Total Corneal Astigmatism Measurements: Agreement Between 2 Rotating Scheimpflug Cameras

被引:2
|
作者
Savini, Giacomo [1 ]
Noeser, Kristian [2 ]
Schiano-Lomoriello, Domenico [1 ]
Ducoli, Pietro [1 ]
机构
[1] Fdn GB Bietti, IRCCS, Via Livenza 3, I-00198 Rome, Italy
[2] Reg Hosp Randers, Dept Ophthalmol, Randers, Denmark
关键词
corneal astigmatism; Scheimpflug camera; vector analysis; TORIC INTRAOCULAR LENSES; POWER CALCULATION; CATARACT-SURGERY; REPEATABILITY; TOPOGRAPHY; REPRODUCIBILITY; TOMOGRAPHY; PENTACAM; SYSTEMS;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate agreement between rotating Scheimpflug camera (Pentacam HR, Oculus) and rotating Scheimpflug camera combined with Placido disc corneal topography (Sirius, CSO) in measuring total corneal astigmatism (TCA). Methods: In this observational study, all patients undergoing cataract surgery with preoperative measurement of TCA by both devices and a validated corneal topographer (Keratron, Optikon 2000) were retrospectively evaluated. Astigmatism analysis was performed with and without vector analysis separately in eyes with with-the-rule, against-the-rule, and oblique astigmatism. Vector analysis was performed using the Naeser polar system. Results: In 130 eyes of 130 subjects, nonvectorial analysis revealed that the mean TCA values obtained with the Sirius were higher than the corresponding values given by the Pentacam HR in all subgroups, although the difference was statistically significant only in eyes with against-the-rule astigmatism (P = 0.0009). This finding was confirmed by vector analysis. A TCA magnitude difference greater than 0.5 diopters was observed in 20.8% of cases, and a TCA axis difference greater than 10 degrees was observed in 45.4% of cases. Axis differences dropped to 18.5% when only eyes with astigmatism >0.75 diopters were analyzed and 3 measurements were averaged. The mean difference in the meridional and torsional power of TCA was close to zero in all subgroups, but with relatively large standard deviations (approximately 0.5 D). Conclusions: Agreement between both devices in measuring TCA is only moderate with respect to both magnitude and axis orientation.
引用
收藏
页码:463 / 469
页数:7
相关论文
共 50 条
  • [31] Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements
    Stella Briggs
    Marwa Bin Moammar
    International Journal of Ophthalmology, 2016, 9 (02) : 239 - 242
  • [32] Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements
    Briggs, Stella
    Bin Moammar, Marwa
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2016, 9 (02) : 239 - 242
  • [33] The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging
    Zhao, Yune
    Chen, Ding
    Savini, Giacomo
    Wang, Qing
    Zhang, Hongfang
    Jin, Yili
    Song, Benhao
    Ning, Rui
    Huang, Jinhai
    Mei, Chenyang
    EYE AND VISION, 2020, 7 (01)
  • [34] Posterior Corneal Astigmatism for Toric IOL Calculations: A Comparison Between the Barrett Toric Calculator Predicted Astigmatism Versus Direct Posterior Corneal Measurement with a Scheimpflug Camera
    Kingrey, Brandon S.
    Kingrey, David
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [35] Posterior and total corneal astigmatism measured with optical coherence tomography-based biometer and dual Scheimpflug analyzer
    Wang, Li
    Cao, Danmin
    Vilar, Cesar
    Koch, Douglas D.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (12): : 1652 - 1658
  • [36] Agreement between Corneal Thickness Measurements Using Pentacam Scheimpflug Camera, Noncontact Specular Microscopy, and Ultrasonographic Pachymetry in Diabetic Patients
    Altay, Yesim
    Balta, Ozgur
    Demirok, Gulizar
    Burcu, Ayse
    Balta, Oyku Bezen
    Ornek, Firdevs
    CURRENT EYE RESEARCH, 2017, 42 (02) : 187 - 194
  • [37] Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography
    Amano, S
    Honda, N
    Amano, Y
    Yamagami, S
    Miyai, T
    Samejima, T
    Ogata, M
    Miyata, K
    OPHTHALMOLOGY, 2006, 113 (06) : 937 - 941
  • [38] Repeatability of posterior and total corneal curvature measurements with a dual Scheimpflug-Placido tomographer
    Kim, Eric J.
    de Oca, Ildamaris Montes
    Wang, Li
    Weikert, Mitchell P.
    Koch, Douglas D.
    Khandelwal, Sumitra S.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (12): : 2731 - 2738
  • [39] Comparison between color light-emitting diode corneal topographer and scheimpflug imager in measuring corneal astigmatism and corneal higher order aberrations
    Hwang, Yousook
    Kim, Da Yeoung
    Chung, So-Hyang
    ACTA OPHTHALMOLOGICA, 2019, 97
  • [40] Agreement between swept-source OCT and Scheimpflug biometers in ocular measurements
    Mclintock, Cameron
    Mckelvie, James
    Niyazmand, Hamed
    Uprety, Samir
    BMC OPHTHALMOLOGY, 2025, 25 (01)