Impact of heart disease and calibration interval on accuracy of pulse transit time-based blood pressure estimation

被引:44
|
作者
Ding, Xiaorong [1 ]
Zhang, Yuanting [1 ]
Tsang, Hon Ki [1 ]
机构
[1] Chinese Univ Hong Kong, Dept Elect Engn, Shatin, Hong Kong, Peoples R China
关键词
heart disease; calibration interval; pulse transit time; blood pressure; estimation accuracy; FINGER PHOTOPLETHYSMOGRAM; ARRIVAL-TIME; VALIDATION; HYPERTENSION; DEVICES;
D O I
10.1088/0967-3334/37/2/227
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Continuous blood pressure (BP) measurement without a cuff is advantageous for the early detection and prevention of hypertension. The pulse transit time (PTT) method has proven to be promising for continuous cuffless BP measurement. However, the problem of accuracy is one of the most challenging aspects before the large-scale clinical application of this method. Since PTT-based BP estimation relies primarily on the relationship between PTT and BP under certain assumptions, estimation accuracy will be affected by cardiovascular disorders that impair this relationship and by the calibration frequency, which may violate these assumptions. This study sought to examine the impact of heart disease and the calibration interval on the accuracy of PTT-based BP estimation. The accuracy of a PTT-BP algorithm was investigated in 37 healthy subjects and 48 patients with heart disease at different calibration intervals, namely 15 min, 2 weeks, and 1 month after initial calibration. The results showed that the overall accuracy of systolic BP estimation was significantly lower in subjects with heart disease than in healthy subjects, but diastolic BP estimation was more accurate in patients than in healthy subjects. The accuracy of systolic and diastolic BP estimation becomes less reliable with longer calibration intervals. These findings demonstrate that both heart disease and the calibration interval can influence the accuracy of PTT-based BP estimation and should be taken into consideration to improve estimation accuracy.
引用
收藏
页码:227 / 237
页数:11
相关论文
共 50 条
  • [41] Noninvasive Cuffless Blood Pressure Estimation Using Pulse Transit Time and Impedance Plethysmography
    Toan Huu Huynh
    Jafari, Roozbeh
    Chung, Wan-Young
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (04) : 967 - 976
  • [42] Optimizing Non-Invasive Blood Pressure Estimation Using Pulse Transit Time
    Jeong, In Cheol
    Finkelstein, Joseph
    MEDINFO 2013: PROCEEDINGS OF THE 14TH WORLD CONGRESS ON MEDICAL AND HEALTH INFORMATICS, PTS 1 AND 2, 2013, 192 : 1198 - 1198
  • [43] Estimation of pulse transit time using two diametric blood pressure waveform measurements
    Hahn, Jin-Oh
    Reisner, Andrew T.
    Asada, H. Harry
    MEDICAL ENGINEERING & PHYSICS, 2010, 32 (07) : 753 - 759
  • [44] Cuffless Blood Pressure Estimation Using Pulse Transit Time and Photoplethysmogram Intensity Ratio
    Gholamhosseini, Hamid
    Baig, Mirza
    Rastegar, Solmaz
    Linden, Maria
    PROCEEDINGS OF THE 15TH INTERNATIONAL CONFERENCE ON WEARABLE MICRO AND NANO TECHNOLOGIES FOR PERSONALIZED HEALTH (PHEALTH 2018), 2018, 249 : 77 - 83
  • [45] The influence of heart rate on the relationship between pulse transit time and systolic blood pressure
    Fu, Zhizhong
    Song, Xinyue
    Qin, Tianyi
    Chen, Yifan
    Ding, Xiaorong
    PHYSIOLOGICAL MEASUREMENT, 2024, 45 (10)
  • [46] Continuous Cuffless Blood Pressure Estimation Based on Pulse Transit Time: A New Evaluation with Invasive BP Reference
    Tang, Zunyi
    Chen, Ying
    Liu, Shing-Hong
    Kobayashi, Toshio
    Chen, Wenxi
    2019 IEEE 10TH INTERNATIONAL CONFERENCE ON AWARENESS SCIENCE AND TECHNOLOGY (ICAST 2019), 2019, : 516 - 520
  • [47] Regression-forests-based Estimation of Blood Pressure using the Pulse Transit Time Obtained by Facial Photoplethysmogram
    Yoshioka, Mototaka
    Bounyong, Souksakhone
    2017 INTERNATIONAL JOINT CONFERENCE ON NEURAL NETWORKS (IJCNN), 2017, : 3248 - 3253
  • [48] Limitation of Pulse Transit Time to Estimate Blood Pressure
    Sklarsky, Amanda
    Garvin, Nathan M.
    Pawelczyk, James A.
    FASEB JOURNAL, 2019, 33
  • [49] Continuous blood pressure measurement based on pulse transit time: a comparison to ambulatory blood pressure measurement
    Kuechler, G.
    Gehring, J.
    Wesely, M.
    Thimel, K.
    Haberl, R.
    EUROPEAN HEART JOURNAL, 2016, 37 : 1267 - 1267
  • [50] Accuracy of pulse interval timing in ambulatory blood pressure measurement
    Sarah A. Kane
    James R. Blake
    Frank J. McArdle
    Philip Langley
    Andrew J. Sims
    Scientific Reports, 6