Impact of heart disease and calibration interval on accuracy of pulse transit time-based blood pressure estimation

被引:44
|
作者
Ding, Xiaorong [1 ]
Zhang, Yuanting [1 ]
Tsang, Hon Ki [1 ]
机构
[1] Chinese Univ Hong Kong, Dept Elect Engn, Shatin, Hong Kong, Peoples R China
关键词
heart disease; calibration interval; pulse transit time; blood pressure; estimation accuracy; FINGER PHOTOPLETHYSMOGRAM; ARRIVAL-TIME; VALIDATION; HYPERTENSION; DEVICES;
D O I
10.1088/0967-3334/37/2/227
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Continuous blood pressure (BP) measurement without a cuff is advantageous for the early detection and prevention of hypertension. The pulse transit time (PTT) method has proven to be promising for continuous cuffless BP measurement. However, the problem of accuracy is one of the most challenging aspects before the large-scale clinical application of this method. Since PTT-based BP estimation relies primarily on the relationship between PTT and BP under certain assumptions, estimation accuracy will be affected by cardiovascular disorders that impair this relationship and by the calibration frequency, which may violate these assumptions. This study sought to examine the impact of heart disease and the calibration interval on the accuracy of PTT-based BP estimation. The accuracy of a PTT-BP algorithm was investigated in 37 healthy subjects and 48 patients with heart disease at different calibration intervals, namely 15 min, 2 weeks, and 1 month after initial calibration. The results showed that the overall accuracy of systolic BP estimation was significantly lower in subjects with heart disease than in healthy subjects, but diastolic BP estimation was more accurate in patients than in healthy subjects. The accuracy of systolic and diastolic BP estimation becomes less reliable with longer calibration intervals. These findings demonstrate that both heart disease and the calibration interval can influence the accuracy of PTT-based BP estimation and should be taken into consideration to improve estimation accuracy.
引用
收藏
页码:227 / 237
页数:11
相关论文
共 50 条
  • [31] Using Individualized Pulse Transit Time Calibration to Monitor Blood Pressure During Exercise
    Jeong, In Cheol
    Wood, Jeffrey
    Finkelstein, Joseph
    INFORMATICS, MANAGEMENT AND TECHNOLOGY IN HEALTHCARE, 2013, 190 : 39 - 41
  • [32] Near threshold pulse transit time processor for central blood pressure estimation
    Veirano, Francisco
    Perez-Nicoli, Pablo
    Gammarano, Nicolas
    Fierro, German
    Silveira, Fernando
    2022 IEEE 13TH LATIN AMERICAN SYMPOSIUM ON CIRCUITS AND SYSTEMS (LASCAS), 2022, : 172 - 175
  • [33] Characters available in photoplethysmogram for blood pressure estimation: beyond the pulse transit time
    Yanjun Li
    Zengli Wang
    Lin Zhang
    Xianglin Yang
    Jinzhong Song
    Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 367 - 376
  • [34] Characters available in photoplethysmogram for blood pressure estimation: beyond the pulse transit time
    Li, Yanjun
    Wang, Zengli
    Zhang, Lin
    Yang, Xianglin
    Song, Jinzhong
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (02) : 367 - 376
  • [35] Restrictive factors of the measurement accuracy to estimate Blood Pressure with Pulse wave Transit Time
    Chen, X. M.
    Li, Y. B.
    Zhang, Y.
    Deng, N.
    BIOINFORMATICS AND BIOMEDICAL ENGINEERING: NEW ADVANCES, 2016, : 435 - 441
  • [36] A Physical Model-Based Approach to One-Point Calibration of Pulse Transit Time to Blood Pressure
    Mousavi, Azin
    Inan, Omer T.
    Mukkamala, Ramakrishna
    Hahn, Jin-Oh
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2024, 71 (02) : 477 - 483
  • [37] Pulse Transit Time Based Continuous Cuffless Blood Pressure Estimation: A New Extension and A Comprehensive Evaluation
    Ding, Xiaorong
    Yan, Bryan P.
    Zhang, Yuan-Ting
    Liu, Jing
    Zhao, Ni
    Tsang, Hon Ki
    SCIENTIFIC REPORTS, 2017, 7
  • [38] Pulse Transit Time Based Continuous Cuffless Blood Pressure Estimation: A New Extension and A Comprehensive Evaluation
    Xiaorong Ding
    Bryan P. Yan
    Yuan-Ting Zhang
    Jing Liu
    Ni Zhao
    Hon Ki Tsang
    Scientific Reports, 7
  • [39] An implantable optical blood pressure sensor based on pulse transit time
    Fiala, Jens
    Bingger, Philipp
    Ruh, Dominic
    Foerster, Katharina
    Heilmann, Claudia
    Beyersdorf, Friedhelm
    Zappe, Hans
    Seifert, Andreas
    BIOMEDICAL MICRODEVICES, 2013, 15 (01) : 73 - 81
  • [40] An implantable optical blood pressure sensor based on pulse transit time
    Jens Fiala
    Philipp Bingger
    Dominic Ruh
    Katharina Foerster
    Claudia Heilmann
    Friedhelm Beyersdorf
    Hans Zappe
    Andreas Seifert
    Biomedical Microdevices, 2013, 15 : 73 - 81