Detection and classification of gaseous microemboli during Pulsatile and nonpulsatile perfusion in a simulated neonatal CPB model

被引:29
|
作者
Uendar, Akif
Ji, Bingyang
Kunselman, Allen R.
Myers, John L.
机构
[1] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Pediat, Hershey, PA USA
[2] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Surg, Hershey, PA USA
[3] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Bioengn, Hershey, PA USA
[4] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Hlth Evaluat Sci, Hershey, PA USA
关键词
D O I
10.1097/MAT.0b013e3181588dc3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We compared the effects of perfusion modes (pulsatile vs. nonpulsatile) on gaseous microemboli delivery using the Emboli Detection and Classification (EDACTM) Quantifier at postpump, postoxygenator, and postarterial filter sites in a simulated pediatric cardiopulmonary bypass (CPB) model. The mock loop was subjected to five different pump flow rates of equal 100 ml/min intervals, ranging from 400 to 800 ml/min. When the target pump flow rate was achieved, 5 cc air was introduced into the venous line. The EDACTM system recorded gaseous microemboli counts simultaneously at three locations in 5-minute intervals. Regardless of the type of perfusion mode, when the pump flow rate was increased, more gaseous microemboli were generated at postpump site. Compared with nonpulsatile flow, pulsatile flow did deliver significantly more gaseous microemboli at postpump site, but there was no difference between two groups at postoxygenator and postarterial filter sites. Capiox Baby-RX hollow-fiber membrane oxygenator significantly reduced the gaseous microemboli counts in both groups at all five pump flow rates with either pulsatile flow or nonpulsatile flow in this model. Our results suggest that using this novel EDAC (TM) system, we could detect the size of gaseous microemboli, as small as 10 mu m, and the percentage of detected gaseous microemboli, <40 mu m, was about 90% in total gaseous microemboli counts at any flow rate with pulsatile or nonpulsatile flow.
引用
收藏
页码:725 / 729
页数:5
相关论文
共 50 条
  • [1] Comparison of two different blood pumps on delivery of gaseous microemboli during pulsatile and nonpulsatile perfusion in a simulated infant CPB model
    Wang, Shigang
    Kunselman, Allen R.
    Myers, John L.
    Undar, Akif
    ASAIO JOURNAL, 2008, 54 (05) : 538 - 541
  • [2] Delivery of gaseous microemboli with vacuum-assisted venous drainage during pulsatile and nonpulsatile perfusion in a simulated neonatal cardiopulmonary bypass model
    Wang, Shigang
    Baer, Larry
    Kunselman, Allen R.
    Myers, John L.
    Uendar, Akif
    ASAIO JOURNAL, 2008, 54 (04) : 416 - 422
  • [3] The capability of trapping gaseous microemboli of two pediatric arterial filters with pulsatile and nonpulsatile flow in a simulated infant CPB model
    Wang, Shigang
    Win, Khin N.
    Kunselman, Allen R.
    Woitas, Karl
    Myers, John L.
    Undar, Akif
    ASAIO JOURNAL, 2008, 54 (05) : 519 - 522
  • [4] Gaseous microemboli detection in a simulated pediatric CPB circuit using a novel ultrasound system
    Miller, Akemi
    Wang, Shigang
    Myers, John L.
    Undar, Akif
    ASAIO JOURNAL, 2008, 54 (05) : 504 - 508
  • [5] Impact of membrane oxygenators on pulsatile versus nonpulsatile perfusion in a neonatal model
    Ündar, A
    Koenig, KM
    Frazier, OH
    Fraser, CD
    PERFUSION-UK, 2000, 15 (02): : 111 - 120
  • [6] Handling ability of gaseous microemboli of two pediatric arterial filters in a simulated CPB model
    Strother, A.
    Wang, S.
    Kunselman, A. R.
    Uendar, A.
    PERFUSION-UK, 2013, 28 (03): : 244 - 252
  • [7] Comparison of four different pediatric 10F aortic cannulae during Pulsatile versus nonpulsatile perfusion in a simulated neonatal model of cardiopulmonary bypass
    Uendar, Akif
    Ji, Bingyang
    Rider, Alan
    Lukic, Branka
    Kunselman, Allen R.
    Weiss, William J.
    Myers, John L.
    ASAIO JOURNAL, 2007, 53 (06) : 778 - 784
  • [8] REGIONAL BLOOD FLOW DURING PULSATILE AND NONPULSATILE PERFUSION
    MANDELBAUM, I
    BERRY, J
    SILBERT, M
    BURNS, WH
    ROTHE, C
    ARCHIVES OF SURGERY, 1965, 91 (05) : 771 - +
  • [9] Cardiac surgery during pregnancy: Pulsatile or nonpulsatile perfusion?
    Jahangiri, M
    Clark, J
    Prefumo, F
    Pumphrey, C
    Ward, D
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2003, 126 (03): : 894 - 895
  • [10] Microemboli detection and classification by innovative ultrasound technology during simulated neonatal cardiopulmonary bypass at different flow rates, perfusion modes, and perfusate temperatures
    Schreiner, Robert S.
    Rider, Alan R.
    Myers, John W.
    Ji, Bingyang
    Kunselman, Allen R.
    Myers, John L.
    Uendar, Akif
    ASAIO JOURNAL, 2008, 54 (03) : 316 - 324