Total Corneal Astigmatism Measurements: Agreement Between 2 Rotating Scheimpflug Cameras

被引:2
|
作者
Savini, Giacomo [1 ]
Noeser, Kristian [2 ]
Schiano-Lomoriello, Domenico [1 ]
Ducoli, Pietro [1 ]
机构
[1] Fdn GB Bietti, IRCCS, Via Livenza 3, I-00198 Rome, Italy
[2] Reg Hosp Randers, Dept Ophthalmol, Randers, Denmark
关键词
corneal astigmatism; Scheimpflug camera; vector analysis; TORIC INTRAOCULAR LENSES; POWER CALCULATION; CATARACT-SURGERY; REPEATABILITY; TOPOGRAPHY; REPRODUCIBILITY; TOMOGRAPHY; PENTACAM; SYSTEMS;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate agreement between rotating Scheimpflug camera (Pentacam HR, Oculus) and rotating Scheimpflug camera combined with Placido disc corneal topography (Sirius, CSO) in measuring total corneal astigmatism (TCA). Methods: In this observational study, all patients undergoing cataract surgery with preoperative measurement of TCA by both devices and a validated corneal topographer (Keratron, Optikon 2000) were retrospectively evaluated. Astigmatism analysis was performed with and without vector analysis separately in eyes with with-the-rule, against-the-rule, and oblique astigmatism. Vector analysis was performed using the Naeser polar system. Results: In 130 eyes of 130 subjects, nonvectorial analysis revealed that the mean TCA values obtained with the Sirius were higher than the corresponding values given by the Pentacam HR in all subgroups, although the difference was statistically significant only in eyes with against-the-rule astigmatism (P = 0.0009). This finding was confirmed by vector analysis. A TCA magnitude difference greater than 0.5 diopters was observed in 20.8% of cases, and a TCA axis difference greater than 10 degrees was observed in 45.4% of cases. Axis differences dropped to 18.5% when only eyes with astigmatism >0.75 diopters were analyzed and 3 measurements were averaged. The mean difference in the meridional and torsional power of TCA was close to zero in all subgroups, but with relatively large standard deviations (approximately 0.5 D). Conclusions: Agreement between both devices in measuring TCA is only moderate with respect to both magnitude and axis orientation.
引用
收藏
页码:463 / 469
页数:7
相关论文
共 50 条
  • [41] Agreement of Corneal Diameter Measurements Obtained by a Swept-source Biometer and a Scheimpflug-based Topographer
    Salouti, Ramin
    Nowroozzadeh, Mohammad Hossein
    Tajbakhsh, Zahra
    Bagheri, Mansoureh
    Ghoreyshi, Maryam
    Azizzadeh, Monir
    Razeghinejad, Mohammad Reza
    CORNEA, 2017, 36 (11) : 1373 - 1376
  • [42] Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices
    Feng, Yue
    Nitter, Tore
    Bertelsen, Geir
    Stojanovic, Aleksandar
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2024, 52 (08): : 800 - 810
  • [43] Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry
    Huang, Jinhai
    Savini, Giacomo
    Wang, Chengfang
    Lu, Weicong
    Gao, Rongrong
    Li, Yuanguang
    Wang, Qinmei
    Zhao, Yune
    JOURNAL OF OPHTHALMOLOGY, 2015, 2015
  • [44] Corneal densitometry measurements comparison between anterior segment OCT and scheimpflug imaging
    Lupardi, Enrico
    Moramarco, Antonio
    Cassini, Federico
    Febbraro, Simone
    Savini, Giacomo
    Fontana, Luigi
    INTERNATIONAL OPHTHALMOLOGY, 2024, 44 (01)
  • [45] Corneal Astigmatism Measurements Comparison among Ray-Tracing Aberrometry, Partial Coherence Interferometry, and Scheimpflug Imaging System
    Zhang, Yaqin
    Dong, Jing
    Zhang, Suhua
    Sun, Bin
    Wang, Xiaoliang
    Tang, Maolong
    Wang, Xiaogang
    JOURNAL OF OPHTHALMOLOGY, 2020, 2020
  • [46] Repeatability and Agreement of Horizontal Corneal Diameter Measurements Between Scanning-Slit Topography, Dual Rotating Scheimpflug Camera With Placido Disc Tomography, Placido Disc Topography, and Optical Coherence Tomography
    Namkung, Soobin
    Boyle, Alexander Buckenham
    Li, Ye
    Gokul, Akilesh
    McGhee, Charles
    Ziaei, Mohammed
    CORNEA, 2022, 41 (11) : 1392 - 1397
  • [47] Evaluation of keratometric and total corneal astigmatism measurements from optical biometers and anterior segment tomographers and mapping to reconstructed corneal astigmatism vector components
    Langenbucher, Achim
    Szentmary, Nora
    Cayless, Alan
    Hoffmann, Peter
    Wendelstein, Jascha
    Pantanelli, Seth
    PLOS ONE, 2025, 20 (01):
  • [48] Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography
    Mohammad Naderan
    Saeed Shoar
    Morteza Naderan
    Mohammad Amin Kamaleddin
    Mohammad Taher Rajabi
    International Journal of Ophthalmology, 2015, (02) : 275 - 280
  • [49] Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography
    Naderan, Mohammad
    Shoar, Saeed
    Naderan, Morteza
    Kamaleddin, Mohammad Anmin
    Rajabi, Mohammad Taher
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2015, 8 (02) : 275 - 280
  • [50] Agreement and repeatability of corneal tomography in healthy eyes using a new swept-source OCT, a rotating Scheimpflug camera, and a dual Scheimpflug-Placido system
    Herber, Robert
    Lenk, Janine
    Pillunat, Lutz E.
    Raiskup, Frederik
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2022, 48 (02): : 190 - 198