Scatter Correction Requirements for Likelihood-Based Attenuation Artifact Correction in PET

被引:5
|
作者
Laymon, C. M. [1 ]
Bowsher, J. E. [2 ]
Carney, J. P. J. [1 ]
Blodgett, T. M. [1 ]
机构
[1] Univ Pittsburgh, Pittsburgh, PA 15260 USA
[2] Duke Univ, Durham, NC 27706 USA
关键词
Positron emission tomography; Xray tomography; Image reconstruction; Attenuation measurement;
D O I
10.1109/NSSMIC.2006.354340
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Artifacts have been observed in CT-based, 511keV attenuation images for PET attenuation correction. Artifact detection and correction algorithms are being developed based on the Poisson loglikelihood function (LLF). The LLF quantifies the level of consistency between an activity and attenuation image pair and the measured emission sinogram. We have shown that the LLF can discriminate more-artifactual from less-artifactual attenuation images at clinical noise levels. Here we use a digital anthropomorphic phantom to examine the effect of scatter on LLF performance. Base sinograms were formed by forward projecting a true activity distribution through a true attenuation distribution. Scatter was simulated by filtering the base sinograms. Noise was considered by producing 128 Poisson-randomized variates of the filtered base sinograms for each of 7 assumed count levels. An artifactual attenuation image was produced by modifying mu-values of regions of the true attenuation image. For each sinogram, two OSEM reconstructions were performed and two LLF values obtained using the artifactual and true attenuation. Accuracy was calculated as the fraction of cases in which the LLF preferred the artifact-free attenuation image. Additional studies were performed to test several scatter correction procedures. Findings show that scatter may render the LLF unreliable even in the limit of high-count scans but that its usefulness is restored by applying an appropriate correction.
引用
收藏
页码:2151 / 2154
页数:4
相关论文
共 50 条
  • [31] Scatter Artifact with Ga-68-PSMA-11 PET: Severity Reduced With Furosemide Diuresis and Improved Scatter Correction
    Lawhn-Heath, Courtney
    Flavell, Robert R.
    Korenchan, David E.
    Deller, Timothy
    Lake, Spencer
    Carroll, Peter R.
    Hope, Thomas A.
    MOLECULAR IMAGING, 2018, 17
  • [32] ATTENUATION AND SCATTER CORRECTION AIMING AT QUANTITATIVE SPECT
    AXELSSON, B
    ISRAELSSON, A
    LARSSON, SA
    MSAKI, P
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1984, 9 (07): : A83 - A83
  • [33] Motion Correction and Attenuation Correction in Thoracic PET Imaging
    Bai, Wenjia
    Brady, Sir Michael
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3593 - 3598
  • [34] Effect of Scatter Correction When Comparing Attenuation Maps: Application to Brain PET/MR
    Burgos, Ninon
    Thielemans, Kris
    Cardoso, M. Jorge
    Markiewicz, Pawel
    Jiao, Jieqing
    Dickson, John
    Duncan, John S.
    Atkinson, David
    Arridge, Simon R.
    Hutton, Brian F.
    Ourselin, Sebastien
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [35] Occurrence of an Artifact in Brain F-18-FDG PET with Calculated Attenuation Correction
    von Borczyskowski, Daniel
    Schulte, Ulrike
    Brenner, Winfried
    Clausen, Malte
    Buchert, Ralph
    JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2006, 34 (04) : 232 - 234
  • [36] Metal artifact reduction strategies for improved attenuation correction in hybrid PET/CT imaging
    Abdoli, Mehrsima
    Dierckx, Rudi A. J. O.
    Zaidi, Habib
    MEDICAL PHYSICS, 2012, 39 (06) : 3343 - 3360
  • [37] Empirical Scatter Correction (ESC): A New CT Scatter Correction Method and its Application to Metal Artifact Reduction
    Meyer, Esther
    Maass, Clemens
    Baer, Matthias
    Raupach, Rainer
    Schmidt, Bernhard
    Kachelriess, Marc
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2036 - 2041
  • [38] Attenuation artifact, attenuation correction, and the future of myocardial perfusion SPECT
    Bhupinder Singh
    Timothy M. Bateman
    James A. Case
    Gary Heller
    Journal of Nuclear Cardiology, 2007, 14
  • [39] Attenuation artifact, attenuation correction, and the future of myocardial perfusion SPELT
    Singh, Bhupinder
    Bateman, Timothy M.
    Case, James A.
    Heller, Gary
    JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (02) : 153 - 164
  • [40] Likelihood-based cointegration analysis in panels of vector error-correction models
    Groen, JJJ
    Kleibergen, F
    JOURNAL OF BUSINESS & ECONOMIC STATISTICS, 2003, 21 (02) : 295 - 318