Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements

被引:0
|
作者
Stella Briggs [1 ]
Marwa Bin Moammar [2 ]
机构
[1] College of Applied Medical Sciences, King Saud University
[2] Ophthalmology Clinic, Security Forces Hospital
关键词
central corneal thickness; fluorescein; Scheimpflug device;
D O I
暂无
中图分类号
R770.4 [眼科诊断学];
学科分类号
100212 ;
摘要
AIM: To assess central corneal thickness(CCT)changes measured with Scheimpflug device following instillation of 2% fluorescein in normal subjects.·METHODS: This was a prospective randomized study of 60 hospital volunteers. After baseline CCT measurements of both eyes of 40 subjects were obtained using Scheimpflug system, a drop of preservative-free2% fluorescein, was instilled in one eye and in other eye,one drop of normal saline(control). Measurements were repeated after 1, 2, 5, 10, 20, 30, 40, 50 and 60min(continuous assessment group). Twenty subjects had baseline CCT taken, then fluorescein was instilled in one eye and measurements were taken at 1min. Ten eyes had saline rinse after 1min and 10 other eyes did not,measurements were repeated at 2min(eye rinse group).· RESULTS: The mean baseline CCT for continuous assessment group was 546.2±32.1 μm(range, 489.0-606.0),control eyes was 546.6±30.7 μm(range, 489.0-602.0). At1 min after fluorescein instillation, CCT significantly increased by 37.0 ±34.0 μm(P <0.001), then decreased gradually, reaching baseline at 60 min. CCT variations were not significant in control group(P >0.05). For eye rinse group, CCT mean differences between baseline and 2min were 18.2 μm(95 % CI:-54.7 to 18.3) with rinse and 26.5 μm(95% CI:-62.9 to 9.9) without rinse; paired sample tests were not significant(P >0.05).·CONCLUSION: The presence of fluorescein increased CCT value to a clinically relevant level of 6.8%. Eye rinse did not significantly reduce the effect at 2min post fluorescein timepoint.
引用
收藏
页码:239 / 242
页数:4
相关论文
共 50 条
  • [1] Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements
    Briggs, Stella
    Bin Moammar, Marwa
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2016, 9 (02) : 239 - 242
  • [2] Effect of Fluorescein Dye Staining of the Tear Film on Scheimpflug Measurements of Central Corneal Thickness
    Hirnschall, Nino
    Crnej, Alja
    Gangwani, Vinod
    Findl, Oliver
    CORNEA, 2012, 31 (01) : 18 - 20
  • [3] Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetry
    Ladi, Jeevan S.
    Shah, Nitant A.
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2010, 58 (05) : 385 - 388
  • [4] Correction on the distortion of Scheimpflug imaging for dynamic central corneal thickness
    Li, Tianjie
    Tian, Lei
    Wang, Like
    Hon, Ying
    Lam, Andrew K. C.
    Huang, Yifei
    Wang, Yuanyuan
    Zheng, Yongping
    JOURNAL OF BIOMEDICAL OPTICS, 2015, 20 (05)
  • [5] Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography
    Amano, S
    Honda, N
    Amano, Y
    Yamagami, S
    Miyai, T
    Samejima, T
    Ogata, M
    Miyata, K
    OPHTHALMOLOGY, 2006, 113 (06) : 937 - 941
  • [6] Corneal curvature and thickness Scheimpflug
    Kovacs, Illes
    ACTA OPHTHALMOLOGICA, 2022, 100
  • [7] Effect of contact lens wear on central corneal thickness measurements
    Braun, DA
    Penno, EEA
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (07): : 1319 - 1322
  • [8] Measurement of central corneal thickness and pre-corneal tear film thickness of rabbits using the Scheimpflug system
    Dong, Jing
    Wu, Qiang
    Wang, Xiao-Gang
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2013, 6 (05) : 584 - 587
  • [9] Comparison of central corneal thickness and endothelial cell measurements by Scheimpflug camera system and two noncontact specular microscopes
    Irmak Karaca
    Suzan Guven Yilmaz
    Melis Palamar
    Halil Ates
    International Ophthalmology, 2018, 38 : 1601 - 1609
  • [10] Agreement of Central Corneal Thickness Measurements between Scheimpflug Photography and Optical Low-Coherence Reflectometry in Children
    Hashemi, Hassan
    Nabovati, Payam
    Khabazkhoob, Mehdi
    Emamian, Mohammad Hassan
    Yekta, Abbasali
    Fotouhi, Akbar
    SEMINARS IN OPHTHALMOLOGY, 2020, 35 (04) : 252 - 256