Effect of X-ray Tube Parameters and Iodine Concentration on Image Quality and Radiation Dose in Cerebral Pediatric and Adult CT Angiography A Phantom Study

被引:22
|
作者
Papadakis, Antonios E. [1 ]
Perisinakis, Kostas [2 ]
Raissaki, Maria [3 ]
Damilakis, John [2 ]
机构
[1] Univ Hosp Herakl, Dept Med Phys, Iraklion 71110, Greece
[2] Univ Crete, Dept Med Phys, Iraklion, Greece
[3] Univ Crete, Dept Radiol, Iraklion, Greece
关键词
CT angiography; automatic exposure control; imaging optimization; radiation protection; COMPUTED-TOMOGRAPHY-ANGIOGRAPHY; AUTOMATIC EXPOSURE CONTROL; CURRENT MODULATION; SCANNING PARAMETERS; RATIONAL APPROACH; 80; KVP; REDUCTION; CONTRAST; PROTOCOLS; VOLTAGE;
D O I
10.1097/RLI.0b013e31827efc17
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The aim of the present phantom study was to investigate the effect of x-ray tube parameters and iodine concentration on image quality and radiation dose in cerebral computed tomographic (CT) angiographic examinations of pediatric and adult individuals. Materials and Methods: Four physical anthropomorphic phantoms that represent the average individual as neonate, 1-year-old, 5-year-old, and 10-year-old children and the RANDO phantom that simulates the average adult individual were used. Cylindrical vessels were bored along the brain-equivalent plugs of each physical phantom. To simulate the brain vasculature, vessels of 0.6, 1, 2, and 3 mm in diameter were created. These vessels were filled with contrast medium (CM) solutions at different iodine concentrations, that is, 5.6, 4.2, 2.7, and 1.4 mg I/mL. The phantom heads were scanned at 120, 100, and 80 kV. The applied quality reference tube current-time product values ranged from a minimum of 45 to a maximum of 680. The CT acquisitions were performed on a 16-slice CT scanner using the automatic exposure control system. Image quality was evaluated on the basis of image noise and contrast-to-noise ratio (CNR) between the contrast-enhanced iodinated vessels and the unenhanced regions of interest. Dose reduction was calculated as the percentage difference of the CT dose index value at the quality reference tube current-time product and the CT dose index at the mean modulated tube current-time product. Results: Image noise that was measured using the preset tube current-time product settings varied significantly among the different phantoms (P < 0.0001). Hounsfield unit number of iodinated vessels was linearly related to CM concentration (r(2) = 0.907) and vessel diameter (r(2) = 0.918). The Hounsfield unit number of iodinated vessels followed a decreasing trend from the neonate phantom to the adult phantom at all kilovoltage settings. For the same image noise level, a CNR improvement of up to 69% and a dose reduction of up to 61% may be achieved when CT acquisition is performed at 80 kV compared with 120 kV. For the same CNR, a reduction by 25% of the administered CM concentration may be achieved when CTacquisition is performed at 80 kV compared with 120 kV. Conclusions: In cerebral CT angiographic studies, appropriate adjustment of the preset tube current-time product settings is required to achieve the same image noise level among participants of different age. Cerebral CT angiography at 80 kV significantly improves CNR and significantly reduces radiation dose. Moreover, at 80 kV, a considerable reduction of the administered amount of the CM may be reached, thus reducing potential risks for contrast-induced nephropathy.
引用
收藏
页码:192 / 199
页数:8
相关论文
共 50 条
  • [31] Estimates of the image quality in accordance with radiation dose for pediatric imaging using deep learning CT: A phantom study
    Jeon, Pil-Hyun
    Kim, Daehong
    Chung, Myung-Ae
    2022 IEEE INTERNATIONAL CONFERENCE ON BIG DATA AND SMART COMPUTING (IEEE BIGCOMP 2022), 2022, : 352 - 356
  • [32] Effect of CT acquisition parameters of spiral CT on image quality and radiation dose
    Manmadhachary, A.
    Kumar, Y. Ravi
    Krishnanand, L.
    MEASUREMENT, 2017, 103 : 18 - 26
  • [33] Gridless adult cervical spine radiography and its' effect on image quality and radiation dose: A phantom study
    Mekis, N.
    Bianchi, T.
    Doyle, C.
    Gauchat, M.
    Geerling, I.
    Linneman, J.
    Staats, S.
    Campeanu, C.
    RADIOGRAPHY, 2024, 30 (01) : 359 - 366
  • [34] Optimization of Image Quality and Organ Absorbed Dose for Pediatric Chest X-Ray Examination: In-House Developed Chest Phantom Study
    Pengpan, Thanyawee
    Rattanarungruangchai, Natch
    Dechjaithat, Juthathip
    Panthim, Phawinee
    Siricharuwong, Puntarika
    Prapan, Ausanai
    RADIOLOGY RESEARCH AND PRACTICE, 2022, 2022
  • [35] Effect of iterative reconstruction techniques on image quality in low radiation dose chest CT: a phantom study
    Xu, Yan
    Zhang, Ting-ting
    Hu, Zhi-hai
    Li, Juan
    Hou, Hong-jun
    Xu, Zu-shan
    He, Wen
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2019, 25 (06) : 442 - 450
  • [36] Effect of x-ray tube current on the accuracy of cerebral perfusion parameters obtained by CT perfusion studies
    Murase, K
    Nanjo, T
    Ii, S
    Miyazaki, S
    Hirata, M
    Sugawara, Y
    Kudo, M
    Sasaki, K
    Mochizuki, T
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (21): : 5019 - 5029
  • [37] Paediatric x-ray radiation dose reduction and image quality analysis
    Martin, L.
    Ruddlesden, R.
    Makepeace, C.
    Robinson, L.
    Mistry, T.
    Starritt, H.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2013, 33 (03) : 621 - 633
  • [38] Investigation of Radiation Dose and Image Quality in X-Ray Radiographic Imaging
    Zainon, Rafidah
    Abu Talib, Nor Syazreena
    Abu Bakar, Siti Nur Amira
    Moner, Nur Hanisah Mohd
    Aziz, Nurul Athirah Abdul
    TRANSACTIONS ON ENGINEERING TECHNOLOGIES, 2015, : 361 - 368
  • [39] Infant Cardiac CT Angiography with 64-Slice and 256-Slice CT: Comparison of Radiation Dose and Image Quality Using a Pediatric Phantom
    Lee, Yi-Wei
    Yang, Ching-Ching
    Mok, Greta S. P.
    Wu, Tung-Hsin
    PLOS ONE, 2012, 7 (11):
  • [40] Effect of dose metrics and radiation risk models when optimizing CT x-ray tube voltage
    Huda, W.
    Ogden, K. M.
    Khorasani, M. R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (17): : 4719 - 4732