Bipolar Radiofrequency Ablation: Development of a New Expandable Device

被引:4
|
作者
Ito, Nobutake [1 ,2 ]
Pfeffer, Jochen [1 ]
Isfort, Peter [1 ,2 ]
Penzkofer, Tobias [1 ,2 ]
Kuhl, Christiane K. [1 ]
Mahnken, Andreas H. [3 ]
Schmitz-Rode, Thomas [2 ]
Bruners, Philipp [1 ]
机构
[1] Rhein Westfal TH Aachen, Dept Diagnost Radiol, Aachen, Germany
[2] Rhein Westfal TH Aachen, Helmholtz Inst Biomed Engn, Aachen, Germany
[3] Univ Marburg, Dept Radiol, Marburg, Germany
关键词
Liver; Minimally invasive therapy; Radiofrequency ablation; INTERNALLY COOLED ELECTRODES; ENHANCED HEPATIC BIPOLAR; VIVO BOVINE LIVER; EFFICIENCY; MONOPOLAR; TUMORS; COMPLICATIONS; APPLICATORS; COAGULATION; PRINCIPLES;
D O I
10.1007/s00270-013-0703-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To test the performance of an expandable bipolar probe as a simple technical solution for extending the coagulation volume. On the basis of a commercially available monopolar radiofrequency (RF) probe (LeVeen), an expandable bipolar RF probe was developed by integrating a second electrode into the probe shaft. The influence of length on the second electrode, and the distance between both electrodes and generator output was investigated by performing ten ablations for each condition on a freshly excised bovine liver. Macroscopically quantified coagulation volumes, lesion shape characteristics, and procedure durations were recorded. Results of the prototype featuring the optimal configuration were compared to the original LeVeen probe and commonly used bipolar RF probe (CelonLabPower). Extension of the shaft electrode length, increasing distance between the shaft electrode and the tip electrode, and reduction of generator output resulted in increasing coagulation volumes. The coagulation volumes the prototype generated were significantly smaller and more elliptically shaped than the monopolar probe (9.4 +/- A 1.5 cm(3) vs. 12.1 +/- A 1.6 cm(3)), but were larger than the commercially available bipolar RF probe (vs. 7.3 +/- A 0.5). The procedure duration of the prototype was comparable to the monopolar probe (467 +/- A 31 s vs. 464 +/- A 17 s) and shorter than the bipolar probe (vs. 2009 +/- A 444 s). In comparison to the commercially available bipolar system, the developed prototype exhibited favorable results. The first benchmark testing of the developed bipolar prototype had promising results. However, further optimization of the applicator design and ablation protocol is needed to enlarge the achievable coagulation volume.
引用
收藏
页码:770 / 776
页数:7
相关论文
共 50 条
  • [1] Bipolar Radiofrequency Ablation: Development of a New Expandable Device
    Nobutake Ito
    Jochen Pfeffer
    Peter Isfort
    Tobias Penzkofer
    Christiane K. Kuhl
    Andreas H. Mahnken
    Thomas Schmitz-Rode
    Philipp Bruners
    CardioVascular and Interventional Radiology, 2014, 37 : 770 - 776
  • [2] Endoscopic Radiofrequency Ablation in Colorectal Cancer: Initial Clinical Results of a New Bipolar Radiofrequency Ablation Device
    Vavra, Petr
    Dostalik, Jan
    Zacharoulis, Dimitris
    Khorsandi, Shirin E.
    Khan, Shahid A.
    Habib, Nagy A.
    DISEASES OF THE COLON & RECTUM, 2009, 52 (02) : 355 - 358
  • [3] Technical advance in silico and in vitro development of a new bipolar radiofrequency ablation device for renal denervation
    Perez, Noel
    Muffly, Karl
    Saddow, Stephen E.
    BMC CARDIOVASCULAR DISORDERS, 2021, 21 (01)
  • [4] Technical advance in silico and in vitro development of a new bipolar radiofrequency ablation device for renal denervation
    Noel Pérez
    Karl Muffly
    Stephen E. Saddow
    BMC Cardiovascular Disorders, 21
  • [5] Radiofrequency ablation of osteoid osteoma:: Initial results with a bipolar ablation device
    Mahnken, Andreas H.
    Tacke, Josef A.
    Wildberger, Joachim E.
    Guenther, Rolf W.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 17 (09) : 1465 - 1470
  • [6] Radiofrequency Ablation of Liver Tumors Using a Novel Bipolar Device
    Eisele, Robert M.
    Neuhaus, Peter
    Schumacher, Guido
    JOURNAL OF LAPAROENDOSCOPIC & ADVANCED SURGICAL TECHNIQUES, 2008, 18 (06): : 857 - 863
  • [7] Surgical ablation of atrial fibrillation with a novel bipolar radiofrequency device
    Benussi, S
    Nascimbene, S
    Calori, G
    Denti, P
    Ziskind, Z
    Kassem, S
    La Canna, G
    Pappone, C
    Alfieri, O
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2005, 130 (02): : 491 - 497
  • [8] RADIOFREQUENCY ABLATION PRIOR TO LIVER TRANSPLANTATION: FEASIBILITY OF A NOVEL BIPOLAR DEVICE
    Eisele, Robert M.
    Zhukowa, Julia
    Chopra, Sascha S.
    Schmidt, Sven C.
    Neuhaus, Peter
    Schumacher, Guido
    LIVER TRANSPLANTATION, 2009, 15 (07) : S253 - S253
  • [9] Prostate Radiofrequency Focal Ablation (ProRAFT) Trial: A Prospective Development Study Evaluating a Bipolar Radiofrequency Device to Treat Prostate Cancer
    Orczyk, Clement
    Barratt, Dean
    Brew-Graves, Chris
    Hu, Yi Peng
    Freeman, Alex
    McCartan, Neil
    Potyka, Ingrid
    Ramachandran, Navin
    Rodell, Rachael
    Williams, Norman R.
    Emberton, Mark
    Ahmed, Hashim U.
    JOURNAL OF UROLOGY, 2021, 205 (04): : 1090 - 1098
  • [10] Bipolar radiofrequency ablation of the kidney: Comparison with monopolar radiofrequency ablation
    Nakada, SY
    Jerde, TJ
    Warner, TF
    Wright, AS
    Haemmerich, D
    Mahvi, DM
    Lee, FT
    JOURNAL OF ENDOUROLOGY, 2003, 17 (10) : 927 - 933