Effect of acquisition technique on radiation dose and image quality in multidetector row computed tomography coronary angiography with submillimeter collimation

被引:77
|
作者
Gerber, TC
Stratmann, BP
Kuzo, RS
Kantor, B
Morin, RL
机构
[1] Mayo Clin, Div Cardiovasc Dis, Jacksonville, FL 32224 USA
[2] Mayo Clin, Dept Diagnost Radiol, Jacksonville, FL 32224 USA
[3] Mayo Clin, Div Cardiovasc Dis, Rochester, MN USA
关键词
computed tomography coronary angiography; image quality; multidetector row computed tomography; radiation dose;
D O I
10.1097/01.rli.0000170628.69792.cb
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: We sought to examine effects of tube voltage and current on radiation dose and image quality for minimally invasive coronary angiography with a 16-slice multidetector row computed tomography (MDCT) scanner. Materials and Methods: We scanned the phantom used in the American College of Radiology Computed Tomography Accreditation Program at tube voltages of 80 and 120 kVp at 550, 650, and 750 mAs(eff), with and without a reduction in radiation dose by electrocardiographically (ECG) controlled tube current modulation (ECG pulsing). Results: Without ECG pulsing, the effective dose was 3 to 13 mSv. On average, a 50% increase in tube voltage led to increased radiation dose (215%), contrast-to-noise ratio (150%), and decreased image noise (-48%). On average, a 17% increase in mAs(eff) led to increased radiation dose (17%) and contrast-to-noise ratio (4%) and decreased image noise (-9%). Dose reduction by ECG pulsing (simulated heart rate, 70 beats per minute) was 28%. With ECG pulsing, noise in images reconstructed during ventricular systole was double that in images reconstructed during ventricular diastole. Conclusions: These quantitative findings about the relationships among scan acquisition parameters, radiation dose, and image quality have practical implications for using ECG pulsing to reduce radiation doses in MDCT coronary angiography.
引用
收藏
页码:556 / 563
页数:8
相关论文
共 50 条
  • [41] Achieving consistent image quality with dose optimization in 64-row multidetector computed tomography prospective ECG gated coronary calcium scoring
    Pan, Zilai
    Pang, Lifang
    Li, Jianying
    Zhang, Huan
    Yang, Wenjie
    Ding, Bei
    Chai, Weimin
    Chen, Kemin
    Yao, Weiwu
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2011, 27 (04): : 611 - 617
  • [42] Multidetector Computed Tomography Chest Examinations With Low-Kilovoltage Protocols in Adults: Effect on Image Quality and Radiation Dose
    Kim, Mi Jung
    Park, Chul Hi
    Choi, Soo Jin
    Hwang, Kyung Hoon
    Kim, Hyung Sik
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2009, 33 (03) : 416 - 421
  • [43] Coronary stenosis quantification in cardiac computed tomography angiography: multi-factorial optimization of image quality and radiation dose
    Zarei, Mojtaba
    Abadi, Ehsan
    Segars, William Paul
    Samei, Ehsan
    JOURNAL OF MEDICAL IMAGING, 2023, 10 (06)
  • [44] Image quality in coronary computed tomography angiography: influence of adaptive statistical iterative reconstruction at various radiation dose levels
    Precht, Helle
    Gerke, Oke
    Thygesen, Jesper
    Egstrup, Kenneth
    Auscher, Soren
    Waaler, Dag
    Lambrechtsen, Jess
    ACTA RADIOLOGICA, 2018, 59 (10) : 1194 - 1202
  • [45] Iterative reconstruction and individualized automatic tube current selection reduce radiation dose while maintaining image quality in 320-multidetector computed tomography coronary angiography
    Williams, M. C.
    Weir, N. W.
    Mirsadraee, S.
    Millar, F.
    Baird, A.
    Minns, F.
    Uren, N. G.
    McKillop, G.
    Bull, R. K.
    van Beek, E. J. R.
    Reid, J. H.
    Newby, D. E.
    CLINICAL RADIOLOGY, 2013, 68 (11) : E570 - E577
  • [46] Estimated radiation dose and image quality comparison of the scan protocols in dual-source computed tomography coronary angiography
    Karaarslan, Ercan
    Ulus, Sila
    Bavbek, Cengiz
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2013, 57 (04) : 407 - 414
  • [47] Radiation dose reduction from low-kilovoltage liver computed tomography using multidetector row computed tomography
    Cho, Pyong-Kon
    RADIATION PROTECTION DOSIMETRY, 2013, 154 (01) : 76 - 80
  • [48] Image quality of coronary computed tomography angiography with 320-Row area detector computed tomography in children with congenital heart disease
    Qin, Jie
    Cai, Songwang
    Peng, Lingrong
    Tang, Wenjie
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 68 (16) : C142 - C142
  • [49] Image Quality of Coronary Computed Tomography Angiography with 320-Row Area Detector Computed Tomography in Children with Congenital Heart Disease
    Tada, Akihiro
    Sato, Shuhei
    Kanie, Yuichiro
    Tanaka, Takashi
    Inai, Ryota
    Akagi, Noriaki
    Morimitsu, Yusuke
    Kanazawa, Susumu
    PEDIATRIC CARDIOLOGY, 2016, 37 (03) : 497 - 503
  • [50] 320-Row multidetector computed tomography coronary angiography in patients presenting with suspected acute coronary syndrome
    Van Velzen, J. E.
    Schuijf, J. D.
    De Graaf, F. R.
    Kroner, E. S. J.
    Kroft, L. J.
    De Roos, A.
    Schalij, M. J.
    Jukema, J. W.
    Van Der Wall, E. E.
    Bax, J. J.
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2011, 13 (0A) : A22 - A22