Effect of Automatic Tube Current Modulation on Radiation Dose and Image Quality for Low Tube Voltage Multidetector Row CT Angiography: Phantom Study

被引:46
|
作者
Schindera, Sebastian T. [1 ]
Nelson, Rendon C. [2 ]
Yoshizumi, Terry [3 ]
Toncheva, Greta [3 ]
Nguyen, Giao [3 ]
DeLong, David M. [2 ]
Szucs-Farkas, Zsolt [1 ]
机构
[1] Univ Bern, Univ Hosp Bern, Inselspital, Dept Diagnost Intervent & Pediat Radiol, CH-3010 Bern, Switzerland
[2] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[3] Duke Univ, Med Ctr, Div Radiat Safety, Durham, NC USA
关键词
CT angiography; automatic tube current modulation; low tube voltage; radiation dose; image quality; COMPUTED-TOMOGRAPHY; PULMONARY-EMBOLISM; CONTRAST MATERIAL; PATIENT SIZE; REDUCTION; SETTINGS; 16-MDCT; ENHANCEMENT; PROTOCOL; WEIGHT;
D O I
10.1016/j.acra.2009.02.021
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. To evaluate the effect of automatic tube current modulation on radiation dose and image quality for low tube voltage computed tomography (CT) angiography. Materials and Methods. An anthropomorphic phantom was scanned with a 64-section CT scanner using following tube voltages: 140 kVp (Protocol A), 120 kVp (Protocol B), 100 kVp (Protocol C), and 80 kVp (Protocol D). To achieve similar noise, combined z-axis and xy-axes automatic tube current modulation was applied. Effective dose (ED) for the four tube voltages was assessed. Three plastic vials filled with different concentrations of iodinated solution were placed on the phantom's abdomen to obtain attenuation measurements. The signal-to-noise ratio (SNR) was calculated and a figure of merit (FOM) for each iodinated solution was computed as SNR2/ED. Results. The ED was kept similar for the four different tube voltages: (A) 5.4 mSv +/- 0.3, (B) 4.1 mSv +/- 0.6, (C) 3.9 mSv +/- 0.5, and (D) 4.2 mSv +/- 0.3 (P >.05). As the tube voltage decreased from 140 to 80 kVp, image noise was maintained (range, 13.8-14.9 HU) (P >.05). SNR increased as the tube voltage decreased, with an overall gain of 119% for the 80-kVp compared to the 140-kVp protocol (P < .05). The FOM results indicated that with a reduction of the tube voltage from 140 to 120, 100, and 80 kVp, at constant SNR, ED was reduced by a factor of 2.1, 3.3, and 5.1, respectively, (P < .001). Conclusions. As tube voltage decreases, automatic tube current modulation for CT angiography yields either a significant quality. increase in image quality at constant radiation dose or a significant decrease in radiation dose at a constant image quality.
引用
收藏
页码:997 / 1002
页数:6
相关论文
共 50 条
  • [41] Automatic tube potential selection with tube current modulation in coronary CT angiography: Can it achieve consistent image quality among various individuals?
    Wang, Xiao-Ping
    Zhu, Xiao-Mei
    Zhu, Yin-Su
    Liu, Wang-Yan
    Yang, Xiao-Han
    Huang, Wei-Wei
    Xu, Yi
    Tang, Li-Jun
    EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2018, 16 (01) : 253 - 259
  • [42] Impact of the automatic tube current modulation (ATCM) system on virtual monoenergetic image quality for dual-source CT: A phantom study
    Dabli, D.
    Durand, Quentin
    Frandon, J.
    de Oliveira, Fabien
    Pastor, Maxime
    Beregi, J. P.
    Greffier, J.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 109
  • [43] Radiation dose reduction in thoracic and abdomen-pelvic CT using tube current modulation: a phantom study
    Sabarudin, Akmal
    Mustafa, Zakira
    Nassir, Khadijah Mohd
    Hamid, Hamzaini Abdul
    Sun, Zhonghua
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (01): : 319 - 328
  • [44] Coronary Computed Tomography Angiography With Very Low Tube Voltage Reduces Radiation Dose and Maintains Image Quality
    Daubert, Melissa A.
    Goldman, Mark
    Chen, John
    Feldmann, Eric
    Ferraro, Summer
    Hellinger, Jeffrey
    Poon, Michael
    CIRCULATION, 2011, 124 (21)
  • [45] Automatic tube current modulation for whole-body polytrauma CT with immobilization devices: is there an increase in radiation dose and degradation of image quality?
    Euler A.
    Stieltjes B.
    Schindera S.T.
    Emergency Radiology, 2017, 24 (1) : 31 - 37
  • [46] Effect of tube current on diagnostic image quality in paediatric cerebral multidetector CT images
    Ledenius, K.
    Gustavsson, M.
    Johansson, S.
    Stalhammar, F.
    Wiklund, L-M
    Thilander-Klang, A.
    BRITISH JOURNAL OF RADIOLOGY, 2009, 82 (976): : 313 - 320
  • [47] The influences of tube voltage and scan direction on combined tube current modulation: a phantom study
    Goo, Hyun Woo
    Suh, Dong Soo
    PEDIATRIC RADIOLOGY, 2006, 36 (08) : 833 - 840
  • [48] 320-row CT coronary angiography: effect of 100-kV tube voltages on image quality, contrast volume, and radiation dose
    Zhang, Chuanchen
    Zhang, Zhaoqi
    Yan, Zixu
    Xu, Lei
    Yu, Wei
    Wang, Rui
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2011, 27 (07): : 1059 - 1068
  • [49] The influences of tube voltage and scan direction on combined tube current modulation: a phantom study
    Hyun Woo Goo
    Dong Soo Suh
    Pediatric Radiology, 2006, 36 : 833 - 840
  • [50] 320-row CT coronary angiography: effect of 100-kV tube voltages on image quality, contrast volume, and radiation dose
    Chuanchen Zhang
    Zhaoqi Zhang
    Zixu Yan
    Lei Xu
    Wei Yu
    Rui Wang
    The International Journal of Cardiovascular Imaging, 2011, 27 : 1059 - 1068