Effect of X-ray Tube Parameters and Iodine Concentration on Image Quality and Radiation Dose in Cerebral Pediatric and Adult CT Angiography A Phantom Study

被引:22
|
作者
Papadakis, Antonios E. [1 ]
Perisinakis, Kostas [2 ]
Raissaki, Maria [3 ]
Damilakis, John [2 ]
机构
[1] Univ Hosp Herakl, Dept Med Phys, Iraklion 71110, Greece
[2] Univ Crete, Dept Med Phys, Iraklion, Greece
[3] Univ Crete, Dept Radiol, Iraklion, Greece
关键词
CT angiography; automatic exposure control; imaging optimization; radiation protection; COMPUTED-TOMOGRAPHY-ANGIOGRAPHY; AUTOMATIC EXPOSURE CONTROL; CURRENT MODULATION; SCANNING PARAMETERS; RATIONAL APPROACH; 80; KVP; REDUCTION; CONTRAST; PROTOCOLS; VOLTAGE;
D O I
10.1097/RLI.0b013e31827efc17
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The aim of the present phantom study was to investigate the effect of x-ray tube parameters and iodine concentration on image quality and radiation dose in cerebral computed tomographic (CT) angiographic examinations of pediatric and adult individuals. Materials and Methods: Four physical anthropomorphic phantoms that represent the average individual as neonate, 1-year-old, 5-year-old, and 10-year-old children and the RANDO phantom that simulates the average adult individual were used. Cylindrical vessels were bored along the brain-equivalent plugs of each physical phantom. To simulate the brain vasculature, vessels of 0.6, 1, 2, and 3 mm in diameter were created. These vessels were filled with contrast medium (CM) solutions at different iodine concentrations, that is, 5.6, 4.2, 2.7, and 1.4 mg I/mL. The phantom heads were scanned at 120, 100, and 80 kV. The applied quality reference tube current-time product values ranged from a minimum of 45 to a maximum of 680. The CT acquisitions were performed on a 16-slice CT scanner using the automatic exposure control system. Image quality was evaluated on the basis of image noise and contrast-to-noise ratio (CNR) between the contrast-enhanced iodinated vessels and the unenhanced regions of interest. Dose reduction was calculated as the percentage difference of the CT dose index value at the quality reference tube current-time product and the CT dose index at the mean modulated tube current-time product. Results: Image noise that was measured using the preset tube current-time product settings varied significantly among the different phantoms (P < 0.0001). Hounsfield unit number of iodinated vessels was linearly related to CM concentration (r(2) = 0.907) and vessel diameter (r(2) = 0.918). The Hounsfield unit number of iodinated vessels followed a decreasing trend from the neonate phantom to the adult phantom at all kilovoltage settings. For the same image noise level, a CNR improvement of up to 69% and a dose reduction of up to 61% may be achieved when CT acquisition is performed at 80 kV compared with 120 kV. For the same CNR, a reduction by 25% of the administered CM concentration may be achieved when CTacquisition is performed at 80 kV compared with 120 kV. Conclusions: In cerebral CT angiographic studies, appropriate adjustment of the preset tube current-time product settings is required to achieve the same image noise level among participants of different age. Cerebral CT angiography at 80 kV significantly improves CNR and significantly reduces radiation dose. Moreover, at 80 kV, a considerable reduction of the administered amount of the CM may be reached, thus reducing potential risks for contrast-induced nephropathy.
引用
收藏
页码:192 / 199
页数:8
相关论文
共 50 条
  • [41] The Effect of Magnification on the Image Quality and the Radiation Dose in X-ray Digital Mammography: a Monte Carlo Simulation Study
    Choi, Yu-Na
    Kim, Hee-Joung
    Park, Hye-Suk
    Lee, Chang-Lae
    Cho, Hyo-Min
    Lee, Seung-Wan
    Ryu, Hyun-Ju
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2010, 57 (03) : 494 - 500
  • [42] A new phantom for remote monitoring of dose and image quality in dental X-ray units
    Goodenough, DJ
    Levy, JR
    Wormuth, JR
    RADIOLOGY, 1997, 205 : 1404 - 1404
  • [43] A phantom study into constant image quality of low-dose CT in the pediatric population
    Caan, G. A.
    Almeida, A. S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S649 - S649
  • [44] Effect of low tube kV on radiation dose and image quality in retrospective ECG-gated coronary CT angiography
    Abou-Issa, Ahmed H.
    Elganayni, Fady
    Al-Azzazy, Mohammed Zakaria
    EGYPTIAN JOURNAL OF RADIOLOGY AND NUCLEAR MEDICINE, 2011, 42 (3-4): : 327 - 333
  • [45] Comparison of in-phantom dose distributions for coronary angiography using an x-ray machine and synchrotron radiation
    Nariyama, N
    Tanaka, S
    Nakane, Y
    Nakashima, H
    Hirayama, H
    Ban, S
    Namito, Y
    Hyodo, K
    Takeda, T
    MEDICAL PHYSICS, 2001, 28 (01) : 16 - 21
  • [46] Dose reduction and image quality in CT angiography for cerebral aneurysm with various tube potentials and current settings
    Imai, K.
    Ikeda, M.
    Kawaura, C.
    Aoyama, T.
    Enchi, Y.
    Yamauchi, M.
    BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1017): : E673 - E681
  • [47] Spectral optimization of iodine-enhanced CT: Quantifying the effect of tube voltage on image quality and radiation exposure determined at an anthropomorphic phantom
    Brix, Gunnar
    Lechel, Ursula
    Sudarski, Sonja
    Trumm, Christoph
    Henzler, Thomas
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (08): : 999 - 1006
  • [48] Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine Intake
    Yue-ying Pan
    Shu-chang Zhou
    Yu-jin Wang
    Qian Li
    Ting-ting Zhu
    Chun-xia Liu
    Han-xiong Guan
    Current Medical Science, 2020, 40 : 178 - 183
  • [49] Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine Intake
    Pan, Yue-ying
    Zhou, Shu-chang
    Wang, Yu-jin
    Li, Qian
    Zhu, Ting-ting
    Liu, Chun-xia
    Guan, Han-xiong
    CURRENT MEDICAL SCIENCE, 2020, 40 (01) : 178 - 183
  • [50] The effect of quality control on some X-ray tube output parameters
    Tabakov, S
    Stoeva, M
    Lewis, C
    PHYSICA MEDICA, 2001, 17 : 42 - 46