Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements

被引:0
|
作者
Stella Briggs [1 ]
Marwa Bin Moammar [2 ]
机构
[1] College of Applied Medical Sciences, King Saud University
[2] Ophthalmology Clinic, Security Forces Hospital
关键词
central corneal thickness; fluorescein; Scheimpflug device;
D O I
暂无
中图分类号
R770.4 [眼科诊断学];
学科分类号
100212 ;
摘要
AIM: To assess central corneal thickness(CCT)changes measured with Scheimpflug device following instillation of 2% fluorescein in normal subjects.·METHODS: This was a prospective randomized study of 60 hospital volunteers. After baseline CCT measurements of both eyes of 40 subjects were obtained using Scheimpflug system, a drop of preservative-free2% fluorescein, was instilled in one eye and in other eye,one drop of normal saline(control). Measurements were repeated after 1, 2, 5, 10, 20, 30, 40, 50 and 60min(continuous assessment group). Twenty subjects had baseline CCT taken, then fluorescein was instilled in one eye and measurements were taken at 1min. Ten eyes had saline rinse after 1min and 10 other eyes did not,measurements were repeated at 2min(eye rinse group).· RESULTS: The mean baseline CCT for continuous assessment group was 546.2±32.1 μm(range, 489.0-606.0),control eyes was 546.6±30.7 μm(range, 489.0-602.0). At1 min after fluorescein instillation, CCT significantly increased by 37.0 ±34.0 μm(P <0.001), then decreased gradually, reaching baseline at 60 min. CCT variations were not significant in control group(P >0.05). For eye rinse group, CCT mean differences between baseline and 2min were 18.2 μm(95 % CI:-54.7 to 18.3) with rinse and 26.5 μm(95% CI:-62.9 to 9.9) without rinse; paired sample tests were not significant(P >0.05).·CONCLUSION: The presence of fluorescein increased CCT value to a clinically relevant level of 6.8%. Eye rinse did not significantly reduce the effect at 2min post fluorescein timepoint.
引用
收藏
页码:239 / 242
页数:4
相关论文
共 50 条
  • [41] Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera
    Gonul, Saban
    Koktekir, Bengu Ekinci
    Bakbak, Berker
    Gedik, Sansal
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2014, 77 (06) : 345 - 350
  • [42] Comparative measurements of central corneal thickness with two ultrasound pachymeters
    Schmidt, E.
    Schwanitz, K.
    Boehm, A. G.
    Pillunat, L. E.
    OPHTHALMOLOGE, 2006, 103 (08): : 688 - 691
  • [43] Reproducibility of Central Corneal Thickness Measurements in Healthy and Glaucomatous Eyes
    Realini, Tony
    Gurka, Matthew J.
    Weinreb, Robert N.
    JOURNAL OF GLAUCOMA, 2017, 26 (09) : 787 - 791
  • [44] Central corneal thickness measurements in central retinal and branch retinal vein occlusions
    Sy, H
    Becker, N
    Brody, JS
    Dwarakanathan, S
    Pathak, N
    Ahuja, RM
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [45] Sirius Scheimpflug-Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
    Jin, Yili
    McAlinden, Colm
    Sun, Yong
    Wen, Daizong
    Wang, Yiran
    Yu, Jinjin
    Feng, Ke
    Song, Benhao
    Wang, Qinmei
    Chen, Shihao
    Huang, Jinhai
    EYE AND VISION, 2021, 8 (01)
  • [46] Comparison of corneal thickness measurements by ultrasound and scheimpflug photography in eyes that have undergone laser in situ keratomileusis
    Ciolino, Joseph B.
    Khachikian, Steven S.
    Belin, Michael W.
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 145 (01) : 75 - 80
  • [47] Reliability and agreement of the central and mid-peripheral corneal thickness measured by a new Scheimpflug based imaging
    Yu, A-Yong
    Ye, Junming
    Savini, Giacomo
    Wang, Yiran
    Zhang, Tianjiao
    Chen, Min
    Wang, Qinmei
    Huang, Jinhai
    ANNALS OF TRANSLATIONAL MEDICINE, 2021, 9 (14)
  • [48] Changes in Central Corneal Thickness With Air-Puff-Induced Corneal Deformation Using a Method to Correct Scheimpflug and Refractive Distortion
    Okon, Monica D.
    Ma, Yanhui
    Liu, Jun
    Roberts, Cynthia J.
    JOURNAL OF REFRACTIVE SURGERY, 2021, 37 (06) : 422 - +
  • [49] Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry
    Huang, Jinhai
    Savini, Giacomo
    Wang, Chengfang
    Lu, Weicong
    Gao, Rongrong
    Li, Yuanguang
    Wang, Qinmei
    Zhao, Yune
    JOURNAL OF OPHTHALMOLOGY, 2015, 2015
  • [50] Factors affecting central corneal thickness measurement agreement between Scheimpflug imaging and ultrasound pachymetry in keratoconus
    Sorkin, Nir
    Ohri, Anamika
    Jung, Hyejung
    Haines, Lacey
    Sorbara, Luigina
    Mimouni, Michael
    Singal, Neera
    Hatch, Wendy
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2021, 105 (10) : 1371 - 1375