was calculated and a figure of merit (FOM) for each iodinated solution was computed as SNR2/ED. Results. The ED was kept similar for the four different tube voltages: (A) 5.4 mSv +/- 0.3, (B) 4.1 mSv +/- 0.6, (C) 3.9 mSv +/- 0.5, and (D) 4.2 mSv +/- 0.3 (P >.05). As the tube voltage decreased from 140 to 80 kVp, image noise was maintained (range, 13.8-14.9 HU) (P >.05). SNR increased as the tube voltage decreased, with an overall gain of 119% for the 80-kVp compared to the 140-kVp protocol (P < .05). The FOM results indicated that with a reduction of the tube voltage from 140 to 120, 100, and 80 kVp, at constant SNR, ED was reduced by a factor of 2.1, 3.3, and 5.1, respectively, (P < .001). Conclusions. As tube voltage decreases, automatic tube current modulation for CT angiography yields either a significant quality. increase in image quality at constant radiation dose or a significant decrease in radiation dose at a constant image quality.
引用
收藏
页码:997 / 1002
页数:6
相关论文
共 50 条
  • [1] Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography A Phantom Study on Radiation Dose and Image Quality
    Papadalds, Antonios E.
    Damilakis, John
    INVESTIGATIVE RADIOLOGY, 2019, 54 (05) : 265 - 272
  • [2] Effect of Low Tube Voltage on Image Quality, Radiation Dose, and Low-Contrast Detectability at Abdominal Multidetector CT: Phantom Study
    Tang, Kun
    Wang, Ling
    Li, Rui
    Lin, Jie
    Zheng, Xiangwu
    Cao, Guoquan
    JOURNAL OF BIOMEDICINE AND BIOTECHNOLOGY, 2012,
  • [3] Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: A phantom study
    Schindera, S. T.
    Winklehner, A.
    Alkadhi, H.
    Goetti, R.
    Fischer, M.
    Gnannt, R.
    Szucs-Farkas, Z.
    CLINICAL RADIOLOGY, 2013, 68 (02) : E79 - E86
  • [4] Comparison of Image Quality and Radiation Dose between Fixed Tube Current and Combined Automatic Tube Current Modulation in Craniocervical CT Angiography
    Lee, E. J.
    Lee, S. K.
    Agid, R.
    Howard, P.
    Bae, J. M.
    terBrugge, K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (09) : 1754 - 1759
  • [5] The application of automatic tube current modulation (ATCM) on image quality and radiation dose at abdominal computed tomography (CT): A phantom study
    Wang, Qian
    Zhao, Xinming
    Song, Junfeng
    Guo, Ning
    Zhu, Ying
    Liu, Jianxin
    Qi, Weiwei
    Wu, Jing
    Liang, Yuan
    Feng, Shichao
    Hu, Mancang
    Zhou, Chunwu
    Wang, Xiaoying
    Hong, Nan
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2013, 21 (04) : 453 - 464
  • [6] Effective dose and image quality evaluations of an automatic CT tube current modulation system with an anthropomorphic phantom
    Rampado, O.
    Marchisio, F.
    Izzo, A.
    Garelli, E.
    Bianchi, C. C.
    Gandini, G.
    Ropolo, R.
    EUROPEAN JOURNAL OF RADIOLOGY, 2009, 72 (01) : 181 - 187
  • [7] Effect of patient size on radiation dose for abdominal MDCT with automatic tube current modulation: Phantom study
    Schindera, Sebastian T.
    Nelson, Rendon C.
    Toth, Thomas L.
    Nguyen, Giao T.
    Toncheva, Greta I.
    DeLong, David M.
    Yoshizumi, Terry T.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (02) : W100 - W105
  • [8] Evaluation of the effect of low tube voltage on radiation dose and image quality
    Din, Nik Norhasrina Nik
    Zainon, Rafidah
    Rahman, A. T. Abdul
    10TH INTERNATIONAL SEMINAR ON MEDICAL PHYSICS & 1ST AMDI INTERNATIONAL ONCOLOGY SYMPOSIUM, 2017, 851
  • [9] Comparison of image quality and radiation dose between combined automatic tube current modulation and fixed tube current technique in CT of abdomen and pelvis
    Lee, Sanghee
    Yoon, Sang-Wook
    Yoo, Seung-Min
    Ji, Young Geon
    Kim, Kyoung Ah
    Kim, Sang Heum
    Lee, Jong Tae
    ACTA RADIOLOGICA, 2011, 52 (10) : 1101 - 1106
  • [10] Effect of Tube Voltage and Radiation Dose on Image Quality in Pediatric Abdominal CT Using Deep Learning Reconstruction: A Phantom Study
    Kim, Daehong
    Jeon, Pil-Hyun
    Lee, Chang-Lae
    Chung, Myung-Ae
    SYMMETRY-BASEL, 2023, 15 (02):