Effect of x-ray tube current on the accuracy of cerebral perfusion parameters obtained by CT perfusion studies

被引:20
|
作者
Murase, K
Nanjo, T
Ii, S
Miyazaki, S
Hirata, M
Sugawara, Y
Kudo, M
Sasaki, K
Mochizuki, T
机构
[1] Osaka Univ, Grad Sch Med, Dept Med Phys & Engn, Suita, Osaka 5650871, Japan
[2] Ehime Univ, Sch Med, Dept Radiol, Touon, Ehime 7910295, Japan
[3] GE Yokogawa Med Syst, Tokyo 1918503, Japan
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2005年 / 50卷 / 21期
关键词
D O I
10.1088/0031-9155/50/21/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to investigate the effect of x-ray tube current on the accuracy of cerebral perfusion parameters obtained by CT perfusion studies using multi-detector row CT (MDCT). Following the standard CT perfusion study protocol, continuous (cine) scans (1 s/rotation x 60 s) consisting of four 5 mm thick contiguous slices were performed using an MDCT scanner with a tube voltage of 80 kVp and a tube current of 200 mA. We generated the simulated images with tube currents of 50 mA, 100 mA and 150 mA by adding the corresponding noise to the raw scan data of the original image acquired above using a noise simulation tool. From the original and simulated images, we generated the functional images of cerebral blood flow (CBF), cerebral blood volume (CBV) and mean transit time (MTT) in seven patients with cerebrovascular disease, and compared the correlation coefficients (CCs) between the perfusion parameter values obtained from the original and simulated images. The coefficients of variation (CVs) in the white matter were also compared. The CC values deteriorated with decreasing tube current. There was a significant difference between 50 mA and 100 mA for all perfusion parameters. The CV values increased with decreasing tube current. There were significant differences between 50 mA and 100 mA and between 100 mA and 150 mA for CBF. For CBV and MTT, there was also a significant difference between 150 mA and 200 mA. This study will be useful for understanding the effect of x-ray tube current on the accuracy of cerebral perfusion parameters obtained by CT perfusion studies using MDCT, and for selecting the tube current.
引用
收藏
页码:5019 / 5029
页数:11
相关论文
共 50 条
  • [1] Effect of input artery choice on perfusion parameters obtained by CT perfusion
    Hoeffner, EG
    Carlos, RC
    Mukherji, SK
    RADIOLOGY, 2002, 225 : 662 - 662
  • [2] Cerebral perfusion imaging of live mice by fluorescent X-ray CT
    Takeda, T
    Wu, J
    Thet-Thet-Lwin
    Sunaguchi, N
    Yuasa, T
    Hyodo, K
    Dilmanian, FA
    Minami, M
    Akatsuka, T
    2005 INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP), VOLS 1-5, 2005, : 2997 - 3000
  • [3] A perfusion procedure for imaging of the mouse cerebral vasculature by X-ray micro-CT
    Ghanavati, Sahar
    Yu, Lisa X.
    Lerch, Jason P.
    Sled, John G.
    JOURNAL OF NEUROSCIENCE METHODS, 2014, 221 : 70 - 77
  • [4] Reproducibility of Perfusion Parameters Obtained From Perfusion CT in Lung Tumors
    Ng, Chaan S.
    Chandler, Adam G.
    Wei, Wei
    Anderson, Ella F.
    Herron, Delise H.
    Charnsangavej, Chusilp
    Kurzrock, Razelle
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (01) : 113 - 121
  • [5] Performance of iterative reconstruction in cerebral perfusion SPECT with x-ray CT based attenuation correction
    Fujishita, T.
    Fujioka, R.
    Kurihara, H.
    Yamaki, N.
    Koura, S.
    Kodama, R.
    Ideguchi, T.
    Ochi, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 : S409 - S409
  • [6] The Effect of X-Ray Tube Voltage Modulation To Quality of Perfusion Images in Cone Beam C-arm CT
    Haseljic, Hana
    Frysch, Robert
    Kulvait, Vojtech
    Pfeiffer, Tim
    Punzet, Daniel
    Hensen, Bennet
    Wacker, Frank
    Brueschc, Inga
    Werncke, Thomas
    Rose, Georg
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [7] The Effect of Scan Duration on the Measurement of Perfusion Parameters in CT Perfusion Studies of Brain Tumors
    Yeung, Timothy Pok Chi
    Yartsev, Slav
    Bauman, Glenn
    He, Wenqing
    Fainardi, Enrico
    Lee, Ting-Yim
    ACADEMIC RADIOLOGY, 2013, 20 (01) : 59 - 65
  • [8] Evaluation of cerebral perfusion parameters measured by perfusion CT in chronic cerebral ischemia: Comparison with xenon CT
    Furukawa, M
    Kashiwagi, S
    Matsunaga, N
    Suzuki, M
    Kishimoto, K
    Shirao, S
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2002, 26 (02) : 272 - 278
  • [9] Aspects on the accuracy of cerebral perfusion parameters obtained by dynamic susceptibility contrast MRI:: a simulation study
    Knutsson, L
    Ståhlberg, F
    Wirestam, R
    MAGNETIC RESONANCE IMAGING, 2004, 22 (06) : 789 - 798
  • [10] Myocardial perfusion imaging with a combined X-ray CT and SPECT system
    Kalki, K
    Blankespoor, SC
    Brown, JK
    Hasegawa, BH
    Dae, MW
    Chin, M
    Stillson, C
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (10) : 1535 - 1540