Comparison of the I-Gel and the Laryngeal Mask Airway Proseal during General Anesthesia: A Systematic Review and Meta-Analysis

被引:37
|
作者
Park, Sun Kyung [1 ]
Choi, Geun Joo [2 ]
Choi, Yun Suk [1 ]
Ahn, Eun Jin [3 ]
Kang, Hyun [2 ]
机构
[1] Jeju Natl Univ, Dept Anesthesiol & Pain Med, Coll Med, Jeju, South Korea
[2] Chung Ang Univ, Coll Med, Dept Anesthesiol & Pain Med, Seoul 156756, South Korea
[3] SungKyunKwan Univ Hosp, Dept Anesthesiol & Pain Med, Seoul, South Korea
来源
PLOS ONE | 2015年 / 10卷 / 03期
关键词
SUPRAGLOTTIC AIRWAY; LMA; CROSSOVER; PRESSURE;
D O I
10.1371/journal.pone.0119469
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objectives Conflicting results have been reported for the i-gel and the laryngeal mask airway proseal (LMA-P) during general anesthesia. The objective of the current investigation was to compare the efficacy and safety of the i-gel vs. the LMA-P during general anesthesia. Methods Two authors performed searches of MEDLINE, EMBASE, CENTRAL, and Google Scholar to identify randomized clinical trials that compared the LMA-P with the i-gel during general anesthesia. A meta -analysis was performed using both random and fixed-effect models. Publication bias was evaluated using Begg's funnel plot and Egger's linear regression test. Results Twelve randomized clinical trials met the eligibility criteria. There were no significant differences in insertion success rate at the first attempt (risk ratio [RR] 1.01, 95% confidence interval [CI] 0.97, 1.06), ease of insertion (RR 1.14, 95% CI 0.93, 1.39), oropharyngeal leak pressure (OLP) (MD -1.98, 95% CI -5.41, 1.45), quality of fiberoptic view (RR 1.00, 95% CI 0.91, 1.10) and success rate of gastric tube insertion (RR 1.07, 95% CI 0.98, 1.18) between the i-gel and the LMA-P, respectively. The i-gel had a shorter insertion time than the LMA-P (MD -3.99, 95% CI -7.13, -0.84) and a lower incidence of blood staining on the device (RR 0.26, 95% CI 0.14, 0.49), sore throat (RR 0.28, 95% CI 0.15, 0.50) and dysphagia (RR 0.27, 95% CI 0.10, 0.74). Conclusions Both devices were comparable in ease of insertion to insert and both had sufficient OLP to provide a reliable airway. Only a few minor complications were reported. The i-gel was found to have fewer complications (blood staining, sore throat, dysphagia) than the LMA-P and offers certain advantages over the LMA-P in adults under general anesthesia.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Pulmonary aspiration associated with supraglottic airways: Proseal laryngeal mask airway and I-Gel(TM)
    Kim, Yoon-Hee
    KOREAN JOURNAL OF ANESTHESIOLOGY, 2012, 63 (06) : 489 - 490
  • [22] Comparison of two supraglottic airway devices, the i-gel airway and the ProSeal laryngeal mask airway, in non-paralysed anaesthetized patients
    Jain, A.
    Luthra, A.
    Bhukal, I.
    Bala, I.
    BRITISH JOURNAL OF ANAESTHESIA, 2016, 116 (06) : E907 - E907
  • [23] A randomised trial to compare i-gel and ProSeal (TM) laryngeal mask airway for airway management in paediatric patients
    Nirupa, R.
    Gombar, Satinder
    Ahuja, Vanita
    Sharma, Preeti
    INDIAN JOURNAL OF ANAESTHESIA, 2016, 60 (10) : 726 - 731
  • [24] Clinical Comparison of I-gel and Laryngeal Mask Airway-Supreme Airway Devices During General Anaesthesia in the Paediatric Population
    Aggarwal, Manish
    Yadav, Rahul
    Singh, Shalendra
    Bansal, Deepali
    TURKISH JOURNAL OF ANAESTHESIOLOGY AND REANIMATION, 2021, 49 (03) : 244 - 249
  • [25] Comparison of oropharyngeal leak pressure and clinical performance of LMA ProSeal and i-gel® in adults: Meta-analysis and systematic review
    Shin, Hye Won
    Yoo, Hae Na
    Bae, Go Eun
    Chang, Jun Chul
    Park, Min Kyung
    You, Hae Seun
    Kim, Hyun Jung
    Ahn, Hyung Sik
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2016, 44 (03) : 405 - 418
  • [26] Comparison of Supreme Laryngeal Mask Airway and ProSeal Laryngeal Mask Airway during Cholecystectomy
    Hosten, Tulay
    Yildiz, Tulay Sahin
    Kus, Alparslan
    Solak, Mine
    Toker, Kamil
    BALKAN MEDICAL JOURNAL, 2012, 29 (03) : 314 - 319
  • [27] Comparison of laryngeal mask airway Supreme and laryngeal mask airway Pro-Seal for controlled ventilation during general anaesthesia in adult patients Systematic review with meta-analysis
    Maitra, Souvik
    Khanna, Puneet
    Baidya, Dalim Kumar
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2014, 31 (05) : 266 - 273
  • [28] A comparative evaluation of ProSeal laryngeal mask airway, I-gel and Supreme laryngeal mask airway in adult patients undergoing elective surgery: A randomised trial
    Singh, Anisha
    Bhalotra, Anju R.
    Anand, Raktima
    INDIAN JOURNAL OF ANAESTHESIA, 2018, 62 (11) : 858 - 864
  • [29] Comparison of the i-gel™ and the Laryngeal Mask Airway Classic™ in terms of clinical performance
    Polat, Reyhan
    Aydin, Gozde Bumin
    Ergil, Julide
    Sayin, Murat
    Kokulu, Tugba
    Ozturk, Ibrahim
    REVISTA BRASILEIRA DE ANESTESIOLOGIA, 2015, 65 (05): : 343 - 348
  • [30] Comparison of the Intersurgical SolusTM laryngeal mask airway and the i-gel supralaryngeal device
    Amini, S.
    Khoshfetrat, M.
    ANAESTHESIA, 2010, 65 (08) : 805 - 809